Recalled · Class II · 2023Unity Toatal Knee System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Corin USA · JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Cleared on Sep 10, 2012 as substantially equivalent to K031729 (Howmedica Osteonics Corp., 2003) and 7 other predicates after a 332-day review. Its family tree reaches back 2 generations to 1992. 8 later devices descend from it. It has been recalled 3 times.
- Decided
- Sep 10, 2012
- Review
- 332 days
- Class
- Class II
- Product code
- JWH
- Recalls
- 3
- Predicates
- 8
Its longest line of descent runs back 2 generations to K915192 (1992). 13 ancestors in all. 8 later devices descend from it, 6 citing it directly.
- Root · 2 generations backK915192 7000 Series Modular Tibial ComponentsOsteonics Corp. · 1992
- 1 generation backK994128 Scorpio Total Stablizer (TS) Total Knee SystemHowmedica Osteonics Corp. · 2000 · Recalled
- This deviceK113060 Unity Toatal Knee SystemCorin USA · 2012
- Descendant · cites it directly · 2 of its ownK173884 Unity Total Knee SystemCorin U.S.A. Limited · 2018
- Descendant · 2 generations on · 1 of its ownK234044 ACS® LD FB Knee SystemImplantcast GmbH · 2024
- Descendant · cites it directly · 1 of its ownK183533 Unity Total Knee SystemCorin U.S.A. Limited · 2019
- Descendant · cites it directly · 0 of its ownK261275 Unity Total Knee System, Unity Knee Cementless CR FemurCorin, Ltd. · 2026
- Descendant · 3 generations on · 0 of its ownK260037 implantcast Packaging System UpdateImplantcast GmbH · 2026
Showing 5 of 8 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 8 · See all- K173884Unity Total Knee SystemCorin U.S.A. Limited · 2018 · cites it directly2descendants
- K234044ACS® LD FB Knee SystemImplantcast GmbH · 2024 · 2 generations on1descendant
- K183533Unity Total Knee SystemCorin U.S.A. Limited · 2019 · cites it directly1descendant
- K261275Unity Total Knee System, Unity Knee Cementless CR FemurCorin, Ltd. · 2026 · cites it directly0descendants
- K260037implantcast Packaging System UpdateImplantcast GmbH · 2026 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K915192 | 7000 Series Modular Tibial Components | Osteonics Corp. | 1992 | — |
| −1 | K994128 | Scorpio Total Stablizer (TS) Total Knee System | Howmedica Osteonics Corp. | 2000 | 8 |
| 0 (this device) | K113060 | Unity Toatal Knee System | Corin USA | 2012 | — |
| +1 | K173884 | Unity Total Knee System | Corin U.S.A. Limited | 2018 | — |
| +2 | K234044 | ACS® LD FB Knee System | Implantcast GmbH | 2024 | — |
| +1 | K183533 | Unity Total Knee System | Corin U.S.A. Limited | 2019 | — |
| +1 | K261275 | Unity Total Knee System, Unity Knee Cementless CR Femur | Corin, Ltd. | 2026 | — |
| +3 | K260037 | implantcast Packaging System Update | Implantcast GmbH | 2026 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1996 | 2 | 0 |
| 1998 | 2 | 0 |
| 2003 | 1 | 0 |
| 2004 | 2 | 0 |
| 2005 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2017 | 1 | 0 |
| 2018 | 1 | 0 |
| 2019 | 1 | 0 |
| 2024 | 3 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JWH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JWH device
- #1 K221479 Persona Personalized Knee Systemscore 22.4 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (8)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K031729 | Triathlon Posteriorly Stabilized (PS) Total Knee SystemSummary page 1…metal/polymer semi- constrained cemented prosthesis (21CFR888.3560) 7. Legally Marketed Devices to which Substantial Equivalence is claimed: Triathlon PS TKA System (KK031729) Triathlon CR TKA System (K040267) Triathlon CR sz 1 Tibial Insert (IK042883) Triathlon PS sz 1 Tibial Insert (K050539) Scorpio C… | 2003 |
| * | K040267 | Triathlon Cruciate Retaining (CR) Total KneeSummary page 1…emented prosthesis (21CFR888.3560) 7. Legally Marketed Devices to which Substantial Equivalence is claimed: Triathlon PS TKA System (KK031729) Triathlon CR TKA System (K040267) Triathlon CR sz 1 Tibial Insert (IK042883) Triathlon PS sz 1 Tibial Insert (K050539) Scorpio CR TKA System (K974556) Scorpio PS… | 2004 |
| * | K042883 | Size 1 Triathlon Cruciate Retaining (CR) Tibial InsertSummary page 1…lly Marketed Devices to which Substantial Equivalence is claimed: Triathlon PS TKA System (KK031729) Triathlon CR TKA System (K040267) Triathlon CR sz 1 Tibial Insert (IK042883) Triathlon PS sz 1 Tibial Insert (K050539) Scorpio CR TKA System (K974556) Scorpio PS TKA System (1K962152) Scorpio NRG System (K… | 2004 |
| * | K050539 | Size 1 Triathlon Posterior Stabilized (PS) Tibial InsertSummary page 1…Equivalence is claimed: Triathlon PS TKA System (KK031729) Triathlon CR TKA System (K040267) Triathlon CR sz 1 Tibial Insert (IK042883) Triathlon PS sz 1 Tibial Insert (K050539) Scorpio CR TKA System (K974556) Scorpio PS TKA System (1K962152) Scorpio NRG System (K042343) Scorpio TS (K994128) 8. Device D… | 2005 |
