# K101534: Conmed Altrus Thermal Tissue Fusion System

- Pathway: 510(k) (Traditional)
- Applicant: Conmedcorp
- Decided: Nov 24, 2010 after 174 days
- Product code: GEI (Electrosurgical, Cutting & Coagulation & Accessories)
- Device class: 2
- Regulation: 21 CFR 878.4400
- Recalls of this device: 8 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 2 generations back to K932293 (1994). 8 ancestors in all.
1 later device descends from it; 1 cites it directly.
0 of 10 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K981916: Ligasure Vessel Sealing System, Valleylab, Inc. (1998) Summary page 1.
- K002906: Ultracision Harmonic Scalpel System, Models GEN01, GEN02, GEN03, HP052, HP053, HSA07, Ethicon Endo-Surgery, Inc. (2000) Summary page 1.

## Most influential descendants
- K140459: Altrus Thermal Tissue Fusion Energy Source, Altrus Thermal Tissue Fusion Handpiece, Altrus Thermal Tissue Fusion System (2014), 0 descendants of its own

## Recalls of this device
- Feb 20, 2014 · Class II · Z-1839-2014: ConMed Corporation initiated an Urgent Medical Device Correction for Altrus Thermal Tissue Fusion Handpieces due to risk of burns.
- Oct 23, 2013 · Class II · Z-0332-2014: The affected lots were exposed to a higher than acceptable level of particulate matter during manufacture and subsequently distributed. ConMed evaluated the potential risk that this issue could presen
- Oct 23, 2013 · Class II · Z-0331-2014: The affected lots were exposed to a higher than acceptable level of particulate matter during manufacture and subsequently distributed. ConMed evaluated the potential risk that this issue could presen
- May 7, 2012 · Not classified · Z-1678-2012: An investigation of the ConMed Electrosurgery Altrus Energy Source revealed that the device could have been calibrated incorrectly resulting in error messages or delays in surgery.
- May 7, 2012 · Not classified · Z-1681-2012: The devices were distributed prior to the firm completing performance qualifications for their package sealing and adhesive bonding.
- May 7, 2012 · Not classified · Z-1680-2012: The devices were distributed prior to the firm completing performance qualifications for their package sealing and adhesive bonding.
- May 7, 2012 · Not classified · Z-1682-2012: The devices were distributed prior to the firm completing performance qualifications for their package sealing and adhesive bonding.
- May 7, 2012 · Not classified · Z-1679-2012: The devices were distributed prior to the firm completing performance qualifications for their package sealing and adhesive bonding.

## Indications for use
The CONMED ALTRUS® Thermal Tissue Fusion System is indicated for open and laparoscopic techniques in general surgical and gynecological procedures for ligating (sealing) and dividing (cutting) of tissue when hemostasis is desired.