510(k) Report

510(k) · Cleared

Cleared

M/29TM - Pressure Injectable

K111939510(k) TraditionalCleared Jul 11, 2012369-day reviewClass II21 CFR 880.5200

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Flexicath, Ltd. · FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

Cleared on Jul 11, 2012 as substantially equivalent to K080793 (Flexicath, Ltd., 2008) and 2 other predicates after a 369-day review. Its family tree reaches back 9 generations to 2000. 2 of the 36 devices in its family, counting itself, have a Class I recall.

Decided
Jul 11, 2012
Review
369 days
Class
Class II
Product code
FOZ
Recalls
None
Predicates
3

Its longest line of descent runs back 9 generations to K001901 (2000). 35 ancestors in all, the oldest K823553 (1983). No later device cites it as a predicate yet. 2 of the 36 devices in this family, counting this one, have a Class I recall.

  1. Root · 9 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
  2. 5 generations between
  3. 3 generations backK051991 Powerpicc, Poly Per-Q-Cath, 6 FR TL Poly Per-Q-Cath, Power Hohn and…C.R. Bard, Inc. · 2005
  4. 2 generations backK080604 Pressure Injfectable PiccArrow International, Inc. · 2008 · Recalled
  5. 1 generation backK100635 Arrowgard Evolution Antimicrobial Peripherally Inserted Central…Arrow Intl., Inc. · 2010
  6. This deviceK111939 M/29TM - Pressure InjectableFlexicath, Ltd. · 2012
Show all 9 generations
  1. Root · 9 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
  2. 8 generations backK012902 Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly…C.R. Bard, Inc. · 2001
  3. 7 generations backK031129 Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc. · 2003
  4. 6 generations backK033389 Powerpicc Catheters, Models 3175155, 3175135C.R. Bard, Inc. · 2004 · Recalled
  5. 5 generations backK050185 Powerhohn and Powerline CathetersC.R. Bard, Inc. · 2005
  6. 4 generations backK051417 6 FR FL Powerhohn and Powerline CatheterC.R. Bard, Inc. · 2005 · Recalled
  7. 3 generations backK051991 Powerpicc, Poly Per-Q-Cath, 6 FR TL Poly Per-Q-Cath, Power Hohn and…C.R. Bard, Inc. · 2005
  8. 2 generations backK080604 Pressure Injfectable PiccArrow International, Inc. · 2008 · Recalled
  9. 1 generation backK100635 Arrowgard Evolution Antimicrobial Peripherally Inserted Central…Arrow Intl., Inc. · 2010
  10. This deviceK111939 M/29TM - Pressure InjectableFlexicath, Ltd. · 2012

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K001901Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath Midline…C.R. Bard, Inc.20002
−8K012902Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly Per-Q-Cath…C.R. Bard, Inc.2001—
−7K031129Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc.2003—
−6K033389Powerpicc Catheters, Models 3175155, 3175135C.R. Bard, Inc.20041
−5K050185Powerhohn and Powerline CathetersC.R. Bard, Inc.2005—
−4K0514176 FR FL Powerhohn and Powerline CatheterC.R. Bard, Inc.20051
−3K051991Powerpicc, Poly Per-Q-Cath, 6 FR TL Poly Per-Q-Cath, Power Hohn and Power LineC.R. Bard, Inc.2005—
−2K080604Pressure Injfectable PiccArrow International, Inc.20081
−1K100635Arrowgard Evolution Antimicrobial Peripherally Inserted Central Catheter…Arrow Intl., Inc.2010—
0 (this device)K111939M/29TM - Pressure InjectableFlexicath, Ltd.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198320
198611
199210
199410
199610
199810
199910
200031
200130
200310
200420
200530
200620
200710
200820
200910
FIG. 1 — Lineage of K111939, 2000–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FOZ device