| * | K974556 | Osteonics Scorpio Posterior Cruciate Retaining Total Knee SystemSummary page 1…on PS TKA System (KK031729) Triathlon CR TKA System (K040267) Triathlon CR sz 1 Tibial Insert (IK042883) Triathlon PS sz 1 Tibial Insert (K050539) Scorpio CR TKA System (K974556) Scorpio PS TKA System (1K962152) Scorpio NRG System (K042343) Scorpio TS (K994128) 8. Device Description: The Corin Unity Kne… | 1998 |
| * | K962152 | Osteonics Scorpio Posteriorly Stabilized Total Knee SystemSummary page 1…hlon CR TKA System (K040267) Triathlon CR sz 1 Tibial Insert (IK042883) Triathlon PS sz 1 Tibial Insert (K050539) Scorpio CR TKA System (K974556) Scorpio PS TKA System (1K962152) Scorpio NRG System (K042343) Scorpio TS (K994128) 8. Device Description: The Corin Unity Knee System is a fixed bearing total… | 1996 |
| * | K042343 | Scorpio NRG Knee SystemSummary page 1…Triathlon CR sz 1 Tibial Insert (IK042883) Triathlon PS sz 1 Tibial Insert (K050539) Scorpio CR TKA System (K974556) Scorpio PS TKA System (1K962152) Scorpio NRG System (K042343) Scorpio TS (K994128) 8. Device Description: The Corin Unity Knee System is a fixed bearing total knee replacement system that… | 2005 |
| * | K994128 | Scorpio Total Stablizer (TS) Total Knee SystemSummary page 1…al Insert (IK042883) Triathlon PS sz 1 Tibial Insert (K050539) Scorpio CR TKA System (K974556) Scorpio PS TKA System (1K962152) Scorpio NRG System (K042343) Scorpio TS (K994128) 8. Device Description: The Corin Unity Knee System is a fixed bearing total knee replacement system that consists of a femora… | 2000 |
What descends from it? (8)
6 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 13, 2023 | Class II | Z-0213-2024 | Process change control | Potential for Unity CR Inserts Right size 6 from batch 529803 to be incorrectly labelled as Unity CR Inserts Right size 7 from batch 532405 and visa versa. |
| Oct 13, 2023 | Class II | Z-0214-2024 | Process change control | Potential for Unity CR Inserts Right size 6 from batch 529803 to be incorrectly labelled as Unity CR Inserts Right size 7 from batch 532405 and visa versa. |
| Oct 31, 2022 | Class II | Z-0273-2023 | Under Investigation by firm | The internal packaging system for devices may have damage to the blisters potentially damaging the device or compromising the sterility of the packaging. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 20 years and passes through 3 companies; a new generation was cleared every 12.5 years on median.
- 8 of the 13 ancestors have a recall of any class.
- Of its descendants, K173884 (Corin U.S.A. Limited, 2018) has gone furthest, with 2 descendants of its own.
- 14 other devices also cite its predicate K031729.
How does its category compare?
All JWH devicesJWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.
FDA decisions per year · JWH
20in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JWH
55in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JWH
2in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JWH
9,306in 2025
MAUDE reports naming product code JWH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 18 | 74 | 14 | 1,305 |
| 2008 | 17 | 81 | 10 | 1,483 |
| 2009 | 11 | 168 | 109 | 1,870 |
| 2010 | 19 | 85 | 32 | 3,286 |
| 2011 | 25 | 87 | 27 | 5,133 |
| 2012 | 33 | 85 | 85 | 5,203 |
| 2013 | 31 | 128 | 31 | 6,135 |
| 2014 | 33 | 98 | 13 | 6,722 |
| 2015 | 27 | 87 | 11 | 6,881 |
| 2016 | 18 | 88 | 17 | 7,586 |
| 2017 | 25 | 118 | 61 | 12,263 |
| 2018 | 25 | 90 | 14 | 14,071 |
| 2019 | 28 | 133 | 5 | 14,987 |
| 2020 | 22 | 59 | 2 | 12,592 |
| 2021 | 24 | 61 | 8 | 9,408 |
| 2022 | 17 | 56 | 12 | 8,364 |
| 2023 | 20 | 57 | 67 | 8,947 |
| 2024 | 19 | 89 | 7 | 10,828 |
| 2025 | 20 | 55 | 2 | 9,306 |
| 2026 | 12 | 63 | 4 | — |
What is it indicated for?
The Unity Total Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is evidence of sufficient sound bone to seat and support the components. General total knee arthroplasty indications include: Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or…
Read the full statement (162 words)
The Unity Total Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is evidence of sufficient sound bone to seat and support the components.
General total knee arthroplasty indications include:
Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis
Post-traumatic loss of knee joint configuration and function
Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function
Revision of previous unsuccessful knee replacement or other procedure, where soft tissue stability is adequate
Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques
The posterior stabilized variant is also indicated for PCL instability requiring implant bearing surface geometries with increased anterior-postetior constraint and absent or non-functioning posterior cruciate ligament
The Unity Total Knee System is indicated for cemented, single use only.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K113060 · printed Nov 6, 2026
Cite this page
510(k) Report. “K113060: Unity Toatal Knee System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113060
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026