510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Endo Gia (TM) Staplers, DST Series (TM) Gia (TM) Staplers, DST Series (TM) TA (TM) Staplers

K111825510(k) TraditionalCleared Oct 25, 2011119-day reviewClass II21 CFR 878.4750Recalls 73

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Covidien, Formerly US Surgical A Divison of Tyco H · GDW Staple, Implantable

Cleared on Oct 25, 2011 as substantially equivalent to K061095 (United States Surgical, A Division of Tyco Healthc, 2006) and 8 other predicates after a 119-day review. Its family tree reaches back 10 generations to 1985. 44 later devices descend from it. It has been recalled 73 times.

Decided
Oct 25, 2011
Review
119 days
Class
Class II
Product code
GDW
Recalls
73
Predicates
9

Its longest line of descent runs back 10 generations to K842066 (1985). 42 ancestors in all, the oldest K843034 (1984). 44 later devices descend from it, 22 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2010: 2 other ancestors2009: 4 other ancestors2007: 2 other ancestors2006: 2 other ancestors2003: 1 other ancestor2004: 2 other ancestors2000: 2 other ancestors2008: 1 other ancestor1999: 1 other ancestor2001: 1 other ancestor1996: 4 other ancestors1994: 4 other ancestors1993: 1 other ancestor1992: 3 other ancestors1997: 1 other ancestor1984: 1 other ancestor2012: 4 descendants cleared2016: 5 descendants cleared2019: 1 descendant cleared2017: 6 descendants cleared2020: 4 descendants cleared2026: 1 descendant cleared2013: 4 descendants cleared2018: 2 descendants cleared2023: 3 descendants cleared2014: 5 descendants cleared2022: 4 descendants cleared2015: 2 descendants cleared2021: 3 descendants clearedK842066 · Peri-Guard Pericardial Patch · Genetic Laboratories, Inc. · 1985K961811 · Peri-Guard Pericardium · Bio-Vascular, Inc. · 1996K983162 · Peri-Guard, CV Peri-Guard, Ocu-Guard, Supple Peri-Guard, Peri-Strips - Sleeve, Peri-Strips-Strips, Peri-Strips Dry, Vasc · Bio-Vascular, Inc. · 1998 · recalledK992537 · Peri-Strips-Sleeve, Peri-Strips-Strip and Peri-Strips Dry (Non-Naoh-Treated and Naoh-Treated · Bio-Vascular, Inc. · 2000K014183 · Afb (Absorbable Foam Buttress) · J&J Corporate Biomaterials Center · 2002K030782 · Seamguard Bioabsorbable Staple Line Reinforcement Material · W. L. Gore & Associates, Inc. · 2003×2: 2 generations cleared close together, K032865 (2003), K040119 (2004) K080898 · Auto Suture Endo Gia Staplers with Endo Gia Single Use Loading Units with Staple Line Reinforcement · Tyco Healthcare Group LP · 2008 · recalledK103263 · Endo Gia Singleuse Duet TRS Reload with TRI0STAPLE Technology · Covidien, Formerly US Surgical A Divison of Tyco H · 201019801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 7 other ancestors2 generations back: 4 other ancestors3 generations back: 3 other ancestors4 generations back: 8 other ancestors5 generations back: 10 other ancestors 1 generation forward: 22 descendants2 generations forward: 18 descendants3 generations forward: 4 descendants K103263 · Endo Gia Singleuse Duet TRS Reload with TRI0STAPLE Technology · Covidien, Formerly US Surgical A Divison of Tyco H · 2010K080898 · Auto Suture Endo Gia Staplers with Endo Gia Single Use Loading Units with Staple Line Reinforcement · Tyco Healthcare Group LP · 2008 · recalledK040119 · Peri-Strips Staple Line Reinforcement-Sleeve Configuration, Peri-Strips Staple Line Reinforcement-Strip Configuration · Synovis Surgical Innovations · 2004K032865 · Gore Seamguard