  1. #1 K243403 BD Nexiva™ Closed IV Catheter Systemscore 14.8 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080793Firmgrip
Summary page 2
…Page I of 3 Predicate Devices: 1) Flexicath'sFirmGripTM PeripherallyInserted Catheter Device, cleared under 5 10(k) number: K080793 and;- 2) Flexicath's FirmsGripTM PeripherallyInserted CathPeterDevice with the SNM (Safety Needle Mechanism) feature,…
Flexicath, Ltd.2008—
*K092629SNM Firmgrip
Summary page 2
…r: K080793 and;- 2) Flexicath's FirmsGripTM PeripherallyInserted CathPeterDevice with the SNM (Safety Needle Mechanism) feature, cleared under 5 10(k) number: K092629 and;- 3) Arrow's Antimicrobial Pressure Injectable P1CC, cleared under 5 10(k) number: K100635. Device description:…
Flexicath, Ltd.2009—
*K100635Arrowgard Evolution Antimicrobial Peripherally Inserted Central Catheter (Picc) Model…
Summary page 2
…Needle Mechanism) feature, cleared under 5 10(k) number: K092629 and;- 3) Arrow's Antimicrobial Pressure Injectable P1CC, cleared under 5 10(k) number: K100635. Device description: Flexicath's M/29r Pressure Injectable device is actually the very same device cleared under K(0807…
Arrow Intl., Inc.2010—

Also cleared against K080793: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K993691Arrowg +Ard Blue Plus Antimicrobial Multi-Lumen Central Venous CatheterArrow Intl., Inc. · 2000 · 5 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K862056[Central Venous][Multi-Lumen] Catheter W/Flex TipArrow Intl., Inc. · 1986 · 6 generations back · ◆ in the 1986 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 12 years and passes through 4 companies; a new generation was cleared every 1.2 years on median.
  • 14 of the 35 ancestors have a recall of any class; 2 have a Class I recall.
  • 1 other device also cites its predicate K080793.

How does its category compare?

All FOZ devices

FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · FOZ

11in 2025

2007: 62008: 92009: 92010: 32011: 62012: 132013: 42014: 92015: 102016: 92017: 92018: 12019: 82020: 102021: 42022: 82023: 92024: 62025: 112026 (partial year): 7
2007high 13 (2012) · low 1 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FOZ

91in 2025

2007: 159 days median2008: 103 days median2009: 30 days median2010: 28 days median2011: 132 days median2012: 122 days median2013: 104 days median2014: 99 days median2015: 107 days median2016: 63 days median2017: 51 days median2019: 140 days median2020: 191 days median2021: 242 days median2022: 183 days median2023: 129 days median2024: 190 days median2025: 91 days median2026 (partial year): 203 days median
2007high 242 (2021) · low 28 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FOZ

4in 2023

2007: 162009: 232010: 1152013: 22014: 42016: 3, 1 Class I2017: 222018: 42019: 32020: 32021: 12022: 4, 2 Class I2023: 4, 2 Class I2026 (partial year): 2
2007high 115 (2010) · low 0 (2008)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · FOZ

5,110in 2025

2007: 3742008: 4222009: 5082010: 3702011: 7862012: 5602013: 6402014: 8152015: 7292016: 9252017: 1,9662018: 3,3052019: 4,3652020: 3,7972021: 4,6742022: 4,1332023: 4,4302024: 4,7722025: 5,110
2007high 5,110 (2025) · low 370 (2010)2025

MAUDE reports naming product code FOZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007615916374
200891030422
200993023508
2010328115370
201161320786
2012131220560
201341042640
20149994815
2015101070729
20169633 (1)925
2017951221,966
20181—43,305
2019814034,365
20201019133,797
2021424214,674
202281834 (2)4,133
202391294 (2)4,430
2024619004,772
2025119105,110
202672032—

What is it indicated for?

The Pressure Injectable M/29TM is a Peripherally Inserted Catheter Device which is intended for use in patients requiring repeated access to the peripheral venous system for infusion or injection intravenous therapies and/or blood sampling and pressure injection applications such as contrast media injection.

The maximum flow rate for Flexicath Pressure Injectable midline catheter may not exceed 5ml/sec.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111939: M/29TM - Pressure Injectable.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111939

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026