510(k) Report

510(k) · Cleared

Cleared

Powder Free Nitrle Examination Gloves with Chemotherapy Labeling Claim

K102189510(k) TraditionalCleared Mar 14, 2012590-day reviewClass I21 CFR 880.6250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Latexx Manufacturing Sdn.Bhd. · LZC Medical Glove, Specialty

Cleared on Mar 14, 2012 as substantially equivalent to K032444 (Cardinalhealth, 2003) and 1 other predicate after a 590-day review. Its family tree reaches back 1 generation to 2002.

Decided
Mar 14, 2012
Review
590 days
Class
Class I
Product code
LZC
Recalls
None
Predicates
2

Its longest line of descent runs back 1 generation to K022765 (2002). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2003: 1 other ancestorK022765 · Nitrile Powder-Free Examination Gloves · Allegiance Healthcare Corp. · 200220002002200420062008201020122014
Other ancestors, by generation 1 generation back: 1 other ancestor K022765 · Nitrile Powder-Free Examination Gloves · Allegiance Healthcare Corp. · 20020
  1. Root · 1 generation backK022765 Nitrile Powder-Free Examination GlovesAllegiance Healthcare Corp. · 2002
  2. This deviceK102189 Powder Free Nitrle Examination Gloves with Chemotherapy Labeling ClaimLatexx Manufacturing Sdn.Bhd. · 2012
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K022765Nitrile Powder-Free Examination GlovesAllegiance Healthcare Corp.2002—
0 (this device)K102189Powder Free Nitrle Examination Gloves with Chemotherapy Labeling ClaimLatexx Manufacturing Sdn.Bhd.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200310
FIG. 1 — Lineage of K102189, 2002–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LZC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZC device

  1. #1 K201390 Medline Powder-Free Light Blue Nitrile Exam…score 19.2 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K032444Nitrile Powder-Free Examination Gloves with Lotion Coating and Use for Chemotherapy
Summary page 1
…ch Equivalence is being Claimed: Primary Predicate: Cardinal Health Esteem Tru-Blu Stretchy Nitrite Exam Gloves 510(k) Number: K032444 Additional Predicate: Powder-Free Blue Nitrite Examination Gloves Tested for Use with Chemotherapy…
Cardinalhealth2003—
*K022765Nitrile Powder-Free Examination Gloves
Summary page 1
…nal Predicate: Powder-Free Blue Nitrite Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim 510(k) Number: K022765 4.0 Description of the Device: Powder-Free Nitrite Examination Gloves Tested for Use with Chemotherapy…
Allegiance Healthcare Corp.2002—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.

How does its category compare?

All LZC devices

LZC Medical Glove, Specialty: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · LZC

1in 2025

2007: 12008: 32009: 52010: 52012: 42015: 12016: 32017: 62018: 62019: 72020: 32021: 32022: 72025: 12026 (partial year): 1
2007high 7 (2019) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZC

85in 2022

2008: 166 days median2009: 253 days median2010: 68 days median2012: 149 days median2016: 258 days median2017: 121 days median2018: 105 days median2019: 200 days median2020: 167 days median2021: 94 days median2022: 85 days median
2007high 258 (2016) · low 68 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZC

1in 2021

2008: 12011: 42021: 12026 (partial year): 1
2007high 4 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LZC

10in 2025

2009: 12011: 12012: 12013: 12014: 32016: 52017: 72018: 62019: 82020: 12021: 22022: 22023: 12024: 162025: 10
2007high 16 (2024) · low 0 (2007)2025

MAUDE reports naming product code LZC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
2008316610
2009525301
201056800
20110—41
2012414901
20130—01
20140—03
20151—00
2016325805
2017612107
2018610506
2019720008
2020316701
202139412
202278502
20230—01
20240—016
20251—010
20261—1—

What is it indicated for?

The Powder Freea Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiners hand or finger to prevent contamination between examiner and patient In addition these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use with chemotherapy drugs. Tested chemotherapy drugs are as…

Read the full statement (188 words)

The Powder Freea Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiners hand or finger to prevent contamination between examiner and patient In addition these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use with chemotherapy drugs. Tested chemotherapy drugs are as follows:

Test Chemical and Concentration Average Breakthrough Detection Time (minutes)

Cisplatin, 1.0 mg/ml : 240

Cyclophosphamide (Cytoxan), 20.0 mg/mI : 240

Cytarabine, 100 mg/rn : 240

Cytovene, 10 mg/mI 240

Dacarbazine (OTIC), 10.0 mg/mI t 240

Docetaxel, 10.0 mg/mI , 4

Doxorubicin Hydrochloride, 2.0 mg/mI : 240

Fluorouracil, 50.0 mg/l t 240

Ifosfamide, 50.0 mg/mI : 240

Mechlorethamine HCI, 1.0 mglml 240

Methotrexate, 25 mg/ml 2t 240

Mitomycin C, 0.5 mg/mI 240

Mitoxantrone, 2 mg/mI 240

Oxaliplatin, Smq/mI 240

Paclitaxel (Taxol), 6.0 mg/mI 240

Vincristine Sulfate, 1.0 mg/mI 240

Vinorelbine, 10 mg/mi :240

Low Permeation Time (minutes) 1.7

Carmustine (BCNU), 3.3 mg/mI ___________________

Please note that the following drug has an extremely low permeation time:

Carmustine (8GNU) 1.7(mins)

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102189: Powder Free Nitrle Examination Gloves with Chemotherapy Labeling Claim.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102189

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026