510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Arctic Sun Temperature Management System, Model 5000

K101092510(k) TraditionalCleared Jul 16, 201087-day reviewClass II21 CFR 870.5900Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medivance, Inc. · DWJ System, Thermal Regulating

Cleared on Jul 16, 2010 as substantially equivalent to K010338 (Medivance, Inc., 2001) and 1 other predicate after an 87-day review. Its family tree reaches back 2 generations to 1981. 8 later devices descend from it. It has been recalled 3 times.

Decided
Jul 16, 2010
Review
87 days
Class
Class II
Product code
DWJ
Recalls
3
Predicates
2

Its longest line of descent runs back 2 generations to K811742 (1981). 4 ancestors in all. 8 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1991: 1 other ancestor2014: 1 descendant cleared2016: 1 descendant cleared2018: 1 descendant cleared2012: 1 descendant cleared2020: 1 descendant cleared2025: 2 descendants cleared2023: 1 descendant clearedK811742 · Hemotherm Model 400 · Cincinnati Sub-Zero · 1981 · recalledK071341 · Modification to Arctic Sun Temperature Management System, Model 2000 · Medivance, Inc. · 2007 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 3 descendants3 generations forward: 1 descendant K071341 · Modification to Arctic Sun Temperature Management System, Model 2000 · Medivance, Inc. · 2007 · recalledK811742 · Hemotherm Model 400 · Cincinnati Sub-Zero · 1981 · recalled0
  1. Root · 2 generations backK811742 Hemotherm Model 400Cincinnati Sub-Zero · 1981 · Recalled
  2. 1 generation backK071341 Modification to Arctic Sun Temperature Management System, Model 2000Medivance, Inc. · 2007 · Recalled
  3. This deviceK101092 Arctic Sun Temperature Management System, Model 5000Medivance, Inc. · 2010
  4. Descendant · cites it directly · 4 of its ownK142702 Arctic Sun Temperature Management System, ArcticGel Pads (Universal…Medivance, Inc. · 2014 · Recalled
  5. Descendant · cites it directly · 3 of its ownK161602 Arctic Sun Temperature Management SystemMedivance, Inc. · 2016 · Recalled
  6. Descendant · 2 generations on · 1 of its ownK200225 Arctic Sun Stat Temperature Management SystemMedivance, Inc. · 2020 · Recalled
  7. Descendant · cites it directly · 1 of its ownK180375 The IQoolTM Warm SystemBraincool AB · 2018
  8. Descendant · 2 generations on · 0 of its ownK251974 Arctic Sun Temperature Management System (Model 5000)Medivance, Inc. · 2025

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2012: 12014: 12016: 12018: 12020: 12023: 12025: 2
2012high 2 (2025) · low 0 (2013)2026

Most influential descendants

Showing 5 of 8 · See all
  • K142702Arctic Sun Temperature Management System, ArcticGel Pads (Universal, XXS, XS, S, M and…Medivance, Inc. · 2014 · cites it directly4descendants
  • K161602Arctic Sun Temperature Management SystemMedivance, Inc. · 2016 · cites it directly3descendants
  • K200225Arctic Sun Stat Temperature Management SystemMedivance, Inc. · 2020 · 2 generations on1descendant
  • K180375The IQoolTM Warm SystemBraincool AB · 2018 · cites it directly1descendant
  • K251974Arctic Sun Temperature Management System (Model 5000)Medivance, Inc. · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K811742Hemotherm Model 400Cincinnati Sub-Zero19812
−1K071341Modification to Arctic Sun Temperature Management System, Model 2000Medivance, Inc.20071
0 (this device)K101092Arctic Sun Temperature Management System, Model 5000Medivance, Inc.2010—
+1K142702Arctic Sun Temperature Management System, ArcticGel Pads (Universal, XXS, XS…Medivance, Inc.20146
+1K161602Arctic Sun Temperature Management SystemMedivance, Inc.20162
+2K200225Arctic Sun Stat Temperature Management SystemMedivance, Inc.20201
+1K180375The IQoolTM Warm SystemBraincool AB2018—
+2K251974Arctic Sun Temperature Management System (Model 5000)Medivance, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201410
201610
201810
202010
202310
202520
FIG. 1 — Lineage of K101092, 1981–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWJ device

