# K101788: Giraffe Omnibed

- Pathway: 510(k) (Special)
- Applicant: Ohmeda Medical
- Decided: Jul 21, 2010 after 26 days
- Product code: FMZ (Incubator, Neonatal)
- Device class: 2
- Regulation: 21 CFR 880.5400
- Recalls of this device: 8 (3 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
4 later devices descend from it; 1 cites it directly.
3 of 5 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. The FDA summary names its predicate without a number.
- (no number) Giraffe OmniBed: named in the summary without a submission number

## Most influential descendants
- K152814: Giraffe OmniBed Carestation CS1 (2016), 3 descendants of its own
- K213553: Giraffe Incubator Carestation CS1 (2022), 2 descendants of its own
- K251663: Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1) (2025), 1 descendant of its own
- K260326: Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1) (2026), 0 descendants of its own

## Recalls of this device
- May 1, 2026 · Class II · Z-2475-2026: GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more
- May 1, 2026 · Class II · Z-2477-2026: GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more
- Sep 3, 2024 · Class I · Z-0047-2025: GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to spe
- May 19, 2023 · Class II · Z-2101-2023: GE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniB
- Apr 28, 2020 · Class II · Z-2235-2020: Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the
- Apr 2, 2020 · Class II · Z-2111-2020: During Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.
- Oct 11, 2019 · Class I · Z-0538-2020: Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
- Jan 8, 2013 · Class I · Z-1051-2013: Neonatal beds and incubator display boards may randomly overwrite stored information and radiant heater and oxygen set points upon power up or recovery from a power failure.

## Indications for use
The OmnniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. it cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. This device may incorporate a Serv o Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).