510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Lifepak 12

K102972510(k) SpecialCleared Dec 22, 201077-day reviewClass III21 CFR 870.5310Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Physio-Control, Inc. · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Dec 22, 2010 as substantially equivalent to K063510 (Medtronic Emergency Response Systems, Inc., 2007) and 8 other predicates after a 77-day review. Its family tree reaches back 9 generations to 1993. It has been recalled once.

Decided
Dec 22, 2010
Review
77 days
Class
Class III
Product code
MKJ
Recalls
1
Predicates
9

Its longest line of descent runs back 9 generations to K921669 (1993). 30 ancestors in all, the oldest K802288 (1980). No later device cites it as a predicate yet.

  1. Root · 9 generations backK921669 Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc. · 1993
  2. 5 generations between
  3. 3 generations backK040775 Lifepak 12Medtronic Physio-Control Corp. · 2004 · Recalled
  4. 2 generations backK041459 Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2004
  5. 1 generation backK063510 Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2007 · Recalled
  6. This deviceK102972 Lifepak 12Physio-Control, Inc. · 2010
Show all 9 generations
  1. Root · 9 generations backK921669 Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc. · 1993
  2. 8 generations backK964750 Eagle 4000 Patient MonitorMarquette Electronics, Inc. · 1997
  3. 7 generations backK973486 Life Pak 12Physio-Control Corp. · 1998 · Recalled
  4. 6 generations backK991910 Lifepak 12 BiphasicPhysio-Control Corp. · 1999 · Recalled
  5. 5 generations backK012274 Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp. · 2002 · Recalled
  6. 4 generations backK033275 Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp. · 2003 · Recalled
  7. 3 generations backK040775 Lifepak 12Medtronic Physio-Control Corp. · 2004 · Recalled
  8. 2 generations backK041459 Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2004
  9. 1 generation backK063510 Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2007 · Recalled
  10. This deviceK102972 Lifepak 12Physio-Control, Inc. · 2010
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K921669Marquette SL Series Transport Remote AcquisitionMarquette Electronics, Inc.1993—
−8K964750Eagle 4000 Patient MonitorMarquette Electronics, Inc.1997—
−7K973486Life Pak 12Physio-Control Corp.19983
−6K991910Lifepak 12 BiphasicPhysio-Control Corp.19993
−5K012274Medtronic Physio-Control Lifepak 20 Defibrillator/Monitor/PacemakerMedtronic Physio-Control Corp.200217
−4K033275Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp.20031
−3K040775Lifepak 12Medtronic Physio-Control Corp.20041
−2K041459Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc.2004—
−1K063510Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc.20072
0 (this device)K102972Lifepak 12Physio-Control, Inc.2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198410
199010
199110
199220
199410
199540
199620
199820
199920
200140
FIG. 1 — Lineage of K102972, 1993–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MKJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (9)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K063510Lifepak 12 Defibrillator/Monitor
Summary page 2
…r/monitor. 51 0(k) numbers for the predicate LIFEPAK 12 defibrillator/monitor are listed below. 510(k) Number Clearance Date K063510 01/26/07 K041459 07/01/04 K040775 04/23/04…
Medtronic Emergency Response Systems, Inc.20072
*K041459Lifepak 12 Defibrillator/Monitor
Summary page 2
…LIFEPAK 12 defibrillator/monitor are listed below. 510(k) Number Clearance Date K063510 01/26/07 K041459 07/01/04 K040775 04/23/04 K033275 11/06/03…
Medtronic Emergency Response Systems, Inc.2004—
*K040775Lifepak 12
Summary page 2
…. 510(k) Number Clearance Date K063510 01/26/07 K041459 07/01/04 K040775 04/23/04 K033275 11/06/03 K010918 03/26/0…
Medtronic Physio-Control Corp.20041
*K033275Lifepak 12, 20, 500, CR Plus
Summary page 2
…Date K063510 01/26/07 K041459 07/01/04 K040775 04/23/04 K033275 11/06/03 K010918 03/26/01 K002445 01/31/0…
Medtronic Physio-Control Corp.20031
*K010918Lifepak 12 Defibrillator/Monitor System
Summary page 2
…/07 K041459 07/01/04 K040775 04/23/04 K033275 11/06/03 K010918 03/26/01 K002445 01/31/01 K990338 09/01/9…
Medtronic Physio-Control Corp.20014
*K002445Lifepak 12DEFIBERLLATOR/Monitor Series
Summary page 2
…04 K040775 04/23/04 K033275 11/06/03 K010918 03/26/01 K002445 01/31/01 K990338 09/01/99 K(991910 06/03/9…
Medtronic Physio-Control Corp.20011
*K990338Nibp and CO2 Options for the Physio-Control Lifepak 12 Defibrillator/Monitor System
Summary page 2
…04 K033275 11/06/03 K010918 03/26/01 K002445 01/31/01 K990338 09/01/99 K(991910 06/03/99 K973486 01/09/9…
Physio-Control Corp.19991
*K991910Lifepak 12 Biphasic
Summary page 2
…K010918 03/26/01 K002445 01/31/01 K990338 09/01/99 K(991910 06/03/99 K973486 01/09/98 Description: The LIFEPAK 12 defibrillator/mo…
Physio-Control Corp.19993
*K973486Life Pak 12
Summary page 2
…01 K002445 01/31/01 K990338 09/01/99 K(991910 06/03/99 K973486 01/09/98 Description: The LIFEPAK 12 defibrillator/monitor was designed for use in.a variety of h…
Physio-Control Corp.19983