Bioabsorbable Staple Line Reinforcement Material · W. L. Gore & Associates, Inc. · 2003K030782 · Seamguard Bioabsorbable Staple Line Reinforcement Material · W. L. Gore & Associates, Inc. · 2003K014183 · Afb (Absorbable Foam Buttress) · J&J Corporate Biomaterials Center · 2002K992537 · Peri-Strips-Sleeve, Peri-Strips-Strip and Peri-Strips Dry (Non-Naoh-Treated and Naoh-Treated · Bio-Vascular, Inc. · 2000K983162 · Peri-Guard, CV Peri-Guard, Ocu-Guard, Supple Peri-Guard, Peri-Strips - Sleeve, Peri-Strips-Strips, Peri-Strips Dry, Vasc · Bio-Vascular, Inc. · 1998 · recalledK961811 · Peri-Guard Pericardium · Bio-Vascular, Inc. · 1996K842066 · Peri-Guard Pericardial Patch · Genetic Laboratories, Inc. · 1985−10−50
  1. Root · 10 generations backK842066 Peri-Guard Pericardial PatchGenetic Laboratories, Inc. · 1985
  2. 6 generations between
  3. 3 generations backK040119 Peri-Strips Staple Line Reinforcement-Sleeve Configuration…Synovis Surgical Innovations · 2004
  4. 2 generations backK080898 Auto Suture Endo Gia Staplers with Endo Gia Single Use Loading Units…Tyco Healthcare Group LP · 2008 · Recalled
  5. 1 generation backK103263 Endo Gia Singleuse Duet TRS Reload with TRI0STAPLE TechnologyCovidien, Formerly US Surgical A Divison of Tyco H · 2010
  6. This deviceK111825 Endo Gia (TM) Staplers, DST Series (TM) Gia (TM) Staplers, DST…Covidien, Formerly US Surgical A Divison of Tyco H · 2011
  7. Descendant · cites it directly · 8 of its ownK121510 Idrive Ultra Powered Handle Endo Gia AdapterCovidien, Formerly US Surgical A Divison of Tyco H · 2012
  8. Descendant · 2 generations on · 5 of its ownK160176 Signia StaplerCovidien · 2016 · Recalled
  9. Descendant · cites it directly · 3 of its ownK192720 GIA Stapler with Tri-Staple TechnologyCovidien · 2020
  10. Descendant · cites it directly · 3 of its ownK161757 Single Use Curved Cutters, Single Use Circular Staplers, Single Use…Victor Medical Instruments Co., Ltd. · 2016
  11. Descendant · cites it directly · 3 of its ownK132581 Microcutter Xchange 30, Microcutter Xchange 30, Blue Staple…Cardica, Inc. · 2014 · Recalled
Show all 10 generations
  1. Root · 10 generations backK842066 Peri-Guard Pericardial PatchGenetic Laboratories, Inc. · 1985
  2. 9 generations backK961811 Peri-Guard PericardiumBio-Vascular, Inc. · 1996
  3. 8 generations backK983162 Peri-Guard, CV Peri-Guard, Ocu-Guard, Supple Peri-Guard, Peri-Strips…Bio-Vascular, Inc. · 1998 · Recalled
  4. 7 generations backK992537 Peri-Strips-Sleeve, Peri-Strips-Strip and Peri-Strips Dry…Bio-Vascular, Inc. · 2000
  5. 6 generations backK014183 Afb (Absorbable Foam Buttress)J&J Corporate Biomaterials Center · 2002
  6. 5 generations backK030782 Seamguard Bioabsorbable Staple Line Reinforcement MaterialW. L. Gore & Associates, Inc. · 2003
  7. 4 generations backK032865 Gore Seamguard Bioabsorbable Staple Line Reinforcement MaterialW. L. Gore & Associates, Inc. · 2003
  8. 3 generations backK040119 Peri-Strips Staple Line Reinforcement-Sleeve Configuration…Synovis Surgical Innovations · 2004
  9. 2 generations backK080898 Auto Suture Endo Gia Staplers with Endo Gia Single Use Loading Units…Tyco Healthcare Group LP · 2008 · Recalled
  10. 1 generation backK103263 Endo Gia Singleuse Duet TRS Reload with TRI0STAPLE TechnologyCovidien, Formerly US Surgical A Divison of Tyco H · 2010
  11. This deviceK111825 Endo Gia (TM) Staplers, DST Series (TM) Gia (TM) Staplers, DST…Covidien, Formerly US Surgical A Divison of Tyco H · 2011

