510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Lasso 2515 Nav Eco Variable Catheter Lasso Nav Eco Catheter Continued From Line 2

K113213510(k) SpecialCleared Dec 5, 201134-day reviewClass II21 CFR 870.1220Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biosense Webster, Inc. · DRF Catheter, Electrode Recording, Or Probe, Electrode Recording

Cleared on Dec 5, 2011 as substantially equivalent to K081258 (Biosense Webster, Inc., 2009) and 1 other predicate after a 34-day review. Its family tree reaches back 4 generations to 2000. 4 later devices descend from it. It has been recalled once.

Decided
Dec 5, 2011
Review
34 days
Class
Class II
Product code
DRF
Recalls
1
Predicates
2

Its longest line of descent runs back 4 generations to K002333 (2000). 5 ancestors in all, the oldest K954390 (1995). 4 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor2013: 1 descendant cleared2022: 1 descendant cleared2016: 2 descendants clearedK002333 · Biosense Webster Lasso Deflectable Circular Mapping Catheter, Model D-1220 · Biosense Webster, Inc. · 2000 · recalledK031161 · Variable Lasso 2515 Circular Mapping Catheter · Biosense Webster, Inc. · 2003 · recalledK081258 · Variable Lasso Nav Catheter, Models: D-1290-01, D-1290-02 · Biosense Webster, Inc. · 2009K093376 · Lasso Nav Catheter · Biosense Webster, Inc. · 20101995200020052010201520202025
  1. Root · 4 generations backK002333 Biosense Webster Lasso Deflectable Circular Mapping Catheter, Model…Biosense Webster, Inc. · 2000 · Recalled
  2. 3 generations backK031161 Variable Lasso 2515 Circular Mapping CatheterBiosense Webster, Inc. · 2003 · Recalled
  3. 2 generations backK081258 Variable Lasso Nav Catheter, Models: D-1290-01, D-1290-02Biosense Webster, Inc. · 2009
  4. 1 generation backK093376 Lasso Nav CatheterBiosense Webster, Inc. · 2010
  5. This deviceK113213 Lasso 2515 Nav Eco Variable Catheter Lasso Nav Eco Catheter…Biosense Webster, Inc. · 2011
  6. Descendant · cites it directly · 0 of its ownK211327 Khelix Steerable Electrophysiology Catheters, Khelix Loop Fixed…CathRx, Ltd. · 2022
  7. Descendant · cites it directly · 0 of its ownK153006 Reprocessed Lasso NAV eco and Lasso 2515 NAV eco Variable…Sterilmed, Inc. · 2016
  8. Descendant · cites it directly · 0 of its ownK153153 Reprocessed Electrophysