510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

Enseal G2 Tissus Sealers

K112033510(k) TraditionalCleared Nov 8, 2011116-day reviewClass II21 CFR 878.4400Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ethicon Endo-Sugery, LLC · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Nov 8, 2011 as substantially equivalent to K072177 (Surgrx, Inc., 2007) and 1 other predicate after a 116-day review. Its family tree reaches back 1 generation to 2007. 76 later devices descend from it. It has been recalled 5 times.

Decided
Nov 8, 2011
Review
116 days
Class
Class II
Product code
GEI
Recalls
5
Predicates
2

Its longest line of descent runs back 1 generation to K072177 (2007). 2 ancestors in all. 76 later devices descend from it, 7 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2013: 4 descendants cleared2021: 10 descendants cleared2018: 6 descendants cleared2017: 7 descendants cleared2015: 4 descendants cleared2016: 11 descendants cleared2014: 1 descendant cleared2023: 8 descendants cleared2020: 6 descendants cleared2025: 5 descendants cleared2022: 4 descendants cleared2026: 5 descendants cleared2024: 5 descendants clearedK072177 · Enseal Vessel Sealing and Hemostasis System · Surgrx, Inc. · 2007 · recalled20052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 7 descendants2 generations forward: 17 descendants3 generations forward: 18 descendants4 generations forward: 19 descendants5 generations forward: 11 descendants6 generations forward: 3 descendants7 generations forward: 1 descendant K072177 · Enseal Vessel Sealing and Hemostasis System · Surgrx, Inc. · 2007 · recalled0+5
  1. Root · 1 generation backK072177 Enseal Vessel Sealing and Hemostasis SystemSurgrx, Inc. · 2007 · Recalled
  2. This deviceK112033 Enseal G2 Tissus SealersEthicon Endo-Sugery, LLC · 2011
  3. Descendant · cites it directly · 56 of its ownK123444 Ligasure Impact Curved, Large Jaw, Open Sealer / DividerCovidien, Formerly Valleylab, A Division of Tyco H · 2013
  4. Descendant · 2 generations on · 40 of its ownK141153 Ligasure; Curved Small Jaw Sealer/Divider, 5MM Blunt Tip…Covidien · 2014
  5. Descendant · cites it directly · 16 of its ownK132612 Harmonic Ace+ Shears 23 CM Length with Advanced Hemostasis, Harmonic…Ethicon Endo-Surgery, LLC · 2013 · Recalled
  6. Descendant · 3 generations on · 14 of its ownK152286 LigaSure Curved, Small Jaw, Open Sealer/DividerCovidien · 2015
  7. Descendant · 3 generations on · 11 of its ownK170869 LigaSure Maryland Jaw Sealer/Divider One-step Sealing, Nano-coatedCovidien · 2017

Showing 5 of 76 descendants. See all descendants

Descendants cleared per year

2013: 42014: 12015: 42016: 112017: 72018: 62020: 62021: 102022: 42023: 82024: 52025: 52026: 5
2013high 11 (2016) · low 0 (2019)2026

Most influential descendants

Showing 5 of 76 · See all
  • K123444Ligasure Impact Curved, Large Jaw, Open Sealer / DividerCovidien, Formerly Valleylab, A Division of Tyco H · 2013 · cites it directly56descendants
  • K141153Ligasure; Curved Small Jaw Sealer/Divider, 5MM Blunt Tip Sealer/Divider, Maryland Jaw…Covidien · 2014 · 2 generations on40descendants
  • K132612Harmonic Ace+ Shears 23 CM Length with Advanced Hemostasis, Harmonic Ace + Laparoscopic…Ethicon Endo-Surgery, LLC · 2013 · cites it directly16descendants
  • K152286LigaSure Curved, Small Jaw, Open Sealer/DividerCovidien · 2015 · 3 generations on14descendants
  • K170869LigaSure Maryland Jaw Sealer/Divider One-step Sealing, Nano-coatedCovidien · 2017 · 3 generations on11descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K072177Enseal Vessel Sealing and Hemostasis SystemSurgrx, Inc.20071
0 (this device)K112033Enseal G2 Tissus SealersEthicon Endo-Sugery, LLC2011—
+1K123444Ligasure Impact Curved, Large Jaw, Open Sealer / DividerCovidien, Formerly Valleylab, A Division of Tyco H2013—
+2K141153Ligasure; Curved Small Jaw Sealer/Divider, 5MM Blunt Tip Sealer/Divider…Covidien2014—
+1K132612Harmonic Ace+ Shears 23 CM Length with Advanced Hemostasis, Harmonic Ace +…Ethicon Endo-Surgery, LLC20131
+3K152286LigaSure Curved, Small Jaw, Open Sealer/DividerCovidien2015—
+3K170869LigaSure Maryland Jaw Sealer/Divider One-step Sealing, Nano-coatedCovidien2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201340
201410
201540
2016110
201770
201860
202060
2021100
202240
202380
202450
202550
202650
FIG. 1 — Lineage of K112033, 2007–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072177Enseal Vessel Sealing and Hemostasis System
Summary page 1
…400, Product Code GEl) * Electrocautery, Gynecologic and Accessories (21 CFR 884.4120, Product Code HGI) Predicate Devices ENSEAL Tissue Sealing Devices cleared under K072 177 on August 29, 2007 and K072493 on September 5, 2007 as part of the ENSEAL Vessel Sealing and Hemostasis System Device Descriptio…
Surgrx, Inc.20071
*K072493Enseal Tissue Sealing Device
Summary page 1
…cautery, Gynecologic and Accessories (21 CFR 884.4120, Product Code HGI) Predicate Devices ENSEAL Tissue Sealing Devices cleared under K072 177 on August 29, 2007 and K072493 on September 5, 2007 as part of the ENSEAL Vessel Sealing and Hemostasis System Device Description: The Ethicon Endo-Surgery EN…
Surgrx, Inc.2007—

