510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Randox Liquid Cardiac Control Levels 1, 2 and 3

K110534510(k) TraditionalCleared Jan 23, 2012332-day reviewClass I21 CFR 862.1660Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Randox Laboratories Limited · JJY Multi-Analyte Controls, All Kinds (Assayed)

Cleared on Jan 23, 2012 after a 332-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has been recalled 11 times.

Decided
Jan 23, 2012
Review
332 days
Class
Class I
Product code
JJY
Recalls
11
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K110534Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JJY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 4, 2019Class IIZ-0412-2020Under Investigation by firmRemoval of claims for N-proBNP in the Liquid Cardiac Control model numbers CQ5051 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.
Oct 4, 2019Class IIZ-0414-2020Under Investigation by firmRemoval of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5053 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.
Oct 4, 2019Class IIZ-0413-2020Under Investigation by firmRemoval of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5052 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.
Mar 15, 2019Class IIZ-1461-2020Device DesignThe form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house monitoring, determined that liquid Cardiac…
Mar 11, 2019Class IIZ-1785-2019Under Investigation by firmCustomers may observe a decrease in recovered concentration with regards to NTproBNP in the Liquid Cardiac Control lots identified, on the Siemens Dimension EXL LOCI.
Oct 2, 2018Class IIZ-0380-2019Under Investigation by firmRandox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of…
Oct 2, 2018Class IIZ-0381-2019Under Investigation by firmRandox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of…
Oct 2, 2018Class IIZ-0382-2019Under Investigation by firmRandox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of…
Jun 8, 2018Class IIZ-2575-2018Under Investigation by firmThe device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting…
Aug 8, 2017Class IIZ-0937-2018Unknown/Undetermined by firmRandox has confirmed that Troponin T and Troponin I in Liquid Cardiac quality control CQ5053 lot 4069CK does not meet the shelf-life claim in the product labelling. As the control may now fall outside the provided control ranges the…
Sep 16, 2016Class IIZ-0762-2017Nonconforming Material/ComponentAn internal real-time stability monitoring program revealed decreased recovery of myoglobin (-20% bias) for Randox Immunoturbidimetric and Siemens Centaur Classic methods and CKMB (-21% bias) for Roche e411, BioMerieux Vidas, and Siemens…

No ancestor or descendant has a Class I recall.

How does its category compare?

All JJY devices

JJY Multi-Analyte Controls, All Kinds (Assayed): FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · JJY

3in 2017

2007: 112008: 72009: 82010: 102011: 142012: 132013: 162014: 102015: 42016: 12017: 3
2007high 16 (2013) · low 0 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JJY

78in 2017

2007: 49 days median2008: 54 days median2009: 80 days median2010: 109 days median2011: 120 days median2012: 40 days median2013: 82 days median2014: 49 days median2015: 153 days median2017: 78 days median
2007high 153 (2015) · low 40 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JJY

2in 2025

2007: 52009: 22010: 62011: 12012: 32013: 22014: 22015: 12016: 32017: 42018: 32019: 42020: 62022: 32023: 32025: 22026 (partial year): 1
2007high 6 (2010) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JJY

13in 2025

2007: 22008: 12009: 32010: 32011: 62012: 202013: 52014: 42015: 42016: 132017: 122018: 72019: 12020: 72021: 12023: 42024: 12025: 13
2007high 20 (2012) · low 0 (2022)2025

MAUDE reports naming product code JJY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007114952
200875401
200988023
20101010963
20111412016
20121340320
2013168225
2014104924
2015415314
20161—313
2017378412
20180—37
20190—41
20200—67
20210—01
20220—30
20230—34
20240—01
20250—213
20260—1—

What is it indicated for?

The Randox Liquid Cardiac Controls Level 1, Level 2 and Level 3 are liquid controls containing BNP, OK MB Mass. Digoxin, D-Dimer, Homacysteine, hsCRP, Myoglobin, NT-ProBNP, Troponin I and Troponin T. They have been developed far use in the quality control of BNP, OK MB Mass, Digoxin, D-Dimer. Homocysteine, hsCRP, Myoglabin, NT-ProBNP, Troponin I and Troponin T assays on various clinical chemistry and immunoassay systems. This in vitro diagnostic device is intended for prescription use only

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K110534: Randox Liquid Cardiac Control Levels 1, 2 and 3.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K110534

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026