510(k) Report

510(k) · Cleared

Cleared

AG-608N Blood Glucose Monitoring System

K110017510(k) TraditionalCleared Jan 18, 2012380-day reviewClass II21 CFR 862.1345

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Andon Medical Co., Ltd. · NBW System, Test, Blood Glucose, Over The Counter

Cleared on Jan 18, 2012 as substantially equivalent to K093262 (Andon Medical Co., Ltd., 2010) after a 380-day review. Its family tree reaches back 3 generations to 2003. 10 later devices descend from it.

Decided
Jan 18, 2012
Review
380 days
Class
Class II
Product code
NBW
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K024194 (2003). 10 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2012: 1 descendant cleared2013: 1 descendant cleared2014: 2 descendants cleared2017: 2 descendants cleared2016: 2 descendants cleared2015: 1 descendant cleared2019: 1 descendant clearedK024194 · One Touch Ultra and One Touch Induo Blood Glucose Monitoring Systems · Lifescan, Inc. · 2003 · recalledK073030 · AG-605 Blood Glucose Monitoring System · Andon Health Co, Ltd. · 2008K093262 · AG-608 Blood Glucose Monitoring System and AG-610 Blood Glucose Monitoring System · Andon Medical Co., Ltd. · 2010200020052010201520202025
Descendants, by generation 1 generation forward: 4 descendants2 generations forward: 2 descendants3 generations forward: 3 descendants4 generations forward: 1 descendant K093262 · AG-608 Blood Glucose Monitoring System and AG-610 Blood Glucose Monitoring System · Andon Medical Co., Ltd. · 2010K073030 · AG-605 Blood Glucose Monitoring System · Andon Health Co, Ltd. · 2008K024194 · One Touch Ultra and One Touch Induo Blood Glucose Monitoring Systems · Lifescan, Inc. · 2003 · recalled0
  1. Root · 3 generations backK024194 One Touch Ultra and One Touch Induo Blood Glucose Monitoring SystemsLifescan, Inc. · 2003 · Recalled
  2. 2 generations backK073030 AG-605 Blood Glucose Monitoring SystemAndon Health Co, Ltd. · 2008
  3. 1 generation backK093262 AG-608 Blood Glucose Monitoring System and AG-610 Blood Glucose…Andon Medical Co., Ltd. · 2010
  4. This deviceK110017 AG-608N Blood Glucose Monitoring SystemAndon Medical Co., Ltd. · 2012
  5. Descendant · cites it directly · 7 of its ownK120813 Blood Glucose Monitoring SystemAndon Medical Co., Ltd. · 2012
  6. Descendant · cites it directly · 3 of its ownK123935 Ihealth BG5/BG5L Wireless Smart Glucose Monitoring SystemAndon Medical Co., Ltd. · 2013
  7. Descendant · 2 generations on · 2 of its ownK153278 Wireless Smart Gluco-Monitoring SystemAndon Health Co, Ltd. · 2016
  8. Descendant · 2 generations on · 2 of its ownK133790 Ihealth Align Mini Gluco-Monitoring SystemAndon Health Co, Ltd. · 2014
  9. Descendant · 3 generations on · 1 of its ownK170231 iHealth Align Gluco-Monitoring System (BG1); iHealth Wireless Smart…Andon Health Co, Ltd. · 2017

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2012: 12013: 12014: 22015: 12016: 22017: 22019: 1
2012high 2 (2014) · low 0 (2018)2026

Most influential descendants

Showing 5 of 10 · See all
  • K120813Blood Glucose Monitoring SystemAndon Medical Co., Ltd. · 2012 · cites it directly7descendants
  • K123935Ihealth BG5/BG5L Wireless Smart Glucose Monitoring SystemAndon Medical Co., Ltd. · 2013 · cites it directly3descendants
  • K153278Wireless Smart Gluco-Monitoring SystemAndon Health Co, Ltd. · 2016 · 2 generations on2descendants
  • K133790Ihealth Align Mini Gluco-Monitoring SystemAndon Health Co, Ltd. · 2014 · 2 generations on2descendants
  • K170231iHealth Align Gluco-Monitoring System (BG1); iHealth Wireless Smart Gluco-Monitoring…Andon Health Co, Ltd. · 2017 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K024194One Touch Ultra and One Touch Induo Blood Glucose Monitoring SystemsLifescan, Inc.20035
−2K073030AG-605 Blood Glucose Monitoring SystemAndon Health Co, Ltd.2008—
−1K093262AG-608 Blood Glucose Monitoring System and AG-610 Blood Glucose Monitoring…Andon Medical Co., Ltd.2010—
0 (this device)K110017AG-608N Blood Glucose Monitoring SystemAndon Medical Co., Ltd.2012—
+1K120813Blood Glucose Monitoring SystemAndon Medical Co., Ltd.2012—
+1K123935Ihealth BG5/BG5L Wireless Smart Glucose Monitoring SystemAndon Medical Co., Ltd.2013—
+2K153278Wireless Smart Gluco-Monitoring SystemAndon Health Co, Ltd.2016—
+2K133790Ihealth Align Mini Gluco-Monitoring SystemAndon Health Co, Ltd.2014—
+3K170231iHealth Align Gluco-Monitoring System (BG1); iHealth Wireless Smart…Andon Health Co, Ltd.2017—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201310
201420
201510
201620
201720
201910
FIG. 1 — Lineage of K110017, 2003–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NBW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NBW device

