510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Immulite 2000 3GALLERGY Specific Ige Assay Kit

K101572510(k) TraditionalCleared Aug 1, 2011423-day reviewClass II21 CFR 866.5750Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Healthcare Diagnostics · DHB System, Test, Radioallergosorbent (Rast) Immunological

Cleared on Aug 1, 2011 as substantially equivalent to K013134 (Diagnostic Products Corp., 2001) and 3 other predicates after a 423-day review. Its family tree reaches back 2 generations to 1991. 2 later devices descend from it. It has been recalled 4 times.

Decided
Aug 1, 2011
Review
423 days
Class
Class II
Product code
DHB
Recalls
4
Predicates
4

Its longest line of descent runs back 2 generations to K911511 (1991). 5 ancestors in all. 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 2 other ancestors2001: 1 other ancestor2012: 1 descendant cleared2014: 1 descendant clearedK911511 · Alastat Allergen Specified Ige System, Modified · Diagnostic Products Corp. · 1991K013134 · Immulite 2000 Allergen-Specific Ige and Specific Allergens, Mdoel L2KUN6 (600 Tests) · Diagnostic Products Corp. · 2001 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K013134 · Immulite 2000 Allergen-Specific Ige and Specific Allergens, Mdoel L2KUN6 (600 Tests) · Diagnostic Products Corp. · 2001 · recalledK911511 · Alastat Allergen Specified Ige System, Modified · Diagnostic Products Corp. · 19910
  1. Root · 2 generations backK911511 Alastat Allergen Specified Ige System, ModifiedDiagnostic Products Corp. · 1991
  2. 1 generation backK013134 Immulite 2000 Allergen-Specific Ige and Specific Allergens, Mdoel…Diagnostic Products Corp. · 2001 · Recalled
  3. This deviceK101572 Immulite 2000 3GALLERGY Specific Ige Assay KitSiemens Healthcare Diagnostics · 2011
  4. Descendant · cites it directly · 1 of its ownK112523 Immulite 2000 3GALLERGY Specific Ige Assay KitSiemens Healthcare Diagnostics, Inc. · 2012
  5. Descendant · 2 generations on · 0 of its ownK132801 Immulite 2000 Systems 3GALLERGY Specific Ige AssaySiemens Healthcare Diagnostics, Inc. · 2014

Most influential descendants

Showing 2 of 2 · See all
  • K112523Immulite 2000 3GALLERGY Specific Ige Assay KitSiemens Healthcare Diagnostics, Inc. · 2012 · cites it directly1descendant
  • K132801Immulite 2000 Systems 3GALLERGY Specific Ige AssaySiemens Healthcare Diagnostics, Inc. · 2014 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K911511Alastat Allergen Specified Ige System, ModifiedDiagnostic Products Corp.1991—
−1K013134Immulite 2000 Allergen-Specific Ige and Specific Allergens, Mdoel L2KUN6 (600…Diagnostic Products Corp.20012
0 (this device)K101572Immulite 2000 3GALLERGY Specific Ige Assay KitSiemens Healthcare Diagnostics2011—
+1K112523Immulite 2000 3GALLERGY Specific Ige Assay KitSiemens Healthcare Diagnostics, Inc.2012—
+2K132801Immulite 2000 Systems 3GALLERGY Specific Ige AssaySiemens Healthcare Diagnostics, Inc.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200220

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201410
FIG. 1 — Lineage of K101572, 1991–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DHB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DHB device

