510(k) Report

510(k) · Cleared

Recalled · 2010

Fastx Sternal Intraosseous Device

K100124510(k) TraditionalCleared Aug 31, 2010224-day reviewClass II21 CFR 880.5570Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pyng Medical Corp. · FMI Needle, Hypodermic, Single Lumen

Cleared on Aug 31, 2010 as substantially equivalent to K080865 (Pyng Medical Corp., 2008) after a 224-day review. Its family tree reaches back 5 generations to 1992. 2 later devices descend from it. It has been recalled once.

Decided
Aug 31, 2010
Review
224 days
Class
Class II
Product code
FMI
Recalls
1
Predicates
1

Its longest line of descent runs back 5 generations to K915409 (1992). 11 ancestors in all, the oldest K883014 (1988). 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2004: 1 other ancestor1997: 1 other ancestor1998: 1 other ancestor1992: 1 other ancestor1988: 1 other ancestor2013: 1 descendant cleared2018: 1 descendant clearedK915409 · Intraosseous Access Needle · Cook, Inc. · 1992K022415 · Bone Injection Gun (B.I.G.) for Pediatrics, Model 1ST · Waismed, Ltd. · 2002K043490 · PD-IO Disposable Intraosseous Infusion Needle · Vidacare Corporation · 2005K052195 · EZ-Mio Manual Driver · Vidacare Corporation · 2005K080865 · FAST1 Intraosseous Infusion System · Pyng Medical Corp. · 20081990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K080865 · FAST1 Intraosseous Infusion System · Pyng Medical Corp. · 2008K052195 · EZ-Mio Manual Driver · Vidacare Corporation · 2005K043490 · PD-IO Disposable Intraosseous Infusion Needle · Vidacare Corporation · 2005K022415 · Bone Injection Gun (B.I.G.) for Pediatrics, Model 1ST · Waismed, Ltd. · 2002K915409 · Intraosseous Access Needle · Cook, Inc. · 1992−50
  1. Root · 5 generations backK915409 Intraosseous Access NeedleCook, Inc. · 1992
  2. 4 generations backK022415 Bone Injection Gun (B.I.G.) for Pediatrics, Model 1STWaismed, Ltd. · 2002
  3. 3 generations backK043490 PD-IO Disposable Intraosseous Infusion NeedleVidacare Corporation · 2005
  4. 2 generations backK052195 EZ-Mio Manual DriverVidacare Corporation · 2005
  5. 1 generation backK080865 FAST1 Intraosseous Infusion SystemPyng Medical Corp. · 2008
  6. This deviceK100124 Fastx Sternal Intraosseous DevicePyng Medical Corp. · 2010
  7. Descendant · cites it directly · 1 of its ownK130487 Fastresponder Sternal Intraosseous DevicePyng Medical Corp. · 2013
  8. Descendant · 2 generations on · 0 of its ownK180588 FAST1™ Intraosseous Infusion System; FASTResponder™ Sternal…Teleflex Medical · 2018

Most influential descendants

Showing 2 of 2 · See all
  • K130487Fastresponder Sternal Intraosseous DevicePyng Medical Corp. · 2013 · cites it directly1descendant
  • K180588FAST1™ Intraosseous Infusion System; FASTResponder™ Sternal Intraosseous DeviceTeleflex Medical · 2018 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K915409Intraosseous Access NeedleCook, Inc.1992—
−4K022415Bone Injection Gun (B.I.G.) for Pediatrics, Model 1STWaismed, Ltd.2002—
−3K043490PD-IO Disposable Intraosseous Infusion NeedleVidacare Corporation2005—
−2K052195EZ-Mio Manual DriverVidacare Corporation2005—
−1K080865FAST1 Intraosseous Infusion SystemPyng Medical Corp.2008—
0 (this device)K100124Fastx Sternal Intraosseous DevicePyng Medical Corp.2010—
+1K130487Fastresponder Sternal Intraosseous DevicePyng Medical Corp.2013—
+2K180588FAST1™ Intraosseous Infusion System; FASTResponder™ Sternal Intraosseous DeviceTeleflex Medical2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199210
199710
199810
200410
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201810
FIG. 1 — Lineage of K100124, 1992–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FMI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMI device

  1. #1 K252886 Pen Needlescore 14.5 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080865FAST1 Intraosseous Infusion System
Summary page 1
…1II. Predicate Devices FASTX' m STERNAL INTRAOSSEOUS DEVICE i5 substantially equivalent to the FAST1I® Intraosseous Infusion System cleared under K080865. IV. Indications for Use of the Device Indications for Use: FASTXTM Sternal Intraosseous Device is indic…
Pyng Medical Corp.2008—

Also cleared against K080865: 2 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K130487Fastresponder Sternal Intraosseous DevicePyng Medical Corp.20131

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 5, 2010Not classifiedZ-2687-2011OtherDevice failed to operate; the insertion of the infusion tube of the FASTx Sternal Intraosseous Device was not achieved in live subjects.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 4 companies; a new generation was cleared every 2.4 years on median.
  • None of the 11 ancestors has a recall.
  • Of its descendants, K130487 (Pyng Medical Corp., 2013) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K080865.

How does its category compare?

All FMI devices

FMI Needle, Hypodermic, Single Lumen: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · FMI

11in 2025

2007: 152008: 102009: 72010: 152011: 122012: 172013: 112014: 142015: 102016: 202017: 92018: 202019: 122020: 112021: 182022: 232023: 182024: 32025: 112026 (partial year): 12
2007high 23 (2022) · low 3 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMI

141in 2025

2007: 71 days median2008: 85 days median2009: 177 days median2010: 119 days median2011: 113 days median2012: 107 days median2013: 87 days median2014: 155 days median2015: 89 days median2016: 217 days median2017: 253 days median2018: 176 days median2019: 186 days median2020: 263 days median2021: 143 days median2022: 189 days median2023: 112 days median2024: 90 days median2025: 141 days median2026 (partial year): 119 days median
2007high 263 (2020) · low 71 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMI

1in 2024

2007: 22008: 42009: 32010: 792011: 52012: 12013: 12014: 32015: 12017: 162019: 52020: 32021: 72023: 22024: 12026 (partial year): 1, 1 Class I
2007high 79 (2010) · low 0 (2016)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · FMI

2,424in 2025

2007: 2082008: 1772009: 2632010: 2492011: 2402012: 2152013: 2552014: 2872015: 5672016: 9412017: 1,5472018: 2,3982019: 4,8962020: 4,2452021: 4,5322022: 4,8852023: 4,9862024: 4,0532025: 2,424
2007high 4,986 (2023) · low 177 (2008)2025

MAUDE reports naming product code FMI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715712208
200810854177
200971773263
20101511979249
2011121135240
2012171071215
201311871255
2014141553287
201510891567
2016202170941
20179253161,547
20182017602,398
20191218654,896
20201126334,245
20211814374,532
20222318904,885
20231811224,986
202439014,053
20251114102,424
2026121191 (1)—

What is it indicated for?

FASTxTM Sternal Intraosseous Device is indicated for use in establishing a sternal intraosseous access route in adult and adolescent patients (12 years of age and older) requiring vascular administration of drugs or fluids to facilitate emergency resuscitation.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K100124: Fastx Sternal Intraosseous Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K100124

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026