Wireless Body Temperature Monitor Model BTM-Dix Series (BTM-Dic,D,E,F,G & H)

K100226510(k) SpecialCleared Mar 18, 201052-day reviewClass II21 CFR 880.2910

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Rio Flexon Technology Co., Ltd. · FLL Continuous Measurement Thermometer

Cleared on Mar 18, 2010 as substantially equivalent to K081256 (Rio Flexon Technology Co., Ltd., 2008) after a 52-day review. Its family tree reaches back 5 generations to 2001. 37 later devices descend from it.

Decided
Mar 18, 2010
Review
52 days
Class
Class II
Product code
FLL
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K010462 (2001). 7 ancestors in all, the oldest K003326 (2001). 37 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 2 other ancestors2012: 1 descendant cleared2014: 1 descendant cleared2016: 2 descendants cleared2021: 5 descendants cleared2018: 2 descendants cleared2023: 6 descendants cleared2019: 1 descendant cleared2022: 3 descendants cleared2024: 7 descendants cleared2020: 1 descendant cleared2025: 6 descendants cleared2026: 2 descendants clearedReference device: K050463 · Clever TD-1107 Ear/Skin/Surface IR Thermometer, Clever TD-1110 Ear/Skin/Surface IR Thermometer · Taidoc Technology Corporation · 2005K010462 · Avita Piccolo/Exato Infared Ear Thermometer (TS2 & TS4 Series) · Avita Corporation · 2001K031503 · Avita TS8/TS9 Series IR Ear/Forehead Thermometer · Avita Corporation · 2003K050463 · Clever TD-1107 Ear/Skin/Surface IR Thermometer, Clever TD-1110 Ear/Skin/Surface IR Thermometer · Taidoc Technology Corporation · 2005K062445 · Monitoring Ear Thermometer, Model BTM-D1A · Rio Flexon Technology Co., Ltd. · 2007K081256 · Wireless Body Temperature Monitor, Models BTM-Dic, D, E, F, G and H · Rio Flexon Technology Co., Ltd. · 2008200020052010201520202025
Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor5 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 3 descendants4 generations forward: 3 descendants5 generations forward: 8 descendants6 generations forward: 10 descendants7 generations forward: 5 descendants8 generations forward: 6 descendants K081256 · Wireless Body Temperature Monitor, Models BTM-Dic, D, E, F, G and H · Rio Flexon Technology Co., Ltd. · 2008K062445 · Monitoring Ear Thermometer, Model BTM-D1A · Rio Flexon Technology Co., Ltd. · 2007K050463 · Clever TD-1107 Ear/Skin/Surface IR Thermometer, Clever TD-1110 Ear/Skin/Surface IR Thermometer · Taidoc Technology Corporation · 2005K031503 · Avita TS8/TS9 Series IR Ear/Forehead Thermometer · Avita Corporation · 2003K010462 · Avita Piccolo/Exato Infared Ear Thermometer (TS2 & TS4 Series) · Avita Corporation · 2001−50+5
  1. Root · 5 generations backK010462 Avita Piccolo/Exato Infared Ear Thermometer (TS2 & TS4 Series)Avita Corporation · 2001
  2. 4 generations backK031503 Avita TS8/TS9 Series IR Ear/Forehead ThermometerAvita Corporation · 2003
  3. 3 generations backK050463 Clever TD-1107 Ear/Skin/Surface IR Thermometer, Clever TD-1110…Taidoc Technology Corporation · 2005
  4. 2 generations backK062445 Monitoring Ear Thermometer, Model BTM-D1ARio Flexon Technology Co., Ltd. · 2007
  5. 1 generation backK081256 Wireless Body Temperature Monitor, Models BTM-Dic, D, E, F, G and HRio Flexon Technology Co., Ltd. · 2008
  6. This deviceK100226 Wireless Body Temperature Monitor Model BTM-Dix Series…Rio Flexon Technology Co., Ltd. · 2010
  7. Descendant · cites it directly · 36 of its ownK121696 Wireless ThermometerRaling Medical Company · 2012
  8. Descendant · 2 generations on · 35 of its ownK132761 Wireless ThermometerRaiing Medical Company · 2014
  9. Descendant · 3 generations on · 27 of its ownK160306 Cloud Smart ThermometerGuangzhou Bosma Technology Co, Ltd. · 2016
  10. Descendant · 4 generations on · 26 of its ownK181013 Fever ScoutVivaLNK, Inc. · 2018
  11. Descendant · 5 generations on · 16 of its ownK191272 Wearable Device, Wearable PackageCurrent Health, Ltd. · 2019

Showing 5 of 37 descendants. See all descendants

Descendants cleared per year

2012: 12014: 12016: 22018: 22019: 12020: 12021: 52022: 32023: 62024: 72025: 62026: 2
2012high 7 (2024) · low 0 (2013)2026

