Aspira Peritoneal Drainage System

K110396510(k) SpecialCleared May 6, 201184-day reviewClass II21 CFR 876.5630Recalls 17

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

C.R. Bard, Inc. · PNG Peritoneal, Drainage Catheter For Refractory Ascites, Long-Term Indwelling

Cleared on May 6, 2011 as substantially equivalent to K081288 (C.R. Bard, Inc., 2008) after an 84-day review. Its family tree reaches back 1 generation to 2008. 10 later devices descend from it. It has been recalled 17 times.

Decided
May 6, 2011
Review
84 days
Class
Class II
Product code
PNG
Recalls
17
Predicates
1

Its longest line of descent runs back 1 generation to K081288 (2008). 10 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2012: 3 descendants cleared2021: 1 descendant cleared2016: 2 descendants cleared2017: 1 descendant cleared2020: 1 descendant cleared2015: 1 descendant cleared2024: 1 descendant clearedK081288 · Aspira Peritoneal Drainage System · C.R. Bard, Inc. · 200820052010201520202025
Descendants, by generation 1 generation forward: 4 descendants2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant K081288 · Aspira Peritoneal Drainage System · C.R. Bard, Inc. · 20080
  1. Root · 1 generation backK081288 Aspira Peritoneal Drainage SystemC.R. Bard, Inc. · 2008
  2. This deviceK110396 Aspira Peritoneal Drainage SystemC.R. Bard, Inc. · 2011
  3. Descendant · cites it directly · 8 of its ownK113854 Pleurx Peritoneal Catheter SystemCare Fusion · 2012
  4. Descendant · cites it directly · 4 of its ownK122422 Pleurx Pleural Catheter System Pleurx Catheter Access Kit Pleurx…Care Fusion · 2012
  5. Descendant · cites it directly · 2 of its ownK160437 PleurX Peritoneal Catheter SystemCare Fusion · 2016
  6. Descendant · 2 generations on · 2 of its ownK121849 Pleurx Pleural Catheter System, Pleurx Catheter Access Kit, Pleurx…Care Fusion · 2012
  7. Descendant · 2 generations on · 1 of its ownK201155 PleurX Peritoneal Catheter SystemCare Fusion · 2020

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2012: 32015: 12016: 22017: 12020: 12021: 12024: 1
2012high 3 (2012) · low 0 (2013)2026

Most influential descendants

Showing 5 of 10 · See all
  • K113854Pleurx Peritoneal Catheter SystemCare Fusion · 2012 · cites it directly8descendants
  • K122422Pleurx Pleural Catheter System Pleurx Catheter Access Kit Pleurx Slide Valved Peelable…Care Fusion · 2012 · cites it directly4descendants
  • K160437PleurX Peritoneal Catheter SystemCare Fusion · 2016 · cites it directly2descendants
  • K121849Pleurx Pleural Catheter System, Pleurx Catheter Access Kit, Pleurx Catheter Insertion…Care Fusion · 2012 · 2 generations on2descendants
  • K201155PleurX Peritoneal Catheter SystemCare Fusion · 2020 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K081288Aspira Peritoneal Drainage SystemC.R. Bard, Inc.2008—
0 (this device)K110396Aspira Peritoneal Drainage SystemC.R. Bard, Inc.2011—
+1K113854Pleurx Peritoneal Catheter SystemCare Fusion2012—
+1K122422Pleurx Pleural Catheter System Pleurx Catheter Access Kit Pleurx Slide Valved…Care Fusion2012—
+1K160437PleurX Peritoneal Catheter SystemCare Fusion2016—
+2K121849Pleurx Pleural Catheter System, Pleurx Catheter Access Kit, Pleurx Catheter…Care Fusion2012—
+2K201155PleurX Peritoneal Catheter SystemCare Fusion2020—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201230
201510
201620
201710
202010
202110
202410
FIG. 1 — Lineage of K110396, 2008–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PNG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PNG device

