DePuy Orthopaedics, Inc. · HSB Rod, Fixation, Intramedullary And Accessories

Cleared on May 11, 2010 as substantially equivalent to K042325 (DePuy Orthopaedics, Inc., 2004) and 1 other predicate after a 104-day review. Its family tree reaches back 4 generations to 1997. 7 later devices descend from it. It has been recalled 250 times.

Decided
May 11, 2010
Review
104 days
Class
Class II
Product code
HSB
Recalls
250
Predicates
2

Its longest line of descent runs back 4 generations to K970097 (1997). 11 ancestors in all, the oldest K871539 (1987). 7 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor2004: 1 other ancestor2003: 1 other ancestor2001: 2 other ancestors1998: 1 other ancestor1987: 1 other ancestor2011: 1 descendant cleared2018: 1 descendant cleared2020: 1 descendant cleared2022: 1 descendant cleared2024: 2 descendants cleared2021: 1 descendant clearedK970097 · Proximal Femoral Nail (PFN) System · Synthes (Usa) · 1997K973240 · Synthes Proximal Femoral Nail (PFN) System Modification · Synthes (Usa) · 1997K013563 · Long Trochanteric Nail System · DePuy Orthopaedics, Inc. · 2001 · recalledK042325 · Trochanteric Nail System · DePuy Orthopaedics, Inc. · 2004 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 3 other ancestors 1 generation forward: 3 descendants2 generations forward: 4 descendants K042325 · Trochanteric Nail System · DePuy Orthopaedics, Inc. · 2004 · recalledK013563 · Long Trochanteric Nail System · DePuy Orthopaedics, Inc. · 2001 · recalledK973240 · Synthes Proximal Femoral Nail (PFN) System Modification · Synthes (Usa) · 1997K970097 · Proximal Femoral Nail (PFN) System · Synthes (Usa) · 19970
  1. Root · 4 generations backK970097 Proximal Femoral Nail (PFN) SystemSynthes (Usa) · 1997
  2. 3 generations backK973240 Synthes Proximal Femoral Nail (PFN) System ModificationSynthes (Usa) · 1997
  3. 2 generations backK013563 Long Trochanteric Nail SystemDePuy Orthopaedics, Inc. · 2001 · Recalled
  4. 1 generation backK042325 Trochanteric Nail SystemDePuy Orthopaedics, Inc. · 2004 · Recalled
  5. This deviceK100238 HFN 125 Degree 9-13 MM, 180MM, HFN 130 Degree 9-13 MM, HFN RH 125…DePuy Orthopaedics, Inc. · 2010
  6. Descendant · cites it directly · 2 of its ownK192003 Auxein Nailing SystemAuxein Medical Private Limited · 2020
  7. Descendant · cites it directly · 2 of its ownK183162 Affixus Hip Fracture NailBiomet, Inc. · 2018
  8. Descendant · 2 generations on · 0 of its ownK230071 Femur Reconstruction Interlocking Nail System, Femur Retrograde…Tdm Co., Ltd. · 2024
  9. Descendant · 2 generations on · 0 of its ownK233910 Vertex Hip Fracture Nailing SystemOrthoxel Dac · 2024
  10. Descendant · 2 generations on · 0 of its ownK221489 Artemis Proximal Femoral Nail SystemGlw, Inc. · 2022

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2011: 12018: 12020: 12021: 12022: 12024: 2
2011high 2 (2024) · low 0 (2012)2026

Most influential descendants

Showing 5 of 7 · See all
  • K192003Auxein Nailing SystemAuxein Medical Private Limited · 2020 · cites it directly2descendants
  • K183162Affixus Hip Fracture NailBiomet, Inc. · 2018 · cites it directly2descendants
  • K230071Femur Reconstruction Interlocking Nail System, Femur Retrograde Interlocking Nail…Tdm Co., Ltd. · 2024 · 2 generations on0descendants
  • K233910Vertex Hip Fracture Nailing SystemOrthoxel Dac · 2024 · 2 generations on0descendants
  • K221489Artemis Proximal Femoral Nail SystemGlw, Inc. · 2022 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K970097Proximal Femoral Nail (PFN) SystemSynthes (Usa)1997—
−3K973240Synthes Proximal Femoral Nail (PFN) System ModificationSynthes (Usa)1997—
−2K013563Long Trochanteric Nail SystemDePuy Orthopaedics, Inc.200154
−1K042325Trochanteric Nail SystemDePuy Orthopaedics, Inc.200411
0 (this device)K100238HFN 125 Degree 9-13 MM, 180MM, HFN 130 Degree 9-13 MM, HFN RH 125 Degree 9-13MMDePuy Orthopaedics, Inc.2010—
+1K192003Auxein Nailing SystemAuxein Medical Private Limited2020—
+1K183162Affixus Hip Fracture NailBiomet, Inc.2018—
+2K230071Femur Reconstruction Interlocking Nail System, Femur Retrograde Interlocking…Tdm Co., Ltd.2024—
+2K233910Vertex Hip Fracture Nailing SystemOrthoxel Dac2024—
+2K221489Artemis Proximal Femoral Nail SystemGlw, Inc.2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199810
200120
200310
200410
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201810
202010
202110
202210
202420
FIG. 1 — Lineage of K100238, 1997–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HSB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HSB device

