510(k) Report

510(k) · Cleared

Recalled · Class I · 2025

Bravo PH Monitoring System and Accessories

K102543510(k) TraditionalCleared Dec 1, 201089-day reviewClass I21 CFR 876.1400Recalls 4 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Given Imaging, Ltd. · FFT Electrode, Ph, Stomach

Cleared on Dec 1, 2010 as substantially equivalent to K002028 (Endonetics, Inc., 2000) after an 89-day review. Its family tree reaches back 1 generation to 2000. 1 of the 2 devices in its family, counting itself, has a Class I recall. It has been recalled 4 times, including a Class I recall.

Decided
Dec 1, 2010
Review
89 days
Class
Class I
Product code
FFT
Recalls
4 (1 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K002028 (2000). No later device cites it as a predicate yet. 1 of the 2 devices in this family, counting this one, has a Class I recall.

K002028 · Bravo PH Monitoring System · Endonetics, Inc. · 2000 · recalled2000200220042006200820102012
  1. Root · 1 generation backK002028 Bravo PH Monitoring SystemEndonetics, Inc. · 2000 · Recalled
  2. This deviceK102543 Bravo PH Monitoring System and AccessoriesGiven Imaging, Ltd. · 2010
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K002028Bravo PH Monitoring SystemEndonetics, Inc.20004
0 (this device)K102543Bravo PH Monitoring System and AccessoriesGiven Imaging, Ltd.2010—
FIG. 1 — Lineage of K102543, 2000–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002028Bravo PH Monitoring System
Summary page 1
…Panel: Gastroenterology FFT - Stomach pH electrode Predicate Device(s): Bravo pH- Monitoring SystemuTm and Accessories (K002028) General Device The Bravo pH Monitoring SysteMm and Accessories is an Description: ambulator…
Endonetics, Inc.20004

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 3, 2025Class IZ-2047-2025Process controlIt has been determined that misapplied adhesive on the Bravo CF capsule delivery device may lead to its malfunction. Specifically, the misapplied adhesive may prevent the capsule from attaching to the patient's esophagus or detaching from…
Feb 26, 2021Class IIZ-1334-2021Process change controlThe capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following…
Feb 26, 2021Class IIZ-1335-2021Process change controlThe capsule may fail to attach to the esophageal mucosa, which in turn may lead to the aspiration of the capsule. In the event of capsule aspiration, immediate intervention to remove the capsule is required. Potential outcomes following…
Jan 11, 2012Not classifiedZ-0966-2012Nonconforming Material/ComponentFailure of the Bravo capsule to attach to the esophagus or alternatively, failure of the capsule to detach from the placement device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.

How does its category compare?

All FFT devices

FFT Electrode, Ph, Stomach: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · FFT

1in 2010

2010: 1
2007high 1 (2010) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FFT

Too few FFT 510(k)s in any year since 2007 to give a median.

Recalls per year · FFT

1in 2007

2007: 12026 (partial year): 1, 1 Class I
2007high 1 (2007) · low 0 (2008)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · FFT

590in 2025

2007: 4992008: 8842009: 3362010: 9282011: 1,5062012: 1,7072013: 3142014: 3062015: 3472016: 4722017: 8962018: 9552019: 6212020: 7702021: 5162022: 2602023: 2592024: 2392025: 590
2007high 1,707 (2012) · low 239 (2024)2025

MAUDE reports naming product code FFT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—1499
20080—0884
20090—0336
20101—0928
20110—01,506
20120—01,707
20130—0314
20140—0306
20150—0347
20160—0472
20170—0896
20180—0955
20190—0621
20200—0770
20210—0516
20220—0260
20230—0259
20240—0239
20250—0590
20260—1 (1)—

What is it indicated for?

The Bravo pH Monitoring System is intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children from 4 years of age. The Bravo pH Capsule can be attached following either endoscopy or manometry. The RAPID pH software is intended to record, store, view, and analyze gastroesophageal pH data.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102543: Bravo PH Monitoring System and Accessories.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102543

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026