Nmi Control Syringe, 7 ML Nmi Control Syringe, 8 ML

K113198510(k) SpecialCleared Nov 28, 201128-day reviewClass II21 CFR 870.1650Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Navilyst Medical, Inc. · DXT Injector And Syringe, Angiographic

Cleared on Nov 28, 2011 as substantially equivalent to K873955 (North American Instrument Corp., 1988) after a 28-day review. Its family tree reaches back 1 generation to 1988. 2 of the 9 devices in its family, counting itself, have a Class I recall. 7 later devices descend from it. It has been recalled once, including a Class I recall.

Decided
Nov 28, 2011
Review
28 days
Class
Class II
Product code
DXT
Recalls
1 (1 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K873955 (1988). 7 later devices descend from it, 2 citing it directly. 2 of the 9 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2014: 1 descendant cleared2017: 1 descendant cleared2019: 1 descendant cleared2024: 1 descendant cleared2021: 1 descendant cleared2022: 1 descendant cleared2025: 1 descendant clearedK873955 · Namic Angiographic Syringe · North American Instrument Corp. · 1988 · Class I recall1990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant K873955 · Namic Angiographic Syringe · North American Instrument Corp. · 1988 · Class I recall0
  1. Root · 1 generation backK873955 Namic Angiographic SyringeNorth American Instrument Corp. · 1988 · Recalled (Class I)
  2. This deviceK113198 Nmi Control Syringe, 7 ML Nmi Control Syringe, 8 MLNavilyst Medical, Inc. · 2011
  3. Descendant · cites it directly · 4 of its ownK140194 Namic RCS SyringeNavilyst Medical, Inc. · 2014
  4. Descendant · 2 generations on · 3 of its ownK191659 Instylla Delivery KitInstylla, Inc. · 2019
  5. Descendant · 3 generations on · 2 of its ownK202544 Instylla Delivery KitInstylla, Inc. · 2021
  6. Descendant · 3 generations on · 1 of its ownK213632 Instylla Delivery KitInstylla, Inc. · 2022
  7. Descendant · cites it directly · 1 of its ownK163084 Merit Coronary Control SyringeMerit Medical Systems, Inc. · 2017

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2014: 12017: 12019: 12021: 12022: 12024: 12025: 1
2014high 1 (2014) · low 0 (2015)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 7 · See all
  • K140194Namic RCS SyringeNavilyst Medical, Inc. · 2014 · cites it directly4descendants
  • K191659Instylla Delivery KitInstylla, Inc. · 2019 · 2 generations on3descendants
  • K202544Instylla Delivery KitInstylla, Inc. · 2021 · 3 generations on2descendants
  • K213632Instylla Delivery KitInstylla, Inc. · 2022 · 3 generations on1descendant
  • K163084Merit Coronary Control SyringeMerit Medical Systems, Inc. · 2017 · cites it directly1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K873955Namic Angiographic SyringeNorth American Instrument Corp.1988Class I
0 (this device)K113198Nmi Control Syringe, 7 ML Nmi Control Syringe, 8 MLNavilyst Medical, Inc.2011—
+1K140194Namic RCS SyringeNavilyst Medical, Inc.2014—
+2K191659Instylla Delivery KitInstylla, Inc.2019—
+3K202544Instylla Delivery KitInstylla, Inc.2021—
+3K213632Instylla Delivery KitInstylla, Inc.2022—
+1K163084Merit Coronary Control SyringeMerit Medical Systems, Inc.2017—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201710
201910
202110
202210
202410
202510
FIG. 1 — Lineage of K113198, 1988–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXT device

  1. #1 K253048 High Pressure Syringe; Pressure Connecting…score 13.6 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K873955Namic Angiographic Syringe
Summary page 4
…ntrol Syringe Piston Syringe 21 CFR §880.5860, Class Il General Hospital NAMIC Angiographic Control Syringe Control Syringe Piston Syringe 21 CFR §880.5860, Class 1} K873955 The NMI Control Syringes are manual control, piston type syringes consisting of a clear, calibrated, hollow barrel accommodating…
North American Instrument Corp.1988Class I

Also cleared against K873955: 1 other device →

What descends from it? (7)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K140194Namic RCS SyringeNavilyst Medical, Inc.20144
K163084Merit Coronary Control SyringeMerit Medical Systems, Inc.20171

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 27, 2026Class IZ-1713-2026Under Investigation by firmMedline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K873955Namic Angiographic SyringeNorth American Instrument Corp. · 1988 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • Its one ancestor has a recall, Class I.
  • Of its descendants, K140194 (Navilyst Medical, Inc., 2014) has gone furthest, with 4 descendants of its own.
  • 1 other device also cites its predicate K873955.

How does its category compare?

All DXT devices

DXT Injector And Syringe, Angiographic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · DXT

1in 2025

2007: 32008: 22009: 62010: 12011: 42013: 22014: 62015: 72016: 82017: 72018: 92019: 42020: 32021: 12023: 12024: 42025: 12026 (partial year): 7
2007high 9 (2018) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXT

138in 2024

2007: 182 days median2009: 137 days median2011: 69 days median2014: 144 days median2015: 153 days median2016: 172 days median2017: 183 days median2018: 93 days median2019: 96 days median2020: 154 days median2024: 138 days median2026 (partial year): 264 days median
2007high 264 (2026) · low 69 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXT

10in 2024

2007: 42008: 32009: 72010: 32011: 32012: 72015: 12018: 22020: 12021: 12023: 22024: 102026 (partial year): 6
2007high 10 (2024) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXT

97in 2025

2007: 1072008: 1682009: 1882010: 1242011: 1332012: 1352013: 1412014: 972015: 1352016: 1522017: 1282018: 1292019: 1212020: 932021: 1322022: 1112023: 1172024: 1182025: 97
2007high 188 (2009) · low 93 (2020)2025

MAUDE reports naming product code DXT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731824107
20082—3168
200961377188
20101—3124
20114693133
20120—7135
20132—0141
20146144097
201571531135
201681720152
201771830128
20189932129
20194960121
20203154193
20211—1132
20220—0111
20231—2117
2024413810118
20251—097
202672646—

What is it indicated for?

The NMI Control Syringe is intended to be used for the intra-arterial or intravenous administration of radiographic contrast media.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K113198: Nmi Control Syringe, 7 ML Nmi Control Syringe, 8 ML.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113198

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026