  1. #1 K241946 PeritX™ 1L Drainage Kit (90-1010); PeritX™ 2L…score 10.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K081288Aspira Peritoneal Drainage System
Summary page 4
…ice ’ Classification Name: Peritoneal Dialysis System and Accessories 21 CFR 876.5630 - Class Il FJS - Peritoneal dialysis system and accessories Premarket Notification: K081288, concumrence date 18 July 2008 Manufacturer. - Bard Access Systems, Inc. Device The Aspira* Peritoneal Drainage System provides p…
C.R. Bard, Inc.2008—

What descends from it? (10)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K113854Pleurx Peritoneal Catheter SystemCare Fusion20128
K122422Pleurx Pleural Catheter System Pleurx Catheter Access Kit Pleurx Slide Valved Peelable…Care Fusion20124
K160437PleurX Peritoneal Catheter SystemCare Fusion20162
K212675Aspira Peritoneal Drainage SystemMerit Medical Systems, Inc.20210

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 22, 2023Class IIZ-1915-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1911-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1913-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1916-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1912-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1920-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1918-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1914-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1909-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1910-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1917-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1908-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
Apr 22, 2023Class IIZ-1919-2023Process controlThere is a potential that valve assemblies will not open. preventing fluid from draining.
May 21, 2011Not classifiedZ-2962-2011Error in labelingLabeling for drainage kit contained incomplete information.
May 21, 2011Not classifiedZ-2959-2011Error in labelingLabeling for drainage kit contained incomplete information.
May 21, 2011Not classifiedZ-2961-2011Error in labelingLabeling for drainage kit contained incomplete information.
May 21, 2011Not classifiedZ-2960-2011Error in labelingLabeling for drainage kit contained incomplete information.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K113854 (Care Fusion, 2012) has gone furthest, with 8 descendants of its own.

How does its category compare?

All PNG devices

PNG Peritoneal, Drainage Catheter For Refractory Ascites, Long-Term Indwelling: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · PNG

1in 2025

2008: 12010: 12011: 12012: 22016: 32020: 12021: 12023: 22024: 12025: 1
2007high 3 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PNG

244in 2016

2016: 244 days median
2007high 244 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PNG

No recall names PNG since 2007.

Adverse event reports per year · PNG

31in 2025

2017: 32018: 62019: 92020: 152021: 152022: 182023: 212024: 182025: 31
2007high 31 (2025) · low 0 (2007)2025

MAUDE reports naming product code PNG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—00
20090—00
20101—00
20111—00
20122—00
20130—00
20140—00
20150—00
2016324400
20170—03
20180—06
20190—09
20201—015
20211—015
20220—018
20232—021
20241—018
20251—031
20260—0—

What is it indicated for?

The Aspira® Peritoneal Drainage System is indicated for intermittent drainage of recurrent and symptomatic malignant ascites. The catheter is intended for long-term access of the peritoneal cavity in order to relieve symptoms such as dyspnea or other symptoms associated with malignant ascites. The Aspira® Drainage Bag is indicated for use only with the Aspira® Drainage Catheter for intermittent drainage. The Aspira® Dressing Kit is indicated for dressing…

Read the full statement (139 words)

The Aspira® Peritoneal Drainage System is indicated for intermittent drainage of recurrent and symptomatic malignant ascites. The catheter is intended for long-term access of the peritoneal cavity in order to relieve symptoms such as dyspnea or other symptoms associated with malignant ascites.

The Aspira® Drainage Bag is indicated for use only with the Aspira® Drainage Catheter for intermittent drainage.

The Aspira® Dressing Kit is indicated for dressing of a catheter and exit site.

The Aspira® Luer/Universal Adapter is intended to provide access to the Aspira® Drainage Catheter. It is used to drain fluid using standard wall suction, water seal drainage system, glass vacuum bottle, syringe or other appropriate method.

The Aspira® valve Assembly attaches to the Aspira® Drainage Catheter. The Aspira® Repair Kit is for the repair of the Aspira® Drainage Catheter and replacement of the Aspira® Valve Assembly.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K110396: Aspira Peritoneal Drainage System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K110396

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026