Showing 5 of 44 descendants. See all descendants

Descendants cleared per year

2012: 42013: 42014: 52015: 22016: 52017: 62018: 22019: 12020: 42021: 32022: 42023: 32026: 1
2012high 6 (2017) · low 0 (2024)2026

Most influential descendants

Showing 5 of 44 · See all
  • K121510Idrive Ultra Powered Handle Endo Gia AdapterCovidien, Formerly US Surgical A Divison of Tyco H · 2012 · cites it directly8descendants
  • K160176Signia StaplerCovidien · 2016 · 2 generations on5descendants
  • K192720GIA Stapler with Tri-Staple TechnologyCovidien · 2020 · cites it directly3descendants
  • K161757Single Use Curved Cutters, Single Use Circular Staplers, Single Use Linear Cutters and…Victor Medical Instruments Co., Ltd. · 2016 · cites it directly3descendants
  • K132581Microcutter Xchange 30, Microcutter Xchange 30, Blue Staple Cartridge, Blue Staple…Cardica, Inc. · 2014 · cites it directly3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K842066Peri-Guard Pericardial PatchGenetic Laboratories, Inc.1985—
−9K961811Peri-Guard PericardiumBio-Vascular, Inc.1996—
−8K983162Peri-Guard, CV Peri-Guard, Ocu-Guard, Supple Peri-Guard, Peri-Strips - Sleeve…Bio-Vascular, Inc.19982
−7K992537Peri-Strips-Sleeve, Peri-Strips-Strip and Peri-Strips Dry (Non-Naoh-Treated…Bio-Vascular, Inc.2000—
−6K014183Afb (Absorbable Foam Buttress)J&J Corporate Biomaterials Center2002—
−5K030782Seamguard Bioabsorbable Staple Line Reinforcement MaterialW. L. Gore & Associates, Inc.2003—
−4K032865Gore Seamguard Bioabsorbable Staple Line Reinforcement MaterialW. L. Gore & Associates, Inc.2003—
−3K040119Peri-Strips Staple Line Reinforcement-Sleeve Configuration, Peri-Strips Staple…Synovis Surgical Innovations2004—
−2K080898Auto Suture Endo Gia Staplers with Endo Gia Single Use Loading Units with…Tyco Healthcare Group LP200824
−1K103263Endo Gia Singleuse Duet TRS Reload with TRI0STAPLE TechnologyCovidien, Formerly US Surgical A Divison of Tyco H2010—
0 (this device)K111825Endo Gia (TM) Staplers, DST Series (TM) Gia (TM) Staplers, DST Series (TM) TA…Covidien, Formerly US Surgical A Divison of Tyco H2011—
+1K121510Idrive Ultra Powered Handle Endo Gia AdapterCovidien, Formerly US Surgical A Divison of Tyco H2012—
+2K160176Signia StaplerCovidien201614
+1K192720GIA Stapler with Tri-Staple TechnologyCovidien2020—
+1K161757Single Use Curved Cutters, Single Use Circular Staplers, Single Use Linear…Victor Medical Instruments Co., Ltd.2016—
+1K132581Microcutter Xchange 30, Microcutter Xchange 30, Blue Staple Cartridge, Blue…Cardica, Inc.20141

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
199230
199310
199440
199640
199710
199910
200020
200110
200310
200420
200620
200720
200810
200940
201020

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201240
201340
201450
201520
201650
201760
201820
201910
202040
202130
202240
202330
202610
FIG. 1 — Lineage of K111825, 1985–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GDW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GDW device