  1. #1 K260545 CONTOUR® PLUS BLUE Blood Glucose Monitoring…score 14.6 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K093262AG-608 Blood Glucose Monitoring System and AG-610 Blood Glucose Monitoring System
Summary page 3
…MULTI Blood Glucose Monitoring System Monitoring System (K093262) Detection Method Amperometry Amperometry Enzyme…
Andon Medical Co., Ltd.2010—

What descends from it? (10)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K120813Blood Glucose Monitoring SystemAndon Medical Co., Ltd.20127
K123935Ihealth BG5/BG5L Wireless Smart Glucose Monitoring SystemAndon Medical Co., Ltd.20133
K153561AG-607 Blood Glucose Monitoring System and AG-607 Multi Blood Glucose Monitoring SystemAndon Health Co, Ltd.20170
K132180Ideal Life Gluco-Manager Blood Glucose Monitoring SystemIdeal Life, Inc.20140

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 3 companies; a new generation was cleared every 2.0 years on median.
  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K120813 (Andon Medical Co., Ltd., 2012) has gone furthest, with 7 descendants of its own.

How does its category compare?

All NBW devices

NBW System, Test, Blood Glucose, Over The Counter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · NBW

2in 2025

2007: 292008: 372009: 382010: 312011: 302012: 302013: 472014: 242015: 322016: 202017: 252018: 192019: 102020: 72021: 72022: 122023: 52024: 72025: 22026 (partial year): 5
2007high 47 (2013) · low 2 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NBW

243in 2024

2007: 85 days median2008: 128 days median2009: 140 days median2010: 132 days median2011: 328 days median2012: 230 days median2013: 167 days median2014: 202 days median2015: 237 days median2016: 233 days median2017: 244 days median2018: 168 days median2019: 127 days median2020: 121 days median2021: 465 days median2022: 410 days median2023: 193 days median2024: 243 days median2026 (partial year): 190 days median
2007high 465 (2021) · low 85 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NBW

1in 2025

2007: 92008: 22009: 52010: 112011: 12012: 12013: 5, 2 Class I2014: 12016: 12019: 142025: 1
2007high 14 (2019) · low 0 (2015)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · NBW

4,745in 2025

2007: 7,9912008: 7,7042009: 10,9632010: 11,6312011: 22,9092012: 23,4492013: 35,4082014: 38,5242015: 62,1112016: 76,6212017: 37,5382018: 16,9112019: 14,4722020: 10,7032021: 9,1982022: 8,1032023: 7,2562024: 6,2202025: 4,745
2007high 76,621 (2016) · low 4,745 (2025)2025

MAUDE reports naming product code NBW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007298597,991
20083712827,704
200938140510,963
2010311321111,631
201130328122,909
201230230123,449
2013471675 (2)35,408
201424202138,524
201532237062,111
201620233176,621
201725244037,538
201819168016,911
2019101271414,472
20207121010,703
2021746509,198
20221241008,103
2023519307,256
2024724306,220
20252—14,745
202651900—

What is it indicated for?

The AG-608N Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh. The AG-608N single Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The AG-608N single Blood Glucose Monitoring System is intended for self testing outside the body (in…

Read the full statement (432 words)

The AG-608N Single Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh. The AG-608N single Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The AG-608N single Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The AG-608N single Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-iQOOON single Test Strips are for use with the AG-608N single Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AG-608N control solutions are intended for use with the AG-608N single Blood Glucose Monitoring System, to check that the glucose meter and test strips are working properly and that you are doing the test correctly. These solutions contain a known range of glucose, as indicated on the bottles.

The AG-608N MULTI Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh. The AG-608N MULTI Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. The system is only used with single-use lancing devices.

The AG-608N MULTI Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, upper arm, calf and thigh should be done only during steady - state times (when glucose is not changing rapidly).

The AGS-100ON MULTI Test Strips are for use with the AG-608N MULTI Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf and thigh.

The AG-608N control solutions are intended for use with the AG-608N MULTI Blood Glucose Monitoring System, to check that the glucose meter and test strips are working properly. These solutions contain a known range of glucose, as indicated on the bottles.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K110017: AG-608N Blood Glucose Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K110017

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026