# K110277: Rapidpoint 405 Neonatal Bilirubin

- Pathway: 510(k) (Traditional)
- Applicant: Siemens Healthcare Diagnostics
- Decided: Aug 15, 2011 after 196 days
- Product code: MQM (Bilirubin (Total And Unbound) In The Neonate Test System)
- Device class: 1
- Regulation: 21 CFR 862.1113
- Recalls of this device: 5 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 7 generations back to K861087 (1986). 22 ancestors in all.
No later device cites it as a predicate yet.
0 of 23 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K073537: Neonatal Bilirubin on Rapidlab Models 1245 and 1265, Siemens Medical Solutions Diagnostics (2008) Summary page 1.
- (no number) RAPIDLab 1245: named in the summary without a submission number
- (no number) RAPIDLab 1265: named in the summary without a submission number

## Recalls of this device
- Sep 2, 2016 · Class III · Z-2913-2016: Error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge
- Aug 11, 2016 · Class II · Z-2798-2016: There is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system
- Jul 28, 2015 · Class II · Z-2479-2015: When both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port causing the data management system to 
- Dec 30, 2014 · Class II · Z-1252-2015: Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 2
- Sep 17, 2012 · Class II · Z-0103-2014: RAPIDPoint 400 Series, RAPIDPoint 500, or RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misiden

## Indications for use
The RAPIDPoint 405 system is intended for in vitro diagnostic use and is designed to provide the determination in whole blood for the following parameters: * Partial pressure of carbon dioxide * Partial pressure of oxygen * pH * Sodium * Potassium * Ionized calcium * Chloride * Glucose * Total hemoglobin and fractions: fO2Hb, fCOHb, fMetHb, fNHb * Neonatal bilirubin This test system is intended for use in point of care or laboratory settings. PCO2, PO2, pH. Measurements of blood gases (pCO2, pO2) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances. Neonate Bilirubin. A bilirubin (total and unbound) in the neonate test system is a device intended to measure the levels of bilirubin (total and unbound) in the blood (serum) of newborn infants to aid in indicating the risk of bilirubin encephalopathy (kernicterus). Sodium. Sodium measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Potassium. Potassium measurements obtained by this device are used to monitor electrolyte balance in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels. Chloride. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. Ionized calcium. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). Glucose. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. Total hemoglobin. Total hemoglobin measurements are used to determine the hemoglobin content of human blood. Oxyhemoglobin. Oxyhemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia. Carboxyhemoglobin. Carboxyhemoglobin measurements are used to determine the carboxyhemoglobin (the compound formed when hemoglobin is exposed to carbon monoxide) content of human blood as an aid in the diagnosis of carbon monoxide poisoning. Sulfhemoglobin. Sulfhemoglobin measurements are used to determine the sulfhemoglobin (a compound of sulfur and hemoglobin) content of human blood as an aid in the diagnosis of sulfhemoglobinemia (presence of sulfhemoglobin in the blood due to drug administration or exposure to a poison).