  1. #1 K253657 Tri-staple 2.0™ Reloads; Endo GIA™ Reloads…score 15.0 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (9)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061095Auto Sutur Endo Gia Staplers
Summary page 1
…a. Panel no and product code: 79-GDW b. Regulation no: 21 CFR 878.4750 PREDICATE DEVICE(S): Endo GIA T M Staplers (K061095, K083519, K(093410, K101444, K103263,1K102291,1K080898)…
United States Surgical, A Division of Tyco Healthc2006—
*K083519Autosuture Endo Gia Staplers with Endo Gia Single Use Loading Units
Summary page 1
…. Panel no and product code: 79-GDW b. Regulation no: 21 CFR 878.4750 PREDICATE DEVICE(S): Endo GIA T M Staplers (K061095, K083519, K(093410, K101444, K103263,1K102291,1K080898)…
Covidien Lp, Formerly Registered AS United States2009—
*K093410Autosuture Endo Gia Single Use Black Reload with Tri-Staple Technology
Summary page 1
…and product code: 79-GDW b. Regulation no: 21 CFR 878.4750 PREDICATE DEVICE(S): Endo GIA T M Staplers (K061095, K083519, K(093410, K101444, K103263,1K102291,1K080898) DST S…
Surgical Devices, A Divsion of Tyco Healthcare2009—
*K101444Autosuture Endo Gia Single Use Tan Reload with Tri-Staple Technology
Summary page 1
…uct code: 79-GDW b. Regulation no: 21 CFR 878.4750 PREDICATE DEVICE(S): Endo GIA T M Staplers (K061095, K083519, K(093410, K101444, K103263,1K102291,1K080898) DST Series m G…
Surgical Devices, A Global Business Unit Tyco2010—
*K103263Endo Gia Singleuse Duet TRS Reload with TRI0STAPLE Technology
Summary page 1
…on no: 21 CFR 878.4750 PREDICATE DEVICE(S): Endo GIA T M Staplers (K061095, K083519, K(093410, K101444, K103263,1K102291,1K080898) DST Series m GIAIV Staplers (K032696) DST Seri…
Covidien, Formerly US Surgical A Divison of Tyco H2010—
*K102291Covidien Endo Gia Radial Reload with Tri-Staple Technology
Summary page 1
…CFR 878.4750 PREDICATE DEVICE(S): Endo GIA T M Staplers (K061095, K083519, K(093410, K101444, K103263,1K102291,1K080898) DST Series m GIAIV Staplers (K032696) DST Series TM TA…
Covidien2010—
*K080898Auto Suture Endo Gia Staplers with Endo Gia Single Use Loading Units with Staple Line…
Summary page 1
…4750 PREDICATE DEVICE(S): Endo GIA T M Staplers (K061095, K083519, K(093410, K101444, K103263,1K102291,1K080898) DST Series m GIAIV Staplers (K032696) DST Series TM TA TM Staple…
Tyco Healthcare Group LP200824
*K032696Auto Suture Endoscopic (& Open) TA Surgical Stapler, Auto Suture Endoscopic (& Open) Gia…
Summary page 1
…083519, K(093410, K101444, K103263,1K102291,1K080898) DST Series m GIAIV Staplers (K032696) DST Series TM TA TM Staplers (K032696) ENDOPATH Linear Cutters a…
United States Surgical, A Division of Tyco Healthc2003—
*K070887Endopath Linear Cutters and Staplers and ECHELON60 Endoscopic Linear Cutters and…
Summary page 1
…s m GIAIV Staplers (K032696) DST Series TM TA TM Staplers (K032696) ENDOPATH Linear Cutters and Staplers (K070887) Echelon Linear Cutters and Staplers (K070887 DEVICE DESCRIPTION: The Endo GIA T…
Ethicon Endo-Surgery, LLC20074

Also cleared against K061095: 6 other devices →

What descends from it? (44)

