510(k) Report

510(k) · Cleared

Cleared

Pearlescent Powder Free Nitrile Examination Glove with Aloe Vera (Yellow, Red, Green, Black, Blue)

K101436510(k) TraditionalCleared Nov 17, 2010177-day reviewClass I21 CFR 880.6250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shen Wei (Usa), Inc. · LZC Medical Glove, Specialty

Cleared on Nov 17, 2010 as substantially equivalent to K092411 (Shen Wei (Usa), Inc., 2009) and 1 other predicate after a 177-day review. Its family tree reaches back 1 generation to 2003.

Decided
Nov 17, 2010
Review
177 days
Class
Class I
Product code
LZC
Recalls
None
Predicates
2

Its longest line of descent runs back 1 generation to K032024 (2003). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2009: 1 other ancestorK032024 · Powder-Free Nitrile Examination Gloves, Blue Color (Tested for Use with Chemotherapy Drugs) · Shen Wei (Usa), Inc. · 20032000200220042006200820102012
Other ancestors, by generation 1 generation back: 1 other ancestor K032024 · Powder-Free Nitrile Examination Gloves, Blue Color (Tested for Use with Chemotherapy Drugs) · Shen Wei (Usa), Inc. · 20030
  1. Root · 1 generation backK032024 Powder-Free Nitrile Examination Gloves, Blue Color (Tested for Use…Shen Wei (Usa), Inc. · 2003
  2. This deviceK101436 Pearlescent Powder Free Nitrile Examination Glove with Aloe Vera…Shen Wei (Usa), Inc. · 2010
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K032024Powder-Free Nitrile Examination Gloves, Blue Color (Tested for Use with…Shen Wei (Usa), Inc.2003—
0 (this device)K101436Pearlescent Powder Free Nitrile Examination Glove with Aloe Vera (Yellow, Red…Shen Wei (Usa), Inc.2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200910
FIG. 1 — Lineage of K101436, 2003–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZC device

  1. #1 K201390 Medline Powder-Free Light Blue Nitrile Exam…score 19.2 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K092411Pearlescent Powder Free Nitrile Examination Gloves with Aloe Vera (Yellow, Red, Green…
Summary page 1
…scent® Pigment, Powder Free Nitrile Examination Gloves with Aloe Vera Regulatory Class I Nitrile patient examination Product code: 8OLZA 5 10(k): K092411 4.0 Description of the Device: Blue and Red with Pearlescent® Pigment, Powder Free Nitrite Examination Gloves with Aloe…
Shen Wei (Usa), Inc.2009—
*K032024Powder-Free Nitrile Examination Gloves, Blue Color (Tested for Use with Chemotherapy…
Summary page 3
…ssion: Characteristic and Shen Wei USA Shen Wei USA Substantial parameters INC. (New Device) INC. K032024 Equivalence (SE) Product Code LZA LZA Indented Use…
Shen Wei (Usa), Inc.2003—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.

How does its category compare?

All LZC devices

LZC Medical Glove, Specialty: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · LZC

1in 2025

2007: 12008: 32009: 52010: 52012: 42015: 12016: 32017: 62018: 62019: 72020: 32021: 32022: 72025: 12026 (partial year): 1
2007high 7 (2019) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZC

85in 2022

2008: 166 days median2009: 253 days median2010: 68 days median2012: 149 days median2016: 258 days median2017: 121 days median2018: 105 days median2019: 200 days median2020: 167 days median2021: 94 days median2022: 85 days median
2007high 258 (2016) · low 68 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZC

1in 2021

2008: 12011: 42021: 12026 (partial year): 1
2007high 4 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZC

10in 2025

2009: 12011: 12012: 12013: 12014: 32016: 52017: 72018: 62019: 82020: 12021: 22022: 22023: 12024: 162025: 10
2007high 16 (2024) · low 0 (2007)2025

MAUDE reports naming product code LZC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
2008316610
2009525301
201056800
20110—41
2012414901
20130—01
20140—03
20151—00
2016325805
2017612107
2018610506
2019720008
2020316701
202139412
202278502
20230—01
20240—016
20251—010
20261—1—

What is it indicated for?

A disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only. Tested for use with chemotherapy drugs. Tested chemotherapy drugs are as follows (Fluorouracil, Etoposide, Cyclophosphamide, Carmustine, Thio-Tepa, Paclitaxel, Doxorubicin Hydrochloride, Dacarbazine, Cisplatin) Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes. The…

Read the full statement (260 words)

A disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only. Tested for use with chemotherapy drugs. Tested chemotherapy drugs are as follows (Fluorouracil, Etoposide, Cyclophosphamide, Carmustine, Thio-Tepa, Paclitaxel, Doxorubicin Hydrochloride, Dacarbazine, Cisplatin)

Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes.

The following chemicals have been tested with these gloves.

Cyclophosphamide >240

Doxorubicin HCl (Adriamycin) >240

*Carmustine 2.10

Etoposide >240

Fluorouracil (Adrucil) >240

Paclitaxel (Taxol) >240

*ThioTEPA 27.17

Dacarbazine >240

Cisplatin >240

Please note that the following drugs have extremely low permeation times of less than 30 minutes, Carmustine has an average breakthrough time of 2.10 minutes. Thiotepa has an average breakthrough time of 27.17 minutes.

A disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only. Tested for use with chemotherapy drugs. Tested chemotherapy drugs are as follows (Fluorouracil, Etoposide, Cyclophospharnide, Carmustine, Thio-Tepa, Paclitaxel, Doxorubicin Hydrochloride, Dacarbazine, Cisplatin)

Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes.

The following chemicals have been tested with these gloves.

Cyclophosphamide >240

Doxorubicin HC1 (Adriamycin) >240

*Carmustine 0.68

Etoposide >240

Fluorouracil (Adrucil) >240

Paclitaxel (Taxol) >240

*ThioTEPA 34.21

Dacarbazine >240

Cisplatin >240

Please note that the following drug have extremely low permeation times of less than 30 minutes, Carmustine has an average breakthrough time of 0.68 minutes.

Please note that the following drug have extremely low permeation times of less than 60 minutes, Thiotepa has an average breakthrough time of 34.21 minutes.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101436: Pearlescent Powder Free Nitrile Examination Glove with Aloe Vera (Yellow, Red, Green, Black, Blue).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101436

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026