510(k) Report

510(k) · Cleared

Cleared

Gyrus Acmi Cyber Wand-Sterilization Trays Model FGT

K102970510(k) TraditionalCleared May 20, 2011226-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Gyrus Acmi, Inc. · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on May 20, 2011 as substantially equivalent to K092682 (Gyrus Acmi, Inc., 2009) and 1 other predicate after a 226-day review. Its family tree reaches back 6 generations to 1984.

Decided
May 20, 2011
Review
226 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
2

Its longest line of descent runs back 6 generations to K844002 (1984). 11 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 6 generations backK844002 HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc. · 1984
  2. 5 generations backK980065 C/T Med-Systems Cassette SystemC/T Med System, Ltd. · 1998
  3. 4 generations backK990761 Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd. · 1999
  4. 3 generations backK012105 Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc. · 2002
  5. 2 generations backK033222 Olympus Sterilization TraysOlympus Winter & Ibe GmbH · 2004
  6. 1 generation backK092682 Gyrus Acmi Flexible Endoscope Storage-Sterilization TraysGyrus Acmi, Inc. · 2009
  7. This deviceK102970 Gyrus Acmi Cyber Wand-Sterilization Trays Model FGTGyrus Acmi, Inc. · 2011
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K844002HU-Friedy Ims Cassette SystemHu-Friedy Mfg. Co., Inc.1984—
−5K980065C/T Med-Systems Cassette SystemC/T Med System, Ltd.1998—
−4K990761Tuttnauer Mini Container Cassette, Tuttnauer Standard Case CassetteTuttnauer U.S.A Co, Ltd.1999—
−3K012105Polyvac Surgical Instrument Delivery SystemSymmetry Medical, Inc.2002—
−2K033222Olympus Sterilization TraysOlympus Winter & Ibe GmbH2004—
−1K092682Gyrus Acmi Flexible Endoscope Storage-Sterilization TraysGyrus Acmi, Inc.2009—
0 (this device)K102970Gyrus Acmi Cyber Wand-Sterilization Trays Model FGTGyrus Acmi, Inc.2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199710
200110
200210
200410
FIG. 1 — Lineage of K102970, 1984–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K092682Gyrus Acmi Flexible Endoscope Storage-Sterilization Trays
Summary page 1
…and Transducer and Accessories Sterilization Trays Generic/Common Name: Sterilization trays Predicate Devices Gyros ACMI Flexible Endoscope Storage Sterilization Trays K092682 ¢ Symmetry Medical PolyVac Instrument Delivery System K040223 Intended Uses The Gyrus ACMI® CyberWand™ Transducer and Accessorie…
Gyrus Acmi, Inc.2009—
*K040223Polyvac Instrument Delivery System
Summary page 1
…mmon Name: Sterilization trays Predicate Devices Gyros ACMI Flexible Endoscope Storage Sterilization Trays K092682 ¢ Symmetry Medical PolyVac Instrument Delivery System K040223 Intended Uses The Gyrus ACMI® CyberWand™ Transducer and Accessories Storage-Sterilization Trays are intended to be used to enclo…
Symmetry Medical USA, Inc.2004—

Also cleared against K092682: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 6 companies; a new generation was cleared every 3.1 years on median.
  • None of the 11 ancestors has a recall.
  • 3 other devices also cite its predicate K092682.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The Gyrus ACMI CyberWand™ Sterilization Trays are intended to be used to enclose and protect only CyberWand™ transducers and accessories during sterilization. The trays are to be used in conjunction with an FDA cleared sterilization wrap. The trays are optional accessories to the CyberWand™ transducers and accessories for which they are designed. The trays are indicated for ETO sterilization of only the following (maximum load) per tray: 2 CyberWand…

Read the full statement (112 words)

The Gyrus ACMI CyberWand™ Sterilization Trays are intended to be used to enclose and protect only CyberWand™ transducers and accessories during sterilization. The trays are to be used in conjunction with an FDA cleared sterilization wrap. The trays are optional accessories to the CyberWand™ transducers and accessories for which they are designed.

The trays are indicated for ETO sterilization of only the following (maximum load) per tray: 2 CyberWand Transducers, 2 CyberWand Probe sets, 1 Wrench, 1 Cleaning stylet.

EtO Sterilization parameters: EtO Sterilize using 100% ethylene oxide: Temp 55 °C, Vacuum 97mmHgV, Preconditioning time 60 minutes, EtO concentration 735-750 mg/L, Exposure time 60 minutes, Humidity 35%-80%, Aeration 12 hours @55 °C.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K102970: Gyrus Acmi Cyber Wand-Sterilization Trays Model FGT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102970

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026