510(k) Report

510(k) · Cleared

Cleared

Aldahol V High Level Disinfectant

K113015510(k) TraditionalCleared Jan 30, 2012111-day reviewClass II21 CFR 880.6885

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Microchem Laboratories, Inc. · MED Sterilant, Medical Devices

Cleared on Jan 30, 2012 as substantially equivalent to K041360 (Healthpoint, Ltd., 2005) after a 111-day review. Its family tree reaches back 1 generation to 2005.

Decided
Jan 30, 2012
Review
111 days
Class
Class II
Product code
MED
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K041360 (2005). No later device cites it as a predicate yet.

K041360 · Aldahol Iii High-Level Disinfectant · Healthpoint, Ltd. · 200520062008201020122014
  1. Root · 1 generation backK041360 Aldahol Iii High-Level DisinfectantHealthpoint, Ltd. · 2005
  2. This deviceK113015 Aldahol V High Level DisinfectantMicrochem Laboratories, Inc. · 2012
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K041360Aldahol Iii High-Level DisinfectantHealthpoint, Ltd.2005—
0 (this device)K113015Aldahol V High Level DisinfectantMicrochem Laboratories, Inc.2012—
FIG. 1 — Lineage of K113015, 2005–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MED? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MED device

  1. #1 K251666 SYSTEM 1E Liquid Chemical Sterilant Processing…score 13.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041360Aldahol Iii High-Level Disinfectant
Summary page 2
…C-1 e. Summary of the substantial eqiuivalence (SE) of Aldahol V High Level Disinfectant to the predicate. Aldahol III High Level Disinifectant (K041360) Aldahol V High Level Disinfectant (HILD) is exactly the same formulation as the predicate, Aldahol III High Level Disinfectant…
Healthpoint, Ltd.2005—

Also cleared against K041360: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • 1 other device also cites its predicate K041360.

How does its category compare?

All MED devices

MED Sterilant, Medical Devices: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · MED

2in 2023

2007: 22008: 12010: 62012: 32013: 12014: 12016: 12017: 12018: 22019: 42021: 22022: 12023: 22026 (partial year): 2
2007high 6 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MED

84in 2019

2010: 281 days median2012: 111 days median2019: 84 days median
2007high 281 (2010) · low 84 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MED

2in 2011

2007: 32008: 22011: 2
2007high 3 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MED

9in 2025

2007: 342008: 792009: 1582010: 892011: 1062012: 3372013: 1402014: 1052015: 1102016: 1352017: 742018: 422019: 352020: 322021: 312022: 162023: 92024: 82025: 9
2007high 337 (2012) · low 8 (2024)2025

MAUDE reports naming product code MED, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—334
20081—279
20090—0158
20106281089
20110—2106
201231110337
20131—0140
20141—0105
20150—0110
20161—0135
20171—074
20182—042
2019484035
20200—032
20212—031
20221—016
20232—09
20240—08
20250—09
20262—0—

What is it indicated for?

Aldahol® V High Level Disinfectant is intended for the high level disinfection of reusable, clean, heat-sensitive, semi-critical medical devices which contact intact mucous membranes when the disinfectant is used at or above its minimum recommended concentration of 1.8% glutaraldehyde for 10.0 min at 20°C. Aldahol V High Level Disinfectant is intended for the sterilization of reusable, clean, heat-sensitive critical and semi-critical medical devices…

Read the full statement (130 words)

Aldahol® V High Level Disinfectant is intended for the high level disinfection of reusable, clean, heat-sensitive, semi-critical medical devices which contact intact mucous membranes when the disinfectant is used at or above its minimum recommended concentration of 1.8% glutaraldehyde for 10.0 min at 20°C.

Aldahol V High Level Disinfectant is intended for the sterilization of reusable, clean, heat-sensitive critical and semi-critical medical devices which contact and potentially penetrate into sterile body areas, for which there is no other practical method of sterilization available with a biological indicator, when the disinfectant is used at or above its minimum recommended concentration of 1.8% glutaraldehyde for 8.0 hrs at 20°C.

Aldahol V High Level Disinfectant may be used in a legally marketed, validated, automatic endoscope reprocessor machine or in a manual bucket system.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113015: Aldahol V High Level Disinfectant.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113015

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026