510(k) Report

510(k) · Cleared

Recalled · 2011

Celltracks Analyer II

K113181510(k) SpecialCleared Dec 12, 201145-day reviewClass II21 CFR 866.6020Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Veridex, LLC · NQI System, Immunomagnetic, Circulating Cancer Cell, Enumeration

Cleared on Dec 12, 2011 as substantially equivalent to K060110 (Immunicon Corp., 2006) after a 45-day review. Its family tree reaches back 3 generations to 2004. 1 later device descends from it. It has been recalled once.

Decided
Dec 12, 2011
Review
45 days
Class
Class II
Product code
NQI
Recalls
1
Predicates
1

Its longest line of descent runs back 3 generations to DEN040001 (2004). 1 later device descends from it, 1 citing it directly.

  1. Root · 3 generations backDEN040001 Cellsearch Epithelial Cell Enrichment Kit; Cellspotter SystemAdvanced Diagnostic Systems · 2004
  2. 2 generations backK050145 Celltracka Analyzer IIImmunicon Corp. · 2005 · Recalled
  3. 1 generation backK060110 Celltracks Analyzer IIImmunicon Corp. · 2006 · Recalled
  4. This deviceK113181 Celltracks Analyer IIVeridex, LLC · 2011
  5. Descendant · cites it directly · 0 of its ownK130794 Celltracks Analyzer II SystemVeridex, LLC · 2013 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K130794Celltracks Analyzer II SystemVeridex, LLC · 2013 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3DEN040001Cellsearch Epithelial Cell Enrichment Kit; Cellspotter SystemAdvanced Diagnostic Systems2004—
−2K050145Celltracka Analyzer IIImmunicon Corp.20052
−1K060110Celltracks Analyzer IIImmunicon Corp.20061
0 (this device)K113181Celltracks Analyer IIVeridex, LLC2011—
+1K130794Celltracks Analyzer II SystemVeridex, LLC20131

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
FIG. 1 — Lineage of K113181, 2004–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NQI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060110Celltracks Analyzer II
Summary page 1
…Product Code NQI) 4. Predicate device The predicate device is the CellTracks Analyzer HI® (K0601 10, March 16, 2006) 510(k) Summary, Continued 5. Device Description The CELLTRACKS ANALYZER II@ is a…
Immunicon Corp.20061

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K130794Celltracks Analyzer II SystemVeridex, LLC20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 17, 2011Not classifiedZ-1004-2012Software designControl cell images may i intermix with Patient cell images while changing between the Control review screen and Patient review screen on the Cell tracks Analyzer II.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 3 companies; a new generation was cleared every 1.1 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K130794 (Veridex, LLC, 2013) has gone furthest, with 0 descendants of its own.

How does its category compare?

All NQI devices

NQI System, Immunomagnetic, Circulating Cancer Cell, Enumeration: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · NQI

1in 2013

2007: 12008: 12010: 12011: 12012: 22013: 1
2007high 2 (2012) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NQI

Too few NQI 510(k)s in any year since 2007 to give a median.

Recalls per year · NQI

1in 2011

2007: 12009: 12011: 1
2007high 1 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NQI

3in 2008

2008: 3
2007high 3 (2008) · low 0 (2007)2025

MAUDE reports naming product code NQI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—10
20081—03
20090—10
20101—00
20111—10
20122—00
20131—00
20140—00
20150—00
20160—00
20170—00
20180—00
20190—00
20200—00
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

The CELLTRACKS ANALYZER II is a semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immunomagnetically selected and aligned. This product is for in vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113181: Celltracks Analyer II.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113181

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026