See all 44

22 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K121510Idrive Ultra Powered Handle Endo Gia AdapterCovidien, Formerly US Surgical A Divison of Tyco H20128
K192720GIA Stapler with Tri-Staple TechnologyCovidien20203
K161757Single Use Curved Cutters, Single Use Circular Staplers, Single Use Linear Cutters and…Victor Medical Instruments Co., Ltd.20163
K132581Microcutter Xchange 30, Microcutter Xchange 30, Blue Staple Cartridge, Blue Staple…Cardica, Inc.20143
K191070Signia Small Diameter ReloadsCovidien20192
K172960easyEndoTM Linear Cutting Stapler and Loading Units for Single Use, easyEndoTM Universal…Ezisurg Medical Co., Ltd.20171
K162707Single Use Circular Stapler, Single Use Hemorrhoidal Circular Stapler, Single Patient…Golden Stapler Surgical Co., Ltd.20171
K151163Signia Loading Unit, Tri-Staple 2.0 CartridgeCovidien, LLC20151
K133938Endo Gia Reinforced Reload with Tri-Staple TechnologyCovidien20141
K131705Endo Gia Radial Reload with Tri-Staple TechnologyCovidien, Formerly US Surgical A Divison of Tyco H20131

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 23, 2020Class IIZ-0747-2021Process controlMay contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue…
Nov 23, 2020Class IIZ-0748-2021Process controlMay contain a damaged internal component which could impede staple deployment and result in incomplete staple formation and a non-functional staple line preventing adequate hemostasis, or leading to anastomotic leak, pneumothorax, tissue…
Aug 12, 2020Class IIZ-0001-2021Under Investigation by firmThe device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Aug 12, 2020Class IIZ-0005-2021Under Investigation by firmThe device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Aug 12, 2020Class IIZ-0004-2021Under Investigation by firmThe device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Aug 12, 2020Class IIZ-0002-2021Under Investigation by firmThe device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Aug 12, 2020Class IIZ-0008-2021Under Investigation by firmThe device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Aug 12, 2020Class IIZ-0007-2021Under Investigation by firmThe device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Aug 12, 2020Class IIZ-0009-2021Under Investigation by firmThe device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Aug 12, 2020Class IIZ-0006-2021Under Investigation by firmThe device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Aug 12, 2020Class IIZ-0003-2021Under Investigation by firmThe device staples may not properly form upon application preventing adequate hemostasis. Use of a product with this assembly error may result in incomplete staple formation.
Jan 17, 2020Class IIZ-1264-2020Package design/selectionPotential for the sterile barrier to be compromised. Products with a compromised sterile barrier may result in a potentially increased risk for infection
May 17, 2019Class IIZ-1829-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1831-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1834-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1833-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1836-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1825-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1832-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1830-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1827-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1835-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1826-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 17, 2019Class IIZ-1828-2019Process controlThe device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the…
May 22, 2018Class IIZ-2503-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
May 22, 2018Class IIZ-2505-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
May 22, 2018Class IIZ-2504-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1855-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1866-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1826-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1840-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1824-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1844-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1859-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1837-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1833-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1852-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1828-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1842-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1822-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1817-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1831-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1843-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1863-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1815-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1849-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1814-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1839-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1850-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Apr 26, 2018Class IIZ-1865-2018Process controlDevice may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 7 companies; a new generation was cleared every 2.3 years on median.
  • 8 of the 42 ancestors have a recall of any class.
  • Of its descendants, K121510 (Covidien, Formerly US Surgical A Divison of Tyco H, 2012) has gone furthest, with 8 descendants of its own.
  • 6 other devices also cite its predicate K061095.

How does its category compare?

