510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Advia Centaur Calibrator 30, Models 10379811/10379810

K100293510(k) AbbreviatedCleared Mar 19, 201045-day reviewClass II21 CFR 862.1150Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Heatlhcare Diagnostics, Inc. · JIT Calibrator, Secondary

Cleared on Mar 19, 2010 as substantially equivalent to K932715 (Ciba Corning Diagnostics Corp., 1993) and 1 other predicate after a 45-day review. Its family tree reaches back 1 generation to 1993. It has been recalled once.

Decided
Mar 19, 2010
Review
45 days
Class
Class II
Product code
JIT
Recalls
1
Predicates
2

Its longest line of descent runs back 1 generation to K932715 (1993). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1995: 1 other ancestorK932715 · Ciba Corning Acs Estradiol Immunoassay · Ciba Corning Diagnostics Corp. · 199319901995200020052010
Other ancestors, by generation 1 generation back: 1 other ancestor K932715 · Ciba Corning Acs Estradiol Immunoassay · Ciba Corning Diagnostics Corp. · 19930
  1. Root · 1 generation backK932715 Ciba Corning Acs Estradiol ImmunoassayCiba Corning Diagnostics Corp. · 1993
  2. This deviceK100293 Advia Centaur Calibrator 30, Models 10379811/10379810Siemens Heatlhcare Diagnostics, Inc. · 2010
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K932715Ciba Corning Acs Estradiol ImmunoassayCiba Corning Diagnostics Corp.1993—
0 (this device)K100293Advia Centaur Calibrator 30, Models 10379811/10379810Siemens Heatlhcare Diagnostics, Inc.2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
FIG. 1 — Lineage of K100293, 1993–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JIT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K932715Ciba Corning Acs Estradiol Immunoassay
Summary page 1
…. Panel: Clinical Chemistry F. Predicate Device: The ADVIA Centaur Calibrator 30 is substantially equivalent to the ADVIA Centaur Calibrator E which was cleared under K932715/K954697 G. Device Description: The ADVIA Centaur® Calibrator 30 is a 2 level human plasma based solutions containing varying con…
Ciba Corning Diagnostics Corp.1993—
*K954697Ciba Corning Acs Estradiol-6 Assay
Summary page 1
…Clinical Chemistry F. Predicate Device: The ADVIA Centaur Calibrator 30 is substantially equivalent to the ADVIA Centaur Calibrator E which was cleared under K932715/K954697 G. Device Description: The ADVIA Centaur® Calibrator 30 is a 2 level human plasma based solutions containing varying concentrati…
Ciba Corning Diagnostics Corp.1995—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 7, 2017Class IIZ-0515-2018Under Investigation by firmSiemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021)…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.

How does its category compare?

All JIT devices

JIT Calibrator, Secondary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · JIT

7in 2017

2007: 52008: 62009: 22010: 62011: 72012: 22013: 52014: 102015: 42016: 42017: 7
2007high 10 (2014) · low 0 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JIT

28in 2017

2007: 78 days median2008: 46 days median2010: 65 days median2011: 48 days median2013: 140 days median2014: 33 days median2015: 30 days median2016: 33 days median2017: 28 days median
2007high 140 (2013) · low 28 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JIT

1in 2023

2007: 12009: 22011: 32012: 32014: 22015: 12018: 52020: 22023: 12026 (partial year): 1
2007high 5 (2018) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JIT

7in 2025

2010: 32011: 22012: 102013: 432014: 92015: 62016: 32017: 72018: 82019: 2152020: 122021: 32022: 42023: 62024: 172025: 7
2007high 215 (2019) · low 0 (2007)2025

MAUDE reports naming product code JIT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200757810
200864600
20092—20
201066503
201174832
20122—310
20135140043
2014103329
201543016
201643303
201772807
20180—58
20190—0215
20200—212
20210—03
20220—04
20230—16
20240—017
20250—07
20260—1—

What is it indicated for?

For in vitro diagnostic use in calibrating the following assays using ADVIA Centaur Systems.

Enhanced Estradiol (eE2)

ADVIA Centaur calibrator 30 is a device intended for medical purposes for use in Estradiol Assay to establish points of reference that are used in the determination of values in the measurement of Estradiol in human serum, Heparinized and EDTA Plasma.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K100293: Advia Centaur Calibrator 30, Models 10379811/10379810.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K100293

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026