Most influential descendants

Showing 5 of 37 · See all
  • K121696Wireless ThermometerRaling Medical Company · 2012 · cites it directly36descendants
  • K132761Wireless ThermometerRaiing Medical Company · 2014 · 2 generations on35descendants
  • K160306Cloud Smart ThermometerGuangzhou Bosma Technology Co, Ltd. · 2016 · 3 generations on27descendants
  • K181013Fever ScoutVivaLNK, Inc. · 2018 · 4 generations on26descendants
  • K191272Wearable Device, Wearable PackageCurrent Health, Ltd. · 2019 · 5 generations on16descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K010462Avita Piccolo/Exato Infared Ear Thermometer (TS2 & TS4 Series)Avita Corporation2001—
−4K031503Avita TS8/TS9 Series IR Ear/Forehead ThermometerAvita Corporation2003—
−3K050463Clever TD-1107 Ear/Skin/Surface IR Thermometer, Clever TD-1110…Taidoc Technology Corporation2005—
−2K062445Monitoring Ear Thermometer, Model BTM-D1ARio Flexon Technology Co., Ltd.2007—
−1K081256Wireless Body Temperature Monitor, Models BTM-Dic, D, E, F, G and HRio Flexon Technology Co., Ltd.2008—
0 (this device)K100226Wireless Body Temperature Monitor Model BTM-Dix Series (BTM-Dic,D,E,F,G & H)Rio Flexon Technology Co., Ltd.2010—
+1K121696Wireless ThermometerRaling Medical Company2012—
+2K132761Wireless ThermometerRaiing Medical Company2014—
+3K160306Cloud Smart ThermometerGuangzhou Bosma Technology Co, Ltd.2016—
+4K181013Fever ScoutVivaLNK, Inc.2018—
+5K191272Wearable Device, Wearable PackageCurrent Health, Ltd.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200120

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201410
201620
201820
201910
202010
202150
202230
202360
202470
202560
202620
FIG. 1 — Lineage of K100226, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLL device

  1. #1 K191829 Microlife Non-Contact Infrared Forehead…score 26.7 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K081256Wireless Body Temperature Monitor, Models BTM-Dic, D, E, F, G and H
Summary page 2
…cording to the US standard ASTM El 112-00 and ASTM E1965-02. Comparison to Predicate Devices: The new BTM-D1X series is the same as the predicate BTM-DIX series under K081256, except for the addition of the high/low alarm function. Prior BTM-…
Rio Flexon Technology Co., Ltd.2008—
†K050463Clever TD-1107 Ear/Skin/Surface IR Thermometer, Clever TD-1110 Ear/Skin/Surface IR…Taidoc Technology Corporation2005—

What descends from it? (37)

See all 37

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K121696Wireless ThermometerRaling Medical Company201236

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 3 companies; a new generation was cleared every 1.8 years on median.
  • None of the 7 ancestors has a recall.
  • Of its descendants, K121696 (Raling Medical Company, 2012) has gone furthest, with 36 descendants of its own.

How does its category compare?

All FLL devices

FLL Continuous Measurement Thermometer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · FLL

6in 2025

2007: 152008: 142009: 132010: 242011: 202012: 172013: 142014: 72015: 32016: 222017: 162018: 352019: 222020: 212021: 432022: 242023: 172024: 132025: 62026 (partial year): 2
2007high 43 (2021) · low 3 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLL

131in 2025

2007: 81 days median2008: 59 days median2009: 84 days median2010: 88 days median2011: 113 days median2012: 126 days median2013: 149 days median2014: 199 days median2015: 270 days median2016: 157 days median2017: 174 days median2018: 209 days median2019: 198 days median2020: 194 days median2021: 218 days median2022: 218 days median2023: 295 days median2024: 223 days median2025: 131 days median
2007high 295 (2023) · low 59 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLL

1in 2025

2011: 12015: 12017: 22018: 12019: 52025: 12026 (partial year): 1
2007high 5 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLL

73in 2025

2007: 42008: 132009: 92010: 92011: 182012: 262013: 162014: 542015: 532016: 582017: 1182018: 512019: 522020: 532021: 1152022: 1422023: 1012024: 1182025: 73
2007high 142 (2022) · low 4 (2007)2025

MAUDE reports naming product code FLL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007158104
20081459013
2009138409
2010248809
201120113118
201217126026
201314149016
20147199054
20153270153
201622157058
2017161742118
201835209151
201922198552
202021194053
2021432180115
2022242180142
2023172950101
2024132230118
20256131173
20262—1—

What is it indicated for?

The Wireless Body Temperature Monitor, model BIM-DIX series (model BTM-D1IC, D, E, F, O and H), is a battery-operated electronic device with intended use of measuring human ear temperature precisely and continuously monitoring armpit temperature via wireless signal transmission of the measuring result. This device is reusable and is intended for ear temperature measurement as well as armpit temperature monitoring for persons over two years old. Model BTM-D1C has two functions, both ear temperature measurement and armpit temperature monitoring. The rest of models, BTM-D1D, E, F, U and H, have only the function of armpit temperature monitoring.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K100226: Wireless Body Temperature Monitor Model BTM-Dix Series (BTM-Dic,D,E,F,G & H).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K100226

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026