510(k) · Cleared
ClearedSwabflsuh (3ML, 5ML, 10ML) Model 10331, 10351, 10311
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Excelsior Medical Corp. · NGT Saline, Vascular Access Flush
Cleared on Feb 10, 2011 as substantially equivalent to K962938 (Excelsior Medical Corp., 1996) and 2 other predicates after a 280-day review. Its family tree reaches back 4 generations to 2003.
- Decided
- Feb 10, 2011
- Review
- 280 days
- Class
- Class II
- Product code
- NGT
- Recalls
- None
- Predicates
- 3
Its longest line of descent runs back 4 generations to K030467 (2003). 9 ancestors in all, the oldest K833182 (1983). No later device cites it as a predicate yet.
- Root · 4 generations backK030467 Gravitech TM Flow Controller Primary Set/Gravitech Flow Controller…Iv Medical, Inc. · 2003
- 3 generations backK063239 Lifeshield Latex-Free Gravitech Flow Controller I.V. Sets, Models…Hospira, Inc. · 2006 · Recalled
- 2 generations backK080579 Effectiv (TM) CapHospira, Inc. · 2008
- 1 generation backK083508 SwabcapExcelsior Medical Corp. · 2009
- This deviceK101270 Swabflsuh (3ML, 5ML, 10ML) Model 10331, 10351, 10311Excelsior Medical Corp. · 2011
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K030467 | Gravitech TM Flow Controller Primary Set/Gravitech Flow Controller Extension Set | Iv Medical, Inc. | 2003 | — |
| −3 | K063239 | Lifeshield Latex-Free Gravitech Flow Controller I.V. Sets, Models 17528… | Hospira, Inc. | 2006 | 7 |
| −2 | K080579 | Effectiv (TM) Cap | Hospira, Inc. | 2008 | — |
| −1 | K083508 | Swabcap | Excelsior Medical Corp. | 2009 | — |
| 0 (this device) | K101270 | Swabflsuh (3ML, 5ML, 10ML) Model 10331, 10351, 10311 | Excelsior Medical Corp. | 2011 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1983 | 1 | 0 |
| 1996 | 1 | 0 |
| 2005 | 1 | 0 |
| 2008 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in NGT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new NGT device
- #1 K250884 0.9 % Sodium Chloride Injection, USP, BD…score 11.6 · 5 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K962938 | Excelsior Disposable Syringe W/Normal Saline (.9% Sodium Chloride)Summary page 1…e following devices as related to intended use, technology and design characteristics: * Excelsior Disposable Syringe w/!Normal Saline (0.9% Sodium Chloride), K962938 (Excelsior Medical Corporation) * Excelsior Saline pre-filled Syringe(s) in Sterile Field Packaging, K082837 (Excelsior… | 1996 |
| * | K082837 | Sterile Field Saline Pre-Filled Syringe (S)Summary page 1…e Syringe w/!Normal Saline (0.9% Sodium Chloride), K962938 (Excelsior Medical Corporation) * Excelsior Saline pre-filled Syringe(s) in Sterile Field Packaging, K082837 (Excelsior Medical Corporation) * SwabCapt K083508 (Excelsior Medical Corporation) Description of the Device SwabFlush… | 2008 |
| * | K083508 | SwabcapSummary page 1…938 (Excelsior Medical Corporation) * Excelsior Saline pre-filled Syringe(s) in Sterile Field Packaging, K082837 (Excelsior Medical Corporation) * SwabCapt K083508 (Excelsior Medical Corporation) Description of the Device SwabFlush T1l is a saline intravenous flush syringe that is combined w… | 2009 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 8 years and passes through 3 companies; a new generation was cleared every 1.8 years on median.
- 4 of the 9 ancestors have a recall of any class.
- 1 other device also cites its predicate K962938.
How does its category compare?
All NGT devicesNGT Saline, Vascular Access Flush: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · NGT
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · NGT
142in 2023
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · NGT
1in 2025
Class I (6 since 2007) Class II/III not classified
Adverse event reports per year · NGT
212in 2025
MAUDE reports naming product code NGT, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 2 | 31 |
| 2008 | 1 | — | 0 | 64 |
| 2009 | 1 | — | 1 | 18 |
| 2010 | 2 | — | 7 | 23 |
| 2011 | 3 | 219 | 1 | 20 |
| 2012 | 4 | 106 | 1 | 5 |
| 2013 | 1 | — | 1 (1) | 25 |
| 2014 | 3 | 87 | 0 | 25 |
| 2015 | 1 | — | 0 | 14 |
| 2016 | 1 | — | 1 (1) | 82 |
| 2017 | 2 | — | 0 | 56 |
| 2018 | 1 | — | 0 | 270 |
| 2019 | 2 | — | 2 | 340 |
| 2020 | 2 | — | 0 | 322 |
| 2021 | 0 | — | 1 (1) | 566 |
| 2022 | 3 | 151 | 1 | 282 |
| 2023 | 6 | 142 | 3 (2) | 313 |
| 2024 | 1 | — | 0 | 332 |
| 2025 | 1 | — | 1 | 212 |
| 2026 | 0 | — | 1 (1) | — |
What is it indicated for?
The flush syringe is intended for flushing of IV catheters and IV tubing.
SwabCap® is intended for use on a swab-able luer access valve as a disinfecting cleaner prior to line access and to act as a physical barrier to contamination between the line accesses. SwabCap® will disinfect the valve five (5) minutes after application and act as a physical barrier to contamination for up to ninety-six (96) hours under normal conditions if not removed.
The SwabFlush product is provided as:
1. 10 ml fill in 10 ml saline flush syringe with integrated SwabCap®.
2. 5 ml fill in 10 ml saline flush syringe with integrated SwabCap®.
3. 3 ml fill in 10 ml saline flush syringe with integrated SwabCap®.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K101270 · printed Nov 6, 2026
Cite this page
510(k) Report. “K101270: Swabflsuh (3ML, 5ML, 10ML) Model 10331, 10351, 10311.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101270
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026