Also cleared against K063510: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 16, 2017Class IIZ-1143-2017Process controlThe End-Tidal CO2 (EtCO2) reading can intermittently show a value of XXX after start-up or during device operation.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 5 companies; a new generation was cleared every 1.7 years on median.
  • 14 of the 30 ancestors have a recall of any class.
  • 1 other device also cites its predicate K063510.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

Manual Defibrillation: Indications Defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Energy delivered in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia. Contraindications Defibrillation is contraindicated…

Read the full statement (385 words)

Manual Defibrillation:

Indications

Defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Energy delivered in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.

Contraindications

Defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA), such as indioventricular or ventricular escape rhythms, and in the treatment of asystole.

Automated External Defibrillation:

Indications

AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the patient's ECG rhythm. In AED mode, the LIFEPAK 12 defibrillator/monitor is not intended for use on pediatric patients less than 8 years old.

Noninvasive Pacing

Indications

Noninvasive pacing is indicated for symptomatic bradycardia in patients with a pulse.

Contraindications: Noninvasive pacing is contraindicated for the treatment of ventricular fibrillation and asystole.

12-lead Electrocardiography:

Indications

The 12-lead electrocardiogram is used to identify, diagnose and treat patients with cardiac disorders and is useful in the early detection and prompt treatment of patients with acute myocardial infarction.

Pulse Oximetry

Indications

Pulse Oximetry is indicated for use in any patient who is at risk of developing hypoxemia.

Noninvasive Blood Pressure Monitoring:

Indications

Noninvasive blood pressure monitoring is indicated for detection in trends of hypertension or hypotension. These include patient conditions indicated by abnormalities in various physiologic parameters such as shock, evaluation of perfusion during dysrhythmias, major fluid shifts, evaluation of response to fluid therapy, and titration of vasoactive and cardiotonic drugs. Noninvasive blood pressure monitoring may be useful during ECG monitoring or for post-defibrillation recovery analysis.

End-Tidal CO2 monitoring:

Indications

EtCO2 monitoring is indicated for detection of trends in the level of expired CO2. It is used for monitoring breathing efficacy and treatment effectiveness in acute cardiopulmonary care, for example, to determine if adequate compressions are being performed during CPR or to rapidly detect whether an endotracheal tube has been placed successfully. It is intended for use on adult and pediatric patients.

Invasive Pressure Monitoring:

Indications

The LIFEPAK 12 invasive pressure monitor is indicated for use in measuring arterial, venous, intracranial and other physiological pressures using and invasive catheter system with a compatible transducer. It may be used on the adult or pediatric patient.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102972: Lifepak 12.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K102972

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026