510(k) Report

510(k) · Cleared

Cleared

Obs Disposable Electrosurgical Pads, Monopolar, Biopolar, Models GBS-DM, GBS-DB

K102372510(k) TraditionalCleared Nov 3, 201075-day reviewClass II21 CFR 878.4400

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Jiangmen City Xinhui Baisheng Medical Equipment CO · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Nov 3, 2010 as substantially equivalent to K030362 (Leonhard Lang GmbH, 2003) and 2 other predicates after a 75-day review. Its family tree reaches back 3 generations to 1985. 2 later devices descend from it.

Decided
Nov 3, 2010
Review
75 days
Class
Class II
Product code
GEI
Recalls
None
Predicates
3

Its longest line of descent runs back 3 generations to K853291 (1985). 6 ancestors in all. 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1986: 1 other ancestor1997: 1 other ancestor2013: 1 descendant cleared2022: 1 descendant clearedK853291 · GP-100 Adult Electrosurgical Ground Pad/200 & 300 · Bio-Detek, Inc. · 1985K993306 · Nikopad Electrosurgical Grounding Pad (*or Sold Under Commercial Names), Model 4760 (Typically) · Paladin Medical, Inc. · 1999K030362 · Skintact Cool Contact Electrosurgical Grounding Plates · Leonhard Lang GmbH · 20031990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant K030362 · Skintact Cool Contact Electrosurgical Grounding Plates · Leonhard Lang GmbH · 2003K993306 · Nikopad Electrosurgical Grounding Pad (*or Sold Under Commercial Names), Model 4760 (Typically) · Paladin Medical, Inc. · 1999K853291 · GP-100 Adult Electrosurgical Ground Pad/200 & 300 · Bio-Detek, Inc. · 19850
  1. Root · 3 generations backK853291 GP-100 Adult Electrosurgical Ground Pad/200 & 300Bio-Detek, Inc. · 1985
  2. 2 generations backK993306 Nikopad Electrosurgical Grounding Pad (*or Sold Under Commercial…Paladin Medical, Inc. · 1999
  3. 1 generation backK030362 Skintact Cool Contact Electrosurgical Grounding PlatesLeonhard Lang GmbH · 2003
  4. This deviceK102372 Obs Disposable Electrosurgical Pads, Monopolar, Biopolar, Models…Jiangmen City Xinhui Baisheng Medical Equipment CO · 2010
  5. Descendant · cites it directly · 1 of its ownK130027 Disposable Neutral Electrode GP202Cathay Manufacturing Corp. · 2013
  6. Descendant · 2 generations on · 0 of its ownK203293 Abbott Medical Grounding Pad, model RF-DGP-ISAbbott Medical · 2022

Most influential descendants

Showing 2 of 2 · See all
  • K130027Disposable Neutral Electrode GP202Cathay Manufacturing Corp. · 2013 · cites it directly1descendant
  • K203293Abbott Medical Grounding Pad, model RF-DGP-ISAbbott Medical · 2022 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K853291GP-100 Adult Electrosurgical Ground Pad/200 & 300Bio-Detek, Inc.1985—
−2K993306Nikopad Electrosurgical Grounding Pad (*or Sold Under Commercial Names), Model…Paladin Medical, Inc.1999—
−1K030362Skintact Cool Contact Electrosurgical Grounding PlatesLeonhard Lang GmbH2003—
0 (this device)K102372Obs Disposable Electrosurgical Pads, Monopolar, Biopolar, Models GBS-DM, GBS-DBJiangmen City Xinhui Baisheng Medical Equipment CO2010—
+1K130027Disposable Neutral Electrode GP202Cathay Manufacturing Corp.2013—
+2K203293Abbott Medical Grounding Pad, model RF-DGP-ISAbbott Medical2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199710
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
202210
FIG. 1 — Lineage of K102372, 1985–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030362Skintact Cool Contact Electrosurgical Grounding Plates
Summary page 1
…inbH Address: Archenweg 56 6010 Innsbruck Austria Proprietary Name: Skintact® Cool Contact Electrosurgical Grounding Plates It's 510(K) number is K030362 Code: FDA-005 IVersion: NO0 Page 1 of 4 Issuing Date: 2009-12-28…
Leonhard Lang GmbH2003—
*K002002Surefit Dispersive Electrode
Summary page 2
…: ConMed Corporation Address: 310 Broad Street Utica, New York 13501 Proprietary Name: SureFiP' Dual Dispersive Electrode (410-2000) It's 510(K) number is K002002 3) Valleylab Company Name: Valleylab,INC. Address: 5910 Longbow Drive Boulder CO 80301 Proprietary Name: PolyHes…
Conmedcorp2000—
*K861036E7509 Polyhesive II Patient Return Electrode
Summary page 2
…ectrode(Mode E7506:solid,with cable.Model E7507:REM,with cable. Model E7509: REM,without cable) It's 510(K) number is K861036 4.0 Device Description OBS Disposable Electrosurgical Pads are disposable, non-sterile dispersive electrode with or without…
Valleylab, Inc.1986—

Also cleared against K030362: 3 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K130027Disposable Neutral Electrode GP202Cathay Manufacturing Corp.20131

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 4 companies; a new generation was cleared every 7.6 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K130027 (Cathay Manufacturing Corp., 2013) has gone furthest, with 1 descendant of its own.
  • 3 other devices also cite its predicate K030362.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

OHS Disposable Electrosurgical Pads are Neutral Electrode which is also known as plate, plate electrode, passive, return, or dispersive electrode. They are indicated for use to adhere to the patient over the entire pad surface to complete an electrical circuit during electrosurgery between the electrosurgical generator, the active electrode and the patient. Solid Electrosurgical Pads are for use with generators that do not have a Contact Quality Monitoring System(CQMS). Split Electrosurgical Pads are for use with generators that have a CQMS(i.e.REM,ARM,NESSY etc.).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102372: Obs Disposable Electrosurgical Pads, Monopolar, Biopolar, Models GBS-DM, GBS-DB.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K102372

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026