  1. #1 K260278 ThermAffyx (TM) Patient Safety Systemscore 12.8 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K010338Artic Sun Temperature Management System, Model 200
Summary page 1
…ent System, Model 5000, is substantially equivalent to the following predicate devices: Arctic Sun Temperature Management System Medivance, Inc. K010338, - Model 2000 K071 341 Intended use of the device The Arctic…
Medivance, Inc.20011
*K071341Modification to Arctic Sun Temperature Management System, Model 2000
Summary page 1
…Arctic Sun Temperature Management System Medivance, Inc. K010338, - Model 2000 K071 341 Intended use of the device The Arctic Sun Temperature Management System is a thermal regulating system, indicated for monitoring…
Medivance, Inc.20071

Also cleared against K010338: 4 other devices →

What descends from it? (8)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K142702Arctic Sun Temperature Management System, ArcticGel Pads (Universal, XXS, XS, S, M and…Medivance, Inc.20144
K161602Arctic Sun Temperature Management SystemMedivance, Inc.20163
K180375The IQoolTM Warm SystemBraincool AB20181
K121720Welkins Emt/Icu Temperature Managerment SystemWelkins, LLC20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 31, 2017Class IIZ-3042-2017Unknown/Undetermined by firmPotential failure of the Arctic Sun 5000 Temperature Management System may occur when the continued use is in excess of 2000 system hours without preventive maintenance.
Nov 10, 2015Class IIZ-0760-2016Use errorImproper targeted temperature therapy was delivered to patients because the Temperature Simulator Key was connected to the patient temperature cable instead of the patient temperature probe.
Nov 9, 2015Class IIZ-1617-2016Unknown/Undetermined by firmThe product may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery responsible for maintaining the system close and Static Random Access Memory (SRAM). This premature coin cell…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 29 years and passes through 2 companies; a new generation was cleared every 26.0 years on median.
  • 4 of the 4 ancestors have a recall of any class.
  • Of its descendants, K142702 (Medivance, Inc., 2014) has gone furthest, with 4 descendants of its own.
  • 4 other devices also cite its predicate K010338.

How does its category compare?

All DWJ devices

DWJ System, Thermal Regulating: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · DWJ

3in 2025

2007: 42008: 32009: 32010: 82011: 42012: 52013: 52014: 42016: 22017: 52018: 32019: 22020: 52021: 22022: 22023: 62024: 22025: 32026 (partial year): 5
2007high 8 (2010) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWJ

131in 2025

2007: 106 days median2008: 73 days median2009: 106 days median2010: 134 days median2011: 180 days median2012: 69 days median2013: 77 days median2014: 131 days median2017: 153 days median2018: 95 days median2020: 30 days median2023: 42 days median2025: 131 days median2026 (partial year): 99 days median
2007high 180 (2011) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DWJ

3in 2024

2009: 62010: 22011: 72012: 12013: 22015: 12017: 32018: 52019: 32020: 22021: 12022: 52024: 32026 (partial year): 3
2007high 7 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DWJ

2,384in 2025

2007: 522008: 402009: 712010: 1042011: 1362012: 1062013: 732014: 1162015: 2452016: 1,4002017: 1,4732018: 1,5172019: 2,2942020: 2,5462021: 2,6892022: 3,2262023: 3,4552024: 3,3292025: 2,384
2007high 3,455 (2023) · low 40 (2008)2025

MAUDE reports naming product code DWJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074106052
2008373040
20093106671
201081342104
201141807136
20125691106
2013577273
201441310116
20150—1245
20162—01,400
2017515331,473
201839551,517
20192—32,294
202053022,546
20212—12,689
20222—53,226
202364203,455
20242—33,329
2025313102,384
20265993—

What is it indicated for?

The Arctic Sun Temperature Management System is a thermal regulating system, indicated for monitoring and controlling patient temperature.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101092: Arctic Sun Temperature Management System, Model 5000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101092

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026