510(k) · Cleared
Recalled · Class I · 2014Foursnare Vascular Retrieval Snare
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Cook, Inc. · MMX Device, Percutaneous Retrieval
Cleared on Nov 2, 2011 as substantially equivalent to K092343 (Merit Medical Systems, Inc., 2009) and 1 other predicate after a 96-day review. Its family tree reaches back 6 generations to 1990. 2 of the 22 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled once, including a Class I recall.
- Decided
- Nov 2, 2011
- Review
- 96 days
- Class
- Class II
- Product code
- MMX
- Recalls
- 1 (1 Class I)
- Predicates
- 2
Its longest line of descent runs back 6 generations to K901502 (1990). 15 ancestors in all. 6 later devices descend from it, 4 citing it directly. 2 of the 22 devices in this family, counting this one, have a Class I recall.
- Root · 6 generations backK901502 Amplatz Goose-Neck SnareMicrovena Corp. · 1990 · Recalled
- 5 generations backK972511 Amplatz Goose Neck Snare Kit/ CatheterMicrovena Corp. · 1998 · Recalled
- 4 generations backK021441 Radius SnareRadius Medical Technologies, Inc. · 2002
- 3 generations backK022201 Radius Micro SnareRadius Medical Technologies, Inc. · 2003
- 2 generations backK032426 Gunhter Tulip Vena Cava Mreye Filter and Retrieval SetCook, Inc. · 2003
- 1 generation backK073374 Cook Clect Vena Cava Filter, Gunther Tulip Vena Cava Filter…William Cook Europe Aps · 2008
- This deviceK112185 Foursnare Vascular Retrieval SnareCook, Inc. · 2011
- Descendant · cites it directly · 1 of its ownK222083 LimFlow V-CeiverLimFlow, Inc. · 2022
- Descendant · cites it directly · 1 of its ownK170987 Captus Vascular Retrieval SystemAvantec Vascular Corporation · 2017
- Descendant · 2 generations on · 0 of its ownK242776 LimFlow V-CeiverLimFlow, Inc. · 2025
- Descendant · 2 generations on · 0 of its ownK212988 ONO Retrieval DeviceOnocor Vascular, LLC · 2022
- Descendant · cites it directly · 0 of its ownK160593 Indy OTW Vascular RetrieverCook Incorporated · 2016
Showing 5 of 6 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K970466 Radius Ptca Guidewire4 generations back · 1997
Most influential descendants
Showing 5 of 6 · See all- K222083LimFlow V-CeiverLimFlow, Inc. · 2022 · cites it directly1descendant
- K170987Captus Vascular Retrieval SystemAvantec Vascular Corporation · 2017 · cites it directly1descendant
- K242776LimFlow V-CeiverLimFlow, Inc. · 2025 · 2 generations on0descendants
- K212988ONO Retrieval DeviceOnocor Vascular, LLC · 2022 · 2 generations on0descendants
- K160593Indy OTW Vascular RetrieverCook Incorporated · 2016 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K901502 | Amplatz Goose-Neck Snare | Microvena Corp. | 1990 | 1 |
| −5 | K972511 | Amplatz Goose Neck Snare Kit/ Catheter | Microvena Corp. | 1998 | 1 |
| −4 | K021441 | Radius Snare | Radius Medical Technologies, Inc. | 2002 | — |
| −3 | K022201 | Radius Micro Snare | Radius Medical Technologies, Inc. | 2003 | — |
| −2 | K032426 | Gunhter Tulip Vena Cava Mreye Filter and Retrieval Set | Cook, Inc. | 2003 | — |
| −1 | K073374 | Cook Clect Vena Cava Filter, Gunther Tulip Vena Cava Filter Retrieval Set | William Cook Europe Aps | 2008 | — |
| 0 (this device) | K112185 | Foursnare Vascular Retrieval Snare | Cook, Inc. | 2011 | — |
| +1 | K222083 | LimFlow V-Ceiver | LimFlow, Inc. | 2022 | — |
| +1 | K170987 | Captus Vascular Retrieval System | Avantec Vascular Corporation | 2017 | — |
| +2 | K242776 | LimFlow V-Ceiver | LimFlow, Inc. | 2025 | — |
| +2 | K212988 | ONO Retrieval Device | Onocor Vascular, LLC | 2022 | — |
| +1 | K160593 | Indy OTW Vascular Retriever | Cook Incorporated | 2016 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1991 | 1 | 0 |
| 1993 | 2 | 0 |
| 1997 | 1 | 1 |
| 1998 | 1 | 0 |
| 2000 | 1 | 0 |
| 2002 | 1 | 0 |
| 2007 | 1 | 0 |
| 2009 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2016 | 2 | 0 |
| 2017 | 1 | 0 |
| 2022 | 2 | 0 |
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MMX device
- #1 K242776 LimFlow V-Ceiverscore 10.9 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K092343 | EN Snare Endovascular Snare SystemSummary page 1…lar Retrieval Snare, or VRS, is substantially equivalent to the EN Snare Endovascular Snare System originally manufactured by Merit Medical under 510(k) clearance number K092343 and the Giinther Tulip Vena-Cava Filter Retrieval Set originally manufactured by Cook Incorporated under 510(k) clearance number… | 2009 |
| * | K073374 | Cook Clect Vena Cava Filter, Gunther Tulip Vena Cava Filter Retrieval Set | 2008 |
What descends from it? (6)
4 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (2 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jul 17, 2014 | Class I | Z-2243-2014 | Device Design | This remedial action is the result of six product complaints associated with separation of the snare from the distal tip of the wire. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K970466Radius Ptca GuidewireRadius Medical Technologies, Inc. · 1997 · 4 generations back · ◆ in the 1997 ancestors bar of FIG. 12 Class I recalls
What does this lineage say?
- The line of descent spans 21 years and passes through 4 companies; a new generation was cleared every 4.3 years on median.
- 3 of the 15 ancestors have a recall of any class; 1 has a Class I recall.
- Of its descendants, K222083 (LimFlow, Inc., 2022) has gone furthest, with 1 descendant of its own.
- 3 other devices also cite its predicate K092343.
How does its category compare?
All MMX devicesMMX Device, Percutaneous Retrieval: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · MMX
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MMX
151in 2022
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MMX
No recall names MMX since 2007.
Adverse event reports per year · MMX
49in 2025
MAUDE reports naming product code MMX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 0 | 2 |
| 2008 | 0 | — | 0 | 6 |
| 2009 | 2 | — | 0 | 6 |
| 2010 | 1 | — | 0 | 11 |
| 2011 | 1 | — | 0 | 15 |
| 2012 | 1 | — | 0 | 13 |
| 2013 | 0 | — | 0 | 10 |
| 2014 | 2 | — | 0 | 11 |
| 2015 | 2 | — | 0 | 5 |
| 2016 | 2 | — | 0 | 16 |
| 2017 | 2 | — | 0 | 29 |
| 2018 | 1 | — | 0 | 15 |
| 2019 | 1 | — | 0 | 17 |
| 2020 | 2 | — | 0 | 14 |
| 2021 | 1 | — | 0 | 22 |
| 2022 | 3 | 151 | 0 | 37 |
| 2023 | 2 | — | 0 | 40 |
| 2024 | 1 | — | 0 | 32 |
| 2025 | 1 | — | 0 | 49 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Vascular Retrieval Snare is intended for use in the cardiovascular system to manipulate and retrieve foreign objects, including, but not limited to, wire guides, coils, balloons, catheters, and filters.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K112185 · printed Nov 6, 2026
Cite this page
510(k) Report. “K112185: Foursnare Vascular Retrieval Snare.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K112185
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026