510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Bracco Diagnostics, Inc. CO2MPACT Endodcopic Insufflator

K111648510(k) SpecialCleared Nov 3, 2011143-day reviewClass II21 CFR 876.1500Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bracco Diagnostic, Inc. · FCX Insufflator, Automatic Carbon-Dioxide For Endoscope

Cleared on Nov 3, 2011 as substantially equivalent to K053008 (E-Z-Em, Inc., 2005) and 1 other predicate after a 143-day review. Its family tree reaches back 4 generations to 1993. 3 later devices descend from it. It has been recalled once.

Decided
Nov 3, 2011
Review
143 days
Class
Class II
Product code
FCX
Recalls
1
Predicates
2

Its longest line of descent runs back 4 generations to K923982 (1993). 9 ancestors in all. 3 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor2001: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2019: 1 descendant cleared2016: 1 descendant cleared2020: 1 descendant clearedK923982 · Olympus Oes Laparoscopy System · Olympus Corp. · 1993 · recalledK955403 · Oes Laparo-Thoraco Videoscope Type V · Olympus America, Inc. · 1996K053382 · Laparo-Thoraco Videoscope XLTF-Vaw · Olympus Medical Systems Corporation · 2006K081173 · Endoscopic CO2 Regulation Unit Ucr · Olympus Medical Systems Corporation · 20081990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor3 generations back: 1 other ancestor4 generations back: 3 other ancestors 1 generation forward: 3 descendants K081173 · Endoscopic CO2 Regulation Unit Ucr · Olympus Medical Systems Corporation · 2008K053382 · Laparo-Thoraco Videoscope XLTF-Vaw · Olympus Medical Systems Corporation · 2006K955403 · Oes Laparo-Thoraco Videoscope Type V · Olympus America, Inc. · 1996K923982 · Olympus Oes Laparoscopy System · Olympus Corp. · 1993 · recalled0
  1. Root · 4 generations backK923982 Olympus Oes Laparoscopy SystemOlympus Corp. · 1993 · Recalled
  2. 3 generations backK955403 Oes Laparo-Thoraco Videoscope Type VOlympus America, Inc. · 1996
  3. 2 generations backK053382 Laparo-Thoraco Videoscope XLTF-VawOlympus Medical Systems Corporation · 2006
  4. 1 generation backK081173 Endoscopic CO2 Regulation Unit UcrOlympus Medical Systems Corporation · 2008
  5. This deviceK111648 Bracco Diagnostics, Inc. CO2MPACT Endodcopic InsufflatorBracco Diagnostic, Inc. · 2011
  6. Descendant · cites it directly · 0 of its ownK201096 CO2mbiLEDKarl Storz Endoscopy America, Inc. · 2020
  7. Descendant · cites it directly · 0 of its ownK190985 BAROnova Insufflation System, BAROnova Accessory KitBaronova, Inc. · 2019
  8. Descendant · cites it directly · 0 of its ownK162332 Nexcore GI InsufflatorNexcore Technology, LLC · 2016

Most influential descendants

Showing 3 of 3 · See all
  • K201096CO2mbiLEDKarl Storz Endoscopy America, Inc. · 2020 · cites it directly0descendants
  • K190985BAROnova Insufflation System, BAROnova Accessory KitBaronova, Inc. · 2019 · cites it directly0descendants
  • K162332Nexcore GI InsufflatorNexcore Technology, LLC · 2016 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K923982Olympus Oes Laparoscopy SystemOlympus Corp.19931
−3K955403Oes Laparo-Thoraco Videoscope Type VOlympus America, Inc.1996—
−2K053382Laparo-Thoraco Videoscope XLTF-VawOlympus Medical Systems Corporation2006—
−1K081173Endoscopic CO2 Regulation Unit UcrOlympus Medical Systems Corporation2008—
0 (this device)K111648Bracco Diagnostics, Inc. CO2MPACT Endodcopic InsufflatorBracco Diagnostic, Inc.2011—
+1K201096CO2mbiLEDKarl Storz Endoscopy America, Inc.2020—
+1K190985BAROnova Insufflation System, BAROnova Accessory KitBaronova, Inc.2019—
+1K162332Nexcore GI InsufflatorNexcore Technology, LLC2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
200110
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201910
202010
FIG. 1 — Lineage of K111648, 1993–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FCX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FCX device

  1. #1 K262152 VS-370CF, VS-370CTscore 15.3 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K053008E-Z-EM Endoscopic CO2 Regulator
Summary page 1
…510(k) Manufacturer/Model Number Bracco Diagnostics, Inc. K053008 E-Z-EM, Inc. (now BDJ) CO 2 MPACT ENDOSCOPIC / CO 2 Effxeient…
E-Z-Em, Inc.2005—
*K081173Endoscopic CO2 Regulation Unit Ucr
Summary page 1
…/ CO 2 Effxeient II4SUFFLATORTM Endoscopic Insufflator K081 173 Olympus Medical Systems, Corp. / UCR…
Olympus Medical Systems Corporation2008—

Also cleared against K053008: 5 other devices →

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K201096CO2mbiLEDKarl Storz Endoscopy America, Inc.20200
K190985BAROnova Insufflation System, BAROnova Accessory KitBaronova, Inc.20190
K162332Nexcore GI InsufflatorNexcore Technology, LLC20160

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 15, 2021Class IIZ-1148-2021Nonconforming Material/ComponentAn internal tubing component in the flow path was made with incorrect material which was not tested for use with this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 4 companies; a new generation was cleared every 3.0 years on median.
  • 2 of the 9 ancestors have a recall of any class.
  • Of its descendants, K201096 (Karl Storz Endoscopy America, Inc., 2020) has gone furthest, with 0 descendants of its own.
  • 5 other devices also cite its predicate K053008.

How does its category compare?

All FCX devices

FCX Insufflator, Automatic Carbon-Dioxide For Endoscope: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · FCX

1in 2024

2007: 12009: 12011: 12012: 12013: 12014: 22016: 12018: 12019: 12020: 12021: 12023: 12024: 12026 (partial year): 1
2007high 2 (2014) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FCX

Too few FCX 510(k)s in any year since 2007 to give a median.

Recalls per year · FCX

1in 2013

2013: 1, 1 Class I
2007high 1 (2013) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · FCX

143in 2025

2007: 52008: 112009: 82010: 202011: 142012: 132013: 192014: 102015: 52016: 42017: 32018: 72019: 42020: 12021: 122022: 292023: 542024: 352025: 143
2007high 143 (2025) · low 1 (2020)2025

MAUDE reports naming product code FCX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—05
20080—011
20091—08
20100—020
20111—014
20121—013
20131—1 (1)19
20142—010
20150—05
20161—04
20170—03
20181—07
20191—04
20201—01
20211—012
20220—029
20231—054
20241—035
20250—0143
20261—0—

What is it indicated for?

The Bracco Diagnostics, Inc. (BDI) CO2 MPACT ENDOSCOPIC INSUFFLATOR is designed to use CO2 as a distention media in the gastrointestinal tract when used in conjunction with a gastrointestinal endoscope.

The BDI CO2 MPACT ENDOSCOPIC INSUFFLATOR allows the physician to modulate and control the CO2 to the patient in an identical manner to that being done currently with room air which is supplied by an air pump in the light source of the endoscopic equipment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111648: Bracco Diagnostics, Inc. CO2MPACT Endodcopic Insufflator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111648

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026