510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Safeport Manifold (TM) (or Stopcock)

K111016510(k) SpecialCleared Jun 9, 201158-day reviewClass II21 CFR 880.5440Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Elcam Medical Acal · FPA Set, Administration, Intravascular

Cleared on Jun 9, 2011 as substantially equivalent to K032393 (Elcam Medical Acal, 2003) after a 58-day review. Its family tree reaches back 3 generations to 1986. 4 later devices descend from it. It has been recalled 12 times.

Decided
Jun 9, 2011
Review
58 days
Class
Class II
Product code
FPA
Recalls
12
Predicates
1

Its longest line of descent runs back 3 generations to K862691 (1986). 4 ancestors in all. 4 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor2013: 1 descendant cleared2022: 1 descendant cleared2018: 1 descendant cleared2025: 1 descendant clearedK862691 · Stopcock · Ram Development · 1986K022895 · Elcam Stopcocks and Manifolds · Medical Device Consultants, Inc. · 2002 · recalledK032393 · Induction and Sampling Manifold (or Stopcock) · Elcam Medical Acal · 2003 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 1 descendant K032393 · Induction and Sampling Manifold (or Stopcock) · Elcam Medical Acal · 2003 · recalledK022895 · Elcam Stopcocks and Manifolds · Medical Device Consultants, Inc. · 2002 · recalledK862691 · Stopcock · Ram Development · 19860
  1. Root · 3 generations backK862691 StopcockRam Development · 1986
  2. 2 generations backK022895 Elcam Stopcocks and ManifoldsMedical Device Consultants, Inc. · 2002 · Recalled
  3. 1 generation backK032393 Induction and Sampling Manifold (or Stopcock)Elcam Medical Acal · 2003 · Recalled
  4. This deviceK111016 Safeport Manifold (TM) (or Stopcock)Elcam Medical Acal · 2011
  5. Descendant · cites it directly · 1 of its ownK123084 Disposable High Pressure Stopcock/Manifold Model HP Stopcock, MP…Elcam Medical Acal · 2013
  6. Descendant · 2 generations on · 0 of its ownK242255 Qitexio® 4-Way Stopcock (QIT014)Medex · 2025
  7. Descendant · cites it directly · 0 of its ownK210516 Clicky CrossYomura Technologies, Inc. · 2022
  8. Descendant · cites it directly · 0 of its ownK172266 Dynarex Three-Way StopcockDynarex Corporation · 2018

Most influential descendants

Showing 4 of 4 · See all
  • K123084Disposable High Pressure Stopcock/Manifold Model HP Stopcock, MP Stopcock, HP Manifold…Elcam Medical Acal · 2013 · cites it directly1descendant
  • K242255Qitexio® 4-Way Stopcock (QIT014)Medex · 2025 · 2 generations on0descendants
  • K210516Clicky CrossYomura Technologies, Inc. · 2022 · cites it directly0descendants
  • K172266Dynarex Three-Way StopcockDynarex Corporation · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K862691StopcockRam Development1986—
−2K022895Elcam Stopcocks and ManifoldsMedical Device Consultants, Inc.200214
−1K032393Induction and Sampling Manifold (or Stopcock)Elcam Medical Acal20032
0 (this device)K111016Safeport Manifold (TM) (or Stopcock)Elcam Medical Acal2011—
+1K123084Disposable High Pressure Stopcock/Manifold Model HP Stopcock, MP Stopcock, HP…Elcam Medical Acal2013—
+2K242255Qitexio® 4-Way Stopcock (QIT014)Medex2025—
+1K210516Clicky CrossYomura Technologies, Inc.2022—
+1K172266Dynarex Three-Way StopcockDynarex Corporation2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201810
202210
202510
FIG. 1 — Lineage of K111016, 1986–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K032393Induction and Sampling Manifold (or Stopcock)
Summary page 1
…E-mail: Llovd.fishmancaelcam-medical corn Predicate Device: Elcam's Induction & Sampling Manifold (or Stopcock) cleared under 510(k) # K032393. Device Name Common/ Usual Name: SafePort Manifold"' (or Stopcock) Proprietary/ Trade Name: SafePort Manif old' (or Stopcock)…
Elcam Medical Acal20032

Also cleared against K032393: 1 other device →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123084Disposable High Pressure Stopcock/Manifold Model HP Stopcock, MP Stopcock, HP Manifold…Elcam Medical Acal20131
K210516Clicky CrossYomura Technologies, Inc.20220
K172266Dynarex Three-Way StopcockDynarex Corporation20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 29, 2025Class IIZ-0629-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0631-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0637-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0664-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0665-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0693-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0695-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0699-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0635-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0656-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Oct 29, 2025Class IIZ-0696-2026Process controlPotential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Aug 8, 2024Class IIZ-3144-2024Process change controlPotential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 3 companies; a new generation was cleared every 7.7 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • Of its descendants, K123084 (Elcam Medical Acal, 2013) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K032393.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—

What is it indicated for?

Safeport Manifold is indicated to serve as a flow control and a conduit device for ILV fluids delivery to the patient's vascular system. The product is intended for delivering 1.V drugs or fluids, allowing gravity feed, sampling, bolus injection and elimination of reflux of fluids during operation

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111016: Safeport Manifold (TM) (or Stopcock).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K111016

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026