510(k) · Cleared
Recalled · Class II · 2015Construx External Fixation Device
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ebi, LLC · KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Cleared on Aug 3, 2011 as substantially equivalent to K953406 (Electro-Biology, Inc., 1995) and 4 other predicates after a 79-day review. Its family tree reaches back 3 generations to 1995. It has been recalled once.
- Decided
- Aug 3, 2011
- Review
- 79 days
- Class
- Class II
- Product code
- KTT
- Recalls
- 1
- Predicates
- 5
Its longest line of descent runs back 3 generations to K953406 (1995). 6 ancestors in all. No later device cites it as a predicate yet.
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K953406 | Ebi X Fix Dynafix System (DFS) | Electro-Biology, Inc. | 1995 | — |
| −2 | K031919 | Ebi Xfix DFS System | Ebi, L.P. | 2003 | — |
| −1 | K040935 | Ebi Xfix DFS System | Ebi, L.P. | 2004 | — |
| 0 (this device) | K111376 | Construx External Fixation Device | Ebi, LLC | 2011 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2000 | 1 | 0 |
| 2002 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in KTT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KTT device
- #1 K242861 TrueLok Elevatescore 21.3 · 14 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (5)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K953406 | Ebi X Fix Dynafix System (DFS)Summary page 2…rux External Fixation System is the same as, or similar to, the predicate devices. Examples of predicate devices include: * Construx External Fixation System (K953406/K021I695/K023324) * DFS Aluminum Ankle Assembly (K953406) " DFS Carbon Ankle Assembly (K03 1… | 1995 |
| * | K021695 | Ebi Dynafix Diaphyseal Correction System | 2002 |
| * | K023324 | Modification to Ebi Xfix DFS SystemSummary page 2…tion System is the same as, or similar to, the predicate devices. Examples of predicate devices include: * Construx External Fixation System (K953406/K021I695/K023324) * DFS Aluminum Ankle Assembly (K953406) " DFS Carbon Ankle Assembly (K03 1919) * DFS Ca… | 2002 |
| * | K031919 | Ebi Xfix DFS SystemSummary page 2…trux External Fixation System (K953406/K021I695/K023324) * DFS Aluminum Ankle Assembly (K953406) " DFS Carbon Ankle Assembly (K03 1919) * DFS Carbon Fixator Straight Clamp (K040935) * DFS Left and Right Telescoping Arms (K953406) * DES Angular H… | 2003 |
| * | K040935 | Ebi Xfix DFS SystemSummary page 2…3324) * DFS Aluminum Ankle Assembly (K953406) " DFS Carbon Ankle Assembly (K03 1919) * DFS Carbon Fixator Straight Clamp (K040935) * DFS Left and Right Telescoping Arms (K953406) * DES Angular Hinge Clamp (K953406) " DES Ring to F… | 2004 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 1, 2015 | Class II | Z-1741-2015 | Process control | Ankle clamp was assembled incorrectly; the offset of the pin to the pivot point is reversed. This prevents complete range of motion and if pushed too far may loosen the pin from the bone. A revision surgery may be necessary to adequately… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 7.1 years on median.
- None of the 6 ancestors has a recall.
- 16 other devices also cite its predicate K953406.
How does its category compare?
All KTT devicesKTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · KTT
10in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KTT
88in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KTT
2in 2021
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · KTT
673in 2025
MAUDE reports naming product code KTT, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 9 | 55 | 2 | 45 |
| 2008 | 11 | 73 | 1 | 66 |
| 2009 | 11 | 94 | 2 | 45 |
| 2010 | 12 | 148 | 1 | 121 |
| 2011 | 10 | 157 | 0 | 196 |
| 2012 | 9 | 140 | 4 | 527 |
| 2013 | 7 | 87 | 7 | 898 |
| 2014 | 14 | 129 | 0 | 445 |
| 2015 | 12 | 60 | 0 | 401 |
| 2016 | 10 | 142 | 0 | 400 |
| 2017 | 11 | 184 | 1 | 485 |
| 2018 | 6 | 149 | 14 | 597 |
| 2019 | 8 | 110 | 3 | 800 |
| 2020 | 9 | 200 | 0 | 942 |
| 2021 | 10 | 83 | 2 | 961 |
| 2022 | 3 | 25 | 0 | 627 |
| 2023 | 7 | 29 | 0 | 554 |
| 2024 | 5 | 90 | 0 | 523 |
| 2025 | 10 | 88 | 0 | 673 |
| 2026 | 5 | 68 | 11 | — |
What is it indicated for?
The Construx External Fixation System is a unilateral external fixation device intended for
use in children and adults in the treatment of bone conditions including leg lengthening,
osteotomies, arthrodesis, fracture fixation, and other bone conditions amenable to treatment
by the use of the external fixation modality.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K111376 · printed Nov 6, 2026
Cite this page
510(k) Report. “K111376: Construx External Fixation Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111376
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026