510(k) · Cleared
Recalled · Class I · 2019Babi*Plus Pressure Relief Manifold - 10CM H20, 12.5CM H20, 15CM H20
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
A Plus Medical · BZD Ventilator, Non-Continuous (Respirator)
Cleared on Aug 1, 2011 as substantially equivalent to K093716 (A Plus Medical, 2010) and 1 other predicate after a 172-day review. Its family tree reaches back 2 generations to 1983. 1 of the 5 devices in its family, counting itself, has a Class I recall. It has been recalled 4 times, including 4 Class I recalls.
- Decided
- Aug 1, 2011
- Review
- 172 days
- Class
- Class II
- Product code
- BZD
- Recalls
- 4 (4 Class I)
- Predicates
- 2
Its longest line of descent runs back 2 generations to K831246 (1983). 4 ancestors in all. No later device cites it as a predicate yet. 1 of the 5 devices in this family, counting this one, has a Class I recall.
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K831246 | Whistler in-Line Pressure Release Valve | Airlife, Inc. | 1983 | — |
| −1 | K040366 | Pressure Manifold, Model BC 110 | Fisher & Paykel Healthcare, Ltd. | 2004 | — |
| 0 (this device) | K110383 | Babi*Plus Pressure Relief Manifold - 10CM H20, 12.5CM H20, 15CM H20 | A Plus Medical | 2011 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1987 | 1 | 0 |
| 2010 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new BZD device
- #1 K250624 myAirscore 15.7 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K093716 | Babi Plus Infant Nasal Cannula SystemSummary page 1…nuous (respirator) Product Code: EZO Regulation: 868.5905 Device Class: 11 Device: UoblAPus Pressure Limitation Manifold Predicate Devices: K Number: K093716 Product Name: Infant Nasal Cannula System Manufacturer: A Plus Medical Product Codes: 20216 & 20217 K Number:… | 2010 |
| * | K040366 | Pressure Manifold, Model BC 110Summary page 1…vices: K Number: K093716 Product Name: Infant Nasal Cannula System Manufacturer: A Plus Medical Product Codes: 20216 & 20217 K Number: K040366 Product Name: BC I 10 Pressure Manifold Manufacturer: Fisher & Paykel Healthcare Product Codes: BC I 10 & RC… | 2004 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 14, 2019 | Class I | Z-0009-2020 | Device Design | Device vented gas below the stated pressure. |
| May 30, 2019 | Class I | Z-0021-2020 | Device Design | Reports of pressure relief manifold venting gas below the standard pressure. |
| May 30, 2019 | Class I | Z-0020-2020 | Device Design | Reports of pressure relief manifold venting gas below the standard pressure. |
| May 30, 2019 | Class I | Z-0022-2020 | Device Design | Reports of pressure relief manifold venting gas below the standard pressure. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 28 years and passes through 3 companies; a new generation was cleared every 20.6 years on median.
- 1 of the 4 ancestors has a recall of any class.
- 1 other device also cites its predicate K093716.
How does its category compare?
All BZD devicesBZD Ventilator, Non-Continuous (Respirator): FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · BZD
7in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · BZD
189in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · BZD
8in 2023
Class I (13 since 2007) Class II/III not classified
Adverse event reports per year · BZD
60,279in 2025
MAUDE reports naming product code BZD, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 21 | 83 | 29 | 65 |
| 2008 | 17 | 108 | 2 | 67 |
| 2009 | 23 | 90 | 1 | 147 |
| 2010 | 12 | 122 | 0 | 143 |
| 2011 | 23 | 125 | 1 | 189 |
| 2012 | 21 | 133 | 0 | 139 |
| 2013 | 23 | 127 | 0 | 82 |
| 2014 | 14 | 172 | 0 | 86 |
| 2015 | 11 | 224 | 0 | 114 |
| 2016 | 12 | 217 | 0 | 118 |
| 2017 | 5 | 221 | 0 | 148 |
| 2018 | 5 | 280 | 0 | 101 |
| 2019 | 6 | 231 | 4 (4) | 188 |
| 2020 | 6 | 137 | 0 | 126 |
| 2021 | 6 | 192 | 0 | 6,066 |
| 2022 | 4 | 253 | 0 | 90,768 |
| 2023 | 5 | 204 | 8 (7) | 27,987 |
| 2024 | 10 | 132 | 0 | 68,262 |
| 2025 | 7 | 189 | 0 | 60,279 |
| 2026 | 10 | 214 | 2 (2) | — |
What is it indicated for?
Single patient use device intended for neonate, infant or child with a body mass of less than 10 Kg requiring a pressure limitation system to eliminate excessive pressure should an obstruction occur between gas supply and exhalation port during continuous gas flow therapy up to 12 liters per minute in hospital critical care unit.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K110383 · printed Nov 6, 2026
Cite this page
510(k) Report. “K110383: Babi*Plus Pressure Relief Manifold - 10CM H20, 12.5CM H20, 15CM H20.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K110383
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026