510(k) · Cleared
Recalled · Class II · 20153.7MM/5.0MM Dynamic Locking Screw
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Synthes USA · HWC Screw, Fixation, Bone
Cleared on Jul 25, 2011 as substantially equivalent to K101696 (Zimmer, Inc., 2010) and 4 other predicates after a 145-day review. Its family tree reaches back 4 generations to 1996. It has been recalled 2 times.
- Decided
- Jul 25, 2011
- Review
- 145 days
- Class
- Class II
- Product code
- HWC
- Recalls
- 2
- Predicates
- 5
Its longest line of descent runs back 4 generations to K961413 (1996). 13 ancestors in all. No later device cites it as a predicate yet.
- Root · 4 generations backK961413 Synthes Anatomical Locking Plate SystemSynthes (Usa) · 1996
- 3 generations backK042695 NCB Plating SystemZimmer GmbH · 2004 · Recalled
- 2 generations backK061211 NCB Plating System, Proximal Tibia PlatesZimmer GmbH · 2006 · Recalled
- 1 generation backK101696 Motionloc Screw for NCB Polyaxial Locking Plating SystemZimmer, Inc. · 2010
- This deviceK110592 3.7MM/5.0MM Dynamic Locking ScrewSynthes USA · 2011
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K961413 | Synthes Anatomical Locking Plate System | Synthes (Usa) | 1996 | — |
| −3 | K042695 | NCB Plating System | Zimmer GmbH | 2004 | 1 |
| −2 | K061211 | NCB Plating System, Proximal Tibia Plates | Zimmer GmbH | 2006 | 4 |
| −1 | K101696 | Motionloc Screw for NCB Polyaxial Locking Plating System | Zimmer, Inc. | 2010 | — |
| 0 (this device) | K110592 | 3.7MM/5.0MM Dynamic Locking Screw | Synthes USA | 2011 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1996 | 1 | 0 |
| 2000 | 3 | 0 |
| 2001 | 1 | 0 |
| 2003 | 2 | 0 |
| 2006 | 1 | 0 |
| 2008 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HWC device
- #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (5)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K101696 | Motionloc Screw for NCB Polyaxial Locking Plating SystemSummary page 1…ss 11 888.3030 - smooth or threaded metallic bone fixation fastener H WC Predicate Device: K 101696 - Zimmer Motion Loc Screw for NCB Polyaxial Locking Plate System… | 2010 |
| * | K961413 | Synthes Anatomical Locking Plate SystemSummary page 1…Predicate Device: K 101696 - Zimmer Motion Loc Screw for NCB Polyaxial Locking Plate System K961413 - Synthes Anatomical Locking Plate System K(963192 - Synthes 3.5/4.0mm Cannulated Screws… | 1996 |
| * | K963192 | Synthes Sterile 3.5 MM and 4.0 MM Cannulated ScrewsSummary page 1…Locking Plate System K961413 - Synthes Anatomical Locking Plate System K(963192 - Synthes 3.5/4.0mm Cannulated Screws K(000682 - Synthes Large Fragment Dynamic Compression Locki… | 1996 |
| * | K000682 | Synthes Large Fragment Dynamic Compression Locking (DCL) SystemSummary page 1…K961413 - Synthes Anatomical Locking Plate System K(963192 - Synthes 3.5/4.0mm Cannulated Screws K(000682 - Synthes Large Fragment Dynamic Compression Locking System K(000… | 2000 |
| * | K000684 | Small Fragment Dynamic Compression Locking (DCL) SystemSummary page 1…ws K(000682 - Synthes Large Fragment Dynamic Compression Locking System K(000684 - Synthes Small Fragment Dynamic Compression Locking System Device Description: The… | 2000 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 11, 2015 | Class II | Z-2794-2015 | Labeling False and Misleading | DLS pin breakage during planned implant removal, after uneventful and successful healing of the fracture. This notice to additional consignees identified after recall Z-0390-2014, related to complaints of breakage located at the bottom of… |
| Jul 22, 2013 | Class II | Z-0390-2014 | Labeling False and Misleading | All lots of 3.7mm and 5.0mm Dynamic Locking Screw (DLS) were recalled due to complaints of breakages at the distal tip of the pin identified after successful healing and during planned removal of the device. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 4.2 years on median.
- 5 of the 13 ancestors have a recall of any class.
- 4 other devices also cite its predicate K101696.
How does its category compare?
All HWC devicesHWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · HWC
15in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HWC
111in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HWC
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HWC
2,893in 2025
MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 35 | 77 | 25 | 669 |
| 2008 | 33 | 76 | 33 | 660 |
| 2009 | 27 | 113 | 29 | 651 |
| 2010 | 23 | 110 | 7 | 765 |
| 2011 | 35 | 88 | 12 | 1,104 |
| 2012 | 20 | 99 | 30 | 3,339 |
| 2013 | 25 | 90 | 14 | 3,850 |
| 2014 | 31 | 88 | 2 | 2,549 |
| 2015 | 41 | 85 | 5 | 2,711 |
| 2016 | 26 | 131 | 4 | 2,285 |
| 2017 | 40 | 127 | 4 | 2,561 |
| 2018 | 20 | 106 | 1 | 3,662 |
| 2019 | 30 | 108 | 5 | 5,463 |
| 2020 | 28 | 75 | 4 | 4,820 |
| 2021 | 26 | 84 | 2 | 4,446 |
| 2022 | 26 | 167 | 13 | 3,934 |
| 2023 | 24 | 104 | 0 | 3,718 |
| 2024 | 17 | 90 | 0 | 3,575 |
| 2025 | 15 | 111 | 1 | 2,893 |
| 2026 | 21 | 125 | 2 | — |
What is it indicated for?
Synthes Dynamic Locking Screws (DLS) in combination with Synthes Locking Compression Plates (LCP) are intended for use in long bone fractures, the fixation of osteopenic bone, the fixation of osteotomies, and for the fixation of non-unions and malunions.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K110592 · printed Nov 6, 2026
Cite this page
510(k) Report. “K110592: 3.7MM/5.0MM Dynamic Locking Screw.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K110592
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026