# K100278: Quadrox-ID Pediatric Diffusion Membrane Oxygenator with Bioline Coating, Models BE-Hmod 30000-Usa, Beq-Hmod 30000-Usa

- Pathway: 510(k) (Traditional)
- Applicant: Maquet Cardiopulmonary, AG
- Decided: Apr 30, 2010 after 88 days
- Product code: DTZ (Oxygenator, Cardiopulmonary Bypass)
- Device class: 2
- Regulation: 21 CFR 870.4350
- Recalls of this device: 4 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 5 generations back to K932252 (1993). 11 ancestors in all.
6 later devices descend from it; 2 cite it directly.
0 of 18 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K072091: D101 Kids Infant Hollow Fiber Membrane Oxygenator with Integrated Hardshell Cardiotomy/Venous Reservoir, Sorin Group Italia S.R.L. (2007) Summary page 1.
- K071774: Jostra Quadrox D Diffusion Membrane Oxygenator with Bioline Coating, Model BE-Hmod 2000, Beq-Hmod 2030, Maquet Cardiopulmonary, AG (2008) Summary page 1.

## Most influential descendants
- K132829: Quadrox-I, Quadrox-ID, Quadrox-IR Oxygenators (2013), 4 descendants of its own
- K150267: QUADROX-i Adult/Small Adult Oxygenators;QUADROX-iD Adult Oxygenators (2015), 2 descendants of its own
- K203424: Quantum Perfusion Blood Oxygenator ECC VT200-E1, Quantum Perfusion Blood Oxygenator ECC VT200-E2, Quantum Perfusion Blood Oxygenator ECC VT160-E1,  Quantum Perfusion Blood Oxygenator ECC VT160-E2 (2021), 1 descendant of its own
- K240934: Quantum Perfusion Blood Oxygenator ECC VT160-E3 (VT160-E3); Quantum Perfusion Blood Oxygenator ECC VT200-E3 (VT200-E3); Quantum Perfusion Blood Oxygenator VT75- E3 (VT75- E3); (2026), 0 descendants of its own
- K133265: Quadrox-IR Adult and Small Adult (2013), 0 descendants of its own

## Recalls of this device
- May 18, 2023 · Class II · Z-1658-2023: Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, 
- Nov 13, 2019 · Class II · Z-0677-2020: Sterile barrier system may be compromised
- Nov 5, 2014 · Class II · Z-0870-2015: Maquet has confirmed through a customer complaint that while the oxygenators and the pouches for a single lot of Quadrox-ID Pediatric Diffusion Membrane Oxygenator with Bioliine Coating are labeled co
- Sep 22, 2011 · Not classified · Z-0689-2012: There are reports of leakage in the Leuer connector of the blood outflow connector of the Quadrox-iD pediatric oygenator.

## Indications for use
The diffusion membrane oxygenator QUADROX-iD Pediatric is intended for use in an extracorporeal perfusion system. The oxygenator is designed for a blood flow rate of 0.2 - 2.8 I/min and is intended for the treatment of pediatric patients. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide and regulates the blood temperature. The utilization period for this device is restricted to six hours. Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.

Source page: https://510kreport.com/device/K100278