  1. #1 K260331 IMMULITE 2000 3gAllergy Specific IgE Vantage…score 12.3 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K013134Immulite 2000 Allergen-Specific Ige and Specific Allergens, Mdoel L2KUN6 (600 Tests)
Summary page 2
…illa 4 F253 Pine Nut 5 F269 Basil 6 F283 Oregano FDA clearance was previously obtained for the assay kit and 196 specific allergens and allergen panels (K013134, K021206, K013135 and K021208). Please note that the FDA clearances indicated above were in the name of Diagnostic Products Corpo…
Diagnostic Products Corp.20012
*K021206Immulite 2000 Specific Allergens
Summary page 2
…F253 Pine Nut 5 F269 Basil 6 F283 Oregano FDA clearance was previously obtained for the assay kit and 196 specific allergens and allergen panels (K013134, K021206, K013135 and K021208). Please note that the FDA clearances indicated above were in the name of Diagnostic Products Corporation wh…
Diagnostic Products Corp.20021
*K013135Immunlite 2000 Allergen-Specific Ige and Mixed Allergen Panels, Model L2KUN6 (600 Tests)
Summary page 2
…ine Nut 5 F269 Basil 6 F283 Oregano FDA clearance was previously obtained for the assay kit and 196 specific allergens and allergen panels (K013134, K021206, K013135 and K021208). Please note that the FDA clearances indicated above were in the name of Diagnostic Products Corporation which was a…
Diagnostic Products Corp.2001—
*K021208Immulite 2000 Mixed Allergen Panels
Summary page 2
…F269 Basil 6 F283 Oregano FDA clearance was previously obtained for the assay kit and 196 specific allergens and allergen panels (K013134, K021206, K013135 and K021208). Please note that the FDA clearances indicated above were in the name of Diagnostic Products Corporation which was acquired by S…
Diagnostic Products Corp.20021

Also cleared against K013134: 7 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K112523Immulite 2000 3GALLERGY Specific Ige Assay KitSiemens Healthcare Diagnostics, Inc.20121

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 6, 2025Class IIZ-0492-2026Under Investigation by firmThe barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument…
Nov 4, 2022Class IIZ-0718-2023Under Investigation by firmPotential for falsely elevated specific IgE mold allergen reactivity with quality control material and patient samples, may result in false reactive result
Nov 4, 2022Class IIZ-0716-2023Under Investigation by firmPotential for falsely elevated specific IgE mold allergen reactivity with quality control material and patient samples, may result in false reactive result
Nov 4, 2022Class IIZ-0717-2023Under Investigation by firmPotential for falsely elevated specific IgE mold allergen reactivity with quality control material and patient samples, may result in false reactive result

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 2 companies; a new generation was cleared every 10.5 years on median.
  • 3 of the 5 ancestors have a recall of any class.
  • Of its descendants, K112523 (Siemens Healthcare Diagnostics, Inc., 2012) has gone furthest, with 1 descendant of its own.
  • 7 other devices also cite its predicate K013134.

How does its category compare?

All DHB devices

DHB System, Test, Radioallergosorbent (Rast) Immunological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · DHB

2in 2025

2007: 12008: 22009: 72010: 32011: 52012: 32013: 52014: 12015: 12016: 22017: 12018: 22019: 32020: 22021: 12022: 22025: 22026 (partial year): 6
2007high 7 (2009) · low 0 (2023)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DHB

90in 2019

2009: 118 days median2010: 247 days median2011: 405 days median2012: 260 days median2013: 257 days median2019: 90 days median2026 (partial year): 138 days median
2007high 405 (2011) · low 90 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DHB

14in 2025

2012: 12014: 12017: 12018: 12022: 32025: 14
2007high 14 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DHB

6in 2025

2008: 12011: 12012: 22015: 12017: 112018: 62019: 42021: 72022: 12023: 62024: 102025: 6
2007high 11 (2017) · low 0 (2007)2025

MAUDE reports naming product code DHB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20082—01
2009711800
2010324700
2011540501
2012326012
2013525700
20141—10
20151—01
20162—00
20171—111
20182—16
201939004
20202—00
20211—07
20222—31
20230—06
20240—010
20252—146
202661380—

What is it indicated for?

For in vitro diagnostic use with the IMMULITE 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101572: Immulite 2000 3GALLERGY Specific Ige Assay Kit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K101572

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026