  1. #1 K200869 Gamma3 Systemscore 19.8 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042325Trochanteric Nail System
Summary page 1
…acture Nail System is substantially equivalent to currently marketed devices: Trochanteric Nail System (K042325) and Gamma3 Nail System (K043431). Device Description: The Hip Fracture Nail System consists of an intrameduilary…
DePuy Orthopaedics, Inc.200411
*K043431GAMMA3 Nail System
Summary page 1
…ly equivalent to currently marketed devices: Trochanteric Nail System (K042325) and Gamma3 Nail System (K043431). Device Description: The Hip Fracture Nail System consists of an intrameduilary na…
Howmedica Osteonics Corp.2005—

Also cleared against K042325: 1 other device →

What descends from it? (7)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K192003Auxein Nailing SystemAuxein Medical Private Limited20202
K183162Affixus Hip Fracture NailBiomet, Inc.20182
K111444Universal and Troch Entry Femoral Nailing Systems Line ExtensionDePuy Orthopaedics, Inc.20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 2, 2018Class IIZ-0449-2019Mixed-up of materials/componentsThe set screw is not able to advance or reverse from the original position in the nail.
Oct 2, 2018Class IIZ-0452-2019Mixed-up of materials/componentsThe set screw is not able to advance or reverse from the original position in the nail.
Oct 2, 2018Class IIZ-0447-2019Mixed-up of materials/componentsThe set screw is not able to advance or reverse from the original position in the nail.
Oct 2, 2018Class IIZ-0454-2019Mixed-up of materials/componentsThe set screw is not able to advance or reverse from the original position in the nail.
Oct 2, 2018Class IIZ-0453-2019Mixed-up of materials/componentsThe set screw is not able to advance or reverse from the original position in the nail.
Oct 2, 2018Class IIZ-0451-2019Mixed-up of materials/componentsThe set screw is not able to advance or reverse from the original position in the nail.
Oct 2, 2018Class IIZ-0448-2019Mixed-up of materials/componentsThe set screw is not able to advance or reverse from the original position in the nail.
Oct 2, 2018Class IIZ-0450-2019Mixed-up of materials/componentsThe set screw is not able to advance or reverse from the original position in the nail.
Aug 8, 2018Class IIZ-0164-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0140-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0151-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0150-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0134-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0139-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0146-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0152-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0157-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0166-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0168-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0124-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0129-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0155-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0154-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0158-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0161-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0163-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0149-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0133-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0159-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0137-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0138-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0136-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0142-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0126-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0156-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0141-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0165-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0148-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0167-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0153-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0169-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0128-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0130-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0125-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0123-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0162-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0131-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0143-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0122-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.
Aug 8, 2018Class IIZ-0144-2019Employee errorThere were complaints indicating that when the product was opened in surgery, the sterile packaging was already open.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 2 companies; a new generation was cleared every 4.1 years on median.
  • 8 of the 11 ancestors have a recall of any class.
  • Of its descendants, K192003 (Auxein Medical Private Limited, 2020) has gone furthest, with 2 descendants of its own.
  • 1 other device also cites its predicate K042325.

How does its category compare?

All HSB devices

HSB Rod, Fixation, Intramedullary And Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · HSB

15in 2025

2007: 82008: 112009: 182010: 142011: 162012: 172013: 182014: 192015: 222016: 82017: 232018: 172019: 132020: 172021: 102022: 122023: 112024: 132025: 152026 (partial year): 18
2007high 23 (2017) · low 8 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HSB

106in 2025

2007: 68 days median2008: 57 days median2009: 86 days median2010: 117 days median2011: 78 days median2012: 112 days median2013: 105 days median2014: 147 days median2015: 111 days median2016: 124 days median2017: 119 days median2018: 134 days median2019: 132 days median2020: 112 days median2021: 94 days median2022: 172 days median2023: 162 days median2024: 195 days median2025: 106 days median2026 (partial year): 151 days median
2007high 195 (2024) · low 57 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HSB

18in 2025

2008: 92009: 812010: 182011: 22012: 72013: 142014: 32015: 32016: 62017: 662018: 562019: 62020: 62021: 22022: 12023: 12024: 12025: 182026 (partial year): 1
2007high 81 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HSB

2,911in 2025

2007: 1622008: 1842009: 3342010: 4542011: 6622012: 1,0792013: 1,7732014: 1,4742015: 1,6712016: 1,9202017: 1,4472018: 3,5262019: 2,5172020: 3,2992021: 4,1112022: 2,7332023: 2,2192024: 2,8692025: 2,911
2007high 4,111 (2021) · low 162 (2007)2025

MAUDE reports naming product code HSB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078680162
200811579184
2009188681334
20101411718454
201116782662
20121711271,079
201318105141,773
20141914731,474
20152211131,671
2016812461,920
201723119661,447
201817134563,526
20191313262,517
20201711263,299
2021109424,111
20221217212,733
20231116212,219
20241319512,869
202515106182,911
2026181511—

What is it indicated for?

The Hip Fracture Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures, including non-union, malunion and tumor resections. The Long Nail system is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone (including prophylactic use) of the trochanteric and diaphyseal areas, impending pathological fractures, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal non-unions, malunions, revision procedures and tumor resections.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K100238: HFN 125 Degree 9-13 MM, 180MM, HFN 130 Degree 9-13 MM, HFN RH 125 Degree 9-13MM.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K100238

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026