# K101347: Magnetom Aera; Magnetom Skyra

- Pathway: 510(k) (Traditional)
- Applicant: Siemens Medical Solutions USA, Inc.
- Decided: Oct 1, 2010 after 141 days
- Product code: LNH (System, Nuclear Magnetic Resonance Imaging)
- Device class: 2
- Regulation: 21 CFR 892.1000
- Recalls of this device: 7 (2 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 7 generations back to K971684 (1997). 11 ancestors in all.
224 later devices descend from it; 4 cite it directly.
22 of 236 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K041112: Magnetom Espree System, Siemens Medical Solutions USA, Inc. (2004) Summary page 2.
- K072237: Magnetom Verio, Siemens Medical Solutions USA, Inc. (2007) Summary page 2.
- K082427: Syngo MR B17, Siemens Medical Solutions USA, Inc. (2008) Summary page 2.

## Most influential descendants
- K111242: Magnetom Aera and Magnetom Skyra (2011), 220 descendants of its own
- K121434: Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/Verio (2012), 215 descendants of its own
- K141977: Magnetom Aera with Software Syngo MR E11A, Magnetom Skyra with Software Syngo MR E11A, Magnetom Skyra with 24 RF Channel (2014), 115 descendants of its own
- K151579: MAGNETOM Aera (24-channel), MAGNETOM Avanto fit, MAGNETOM Skyra fit, MAGNETOM Prisma, MAGNETOM Prisma fit (2015), 85 descendants of its own
- K153343: MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma, MAGNETOM Prisma fit (2016), 84 descendants of its own

## Recalls of this device
- Aug 28, 2025 · Class I · Z-2660-2025: There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, l
- Aug 28, 2025 · Class I · Z-2655-2025: There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, l
- Oct 17, 2016 · Class II · Z-1770-2017: Due to a potential manufacturing error, a rare possibility exists for a sudden drop in table height of the Combi Dockable Table Neurosurgery.
- Apr 13, 2016 · Class II · Z-1598-2016: The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
- Sep 12, 2012 · Class II · Z-2464-2012: When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with resp
- Jan 26, 2012 · Not classified · Z-0940-2012: During product monitoring firm has discovered that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, maki
- Jun 20, 2011 · Not classified · Z-2979-2011: Firm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making s

## Indications for use
The MAGNETOM systems described above are indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM systems described above may also be used for imaging during interventional procedures when performed with MR compatible devices such as inroom display and MR-safe biopsy needles.

Source page: https://510kreport.com/device/K101347