Also cleared against K072177: 5 other devices →

What descends from it? (76)

See all 76

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123444Ligasure Impact Curved, Large Jaw, Open Sealer / DividerCovidien, Formerly Valleylab, A Division of Tyco H201356
K132612Harmonic Ace+ Shears 23 CM Length with Advanced Hemostasis, Harmonic Ace + Laparoscopic…Ethicon Endo-Surgery, LLC201316
K131435Enseal G2 Tissue Sealer, Straight Jaw, Enseal G2 Tissue Sealer, Curved Jaw, Enseal Trio…Ethicon Endo-Surgery, LLC20135
K201696ENSEAL X1 Tissue Sealer, Straight Jaw, 25 cm Shaft Length, ENSEAL X1 Tissue Sealer…Ethicon Endo-Surgery, LLC20211
K180403ENSEAL X1 Tissue Sealer, Curved Jaw, Articulating 37 cm Shaft Length; ENSEAL X1 Tissue…Ethicon Endo-Surgery, LLC20180
K172580ENSEAL X1 Tissue Sealer, Straight Jaw, 25 cm Shaft Length, ENSEAL X1 Tissue Sealer…Ethicon Endo-Surgery, LLC20170
K122797Enseal G2 Articulating Tissue Sealers, Straight Jaw, Enseal G2 Articulating Tissue…Ethicon Endo-Surgery, LLC20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 6, 2012Class IIZ-0758-2013Process controlPotential for self activation while the device is in use, which may result in unintended thermal damage.
Jan 26, 2012Not classifiedZ-1245-2012Nonconforming Material/ComponentEthicon Endo-Surgery initiated a voluntary global recall for specific production lots of ENSEAL G2 Curved and Straight Tissue Sealers due to two potential issues that may occur related to the activation button: (i) Continuous Activation…
Jan 26, 2012Not classifiedZ-1244-2012Nonconforming Material/ComponentEthicon Endo-Surgery initiated a voluntary global recall for specific production lots of ENSEAL G2 Curved and Straight Tissue Sealers due to two potential issues that may occur related to the activation button: (i) Continuous Activation…
Jan 26, 2012Not classifiedZ-1246-2012Nonconforming Material/ComponentEthicon Endo-Surgery initiated a voluntary global recall for specific production lots of ENSEAL G2 Curved and Straight Tissue Sealers due to two potential issues that may occur related to the activation button: (i) Continuous Activation…
Jan 26, 2012Not classifiedZ-1247-2012Nonconforming Material/ComponentEthicon Endo-Surgery initiated a voluntary global recall for specific production lots of ENSEAL G2 Curved and Straight Tissue Sealers due to two potential issues that may occur related to the activation button: (i) Continuous Activation…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K123444 (Covidien, Formerly Valleylab, A Division of Tyco H, 2013) has gone furthest, with 56 descendants of its own.
  • 5 other devices also cite its predicate K072177.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The ENSEAL® G2 Tissue Sealers are indicated for bipolar coagulation and mechanical transection of tissue during laparoscopic and open procedures. The devices are bipolar electrosurgical instruments for use with an electrosurgical generator. They are intended for use during open or laparoscopic, general and gynecological surgery to cut and seal vessels, and to cut, grasp and dissect tissue during surgery. Indications for use include open and laparoscopic…

Read the full statement (160 words)

The ENSEAL® G2 Tissue Sealers are indicated for bipolar coagulation and mechanical transection of tissue during laparoscopic and open procedures.

The devices are bipolar electrosurgical instruments for use with an electrosurgical generator. They are intended for use during open or laparoscopic, general and gynecological surgery to cut and seal vessels, and to cut, grasp and dissect tissue during surgery.

Indications for use include open and laparoscopic, general and gynecological surgical procedures (including urologic, thoracic, plastic and reconstructive, bowel resections, hysterectomies, cholecystectornies, gall bladder procedures, Nissen fundoplication, adhesiolysis, oophorectomies, etc.), or any procedure where vessel ligation (cutting and sealing), tissue grasping and dissection is performed. The devices can be used on vessels up to (and including) 7 mum and bundles as large as will fit in the jaws of the instruments.

The ENS EALO 02 Tissue Sealers have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use these devices for these procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K112033: Enseal G2 Tissus Sealers.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K112033

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026