All GDW devices

GDW Staple, Implantable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · GDW

3in 2025

2007: 82008: 72009: 132010: 102011: 112012: 92013: 132014: 152015: 82016: 112017: 132018: 72019: 122020: 102021: 92022: 72023: 42024: 22025: 32026 (partial year): 3
2007high 15 (2014) · low 2 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GDW

111in 2025

2007: 56 days median2008: 57 days median2009: 27 days median2010: 39 days median2011: 49 days median2012: 45 days median2013: 49 days median2014: 63 days median2015: 111 days median2016: 96 days median2017: 103 days median2018: 90 days median2019: 205 days median2020: 88 days median2021: 233 days median2022: 89 days median2023: 267 days median2025: 111 days median2026 (partial year): 141 days median
2007high 267 (2023) · low 27 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GDW

3in 2025

2007: 92009: 222010: 22011: 92012: 11, 1 Class I2013: 12015: 62016: 22017: 32018: 32019: 4, 4 Class I2020: 112023: 12025: 3
2007high 22 (2009) · low 0 (2008)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · GDW

11,905in 2025

2007: 2,4012008: 2,3802009: 4,0142010: 3,7682011: 3,8262012: 3,7172013: 3,7772014: 4,7842015: 4,8992016: 5,1062017: 11,6432018: 14,9142019: 16,6002020: 12,3322021: 13,0152022: 13,6052023: 14,6232024: 13,2802025: 11,905
2007high 16,600 (2019) · low 2,380 (2008)2025

MAUDE reports naming product code GDW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200785692,401
200875702,380
20091327224,014
2010103923,768
2011114993,826
201294511 (1)3,717
2013134913,777
2014156304,784
2015811164,899
2016119625,106
201713103311,643
2018790314,914
2019122054 (4)16,600
202010881112,332
20219233013,015
2022789013,605
20234267114,623
20242—013,280
20253111311,905
202631410—

What is it indicated for?

Endo GIA TM Staplers: The following indications are applicable to the Endo GIA TM Ultra Universal Stapler with Endo GIA TM Reloads with Tni-Staple TM Technology including Curved Tip Reloads: The Endo GIA TM Ultra Universal Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and…

Read the full statement (556 words)

Endo GIA TM Staplers:

The following indications are applicable to the Endo GIA TM Ultra Universal Stapler with Endo GIA TM Reloads with Tni-Staple TM Technology including Curved Tip Reloads:

The Endo GIA TM Ultra Universal Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.

The Endo GIA TM Ultra Universal Staplers when used with the Endo GIA TM Curved-Tip Reloads can be used to blunt dissect or separate target tissue from other certain tissue.

The following indications are applicable to the Endo GIA TM Ultra Universal Stapler with Duet TRS TM Reloads with Tni-Staple TM Technology including Duet TRS TM Curved Tip Reloads:

The Endo GIA TM Ultra Universal Staplers with Duet TRS TM Reloads with Tni-Staple TM Technology have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.

The Endo GIA TM Ultra Universal Staplers when used with the Duet TRS TM Curved-Tip Reloads can be used to blunt dissect or separate target tissue from other certain tissue.

The following indications are applicable to the Endo GIA TM Ultra Universal Stapler with Radial Reload with Tri-Staple TM Technology:

The Endo GIA TM Radial Reload with Tni-Staple TM Technology has application in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e. low anterior resection.

It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.

The following indications are applicable to the Endo GIA TM Universal Stapler with Endo GIA TM Straight and Roticulator TM SULUs:

The Endo GIA TM Universal Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.

The following indications are applicable to the Endo GIA TM Universal Stapler with Duet TRS TM Reloads:

The Endo GIA TM Ultra Universal Staplers with Duet TRS TM Reloads have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas.

DST Series GIA TM:

The Auto Suture GIA TM Reloadable Staplers and the SGIA TM Knifeless Stapler have applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis. The SGIA TM Knifeless stapler may be used for occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas.

DST Series TA TM:

The Auto Suture TA TM Reloadable staplers have applications in abdominal, gynecological, pediatric and thoracic surgical procedures for resection or transection of tissue and creation of anastomosis, including occlusion of the left atrial appendage in open procedures. They may be used for transection and resection of pancreas.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111825: Endo Gia (TM) Staplers, DST Series (TM) Gia (TM) Staplers, DST Series (TM) TA (TM) Staplers.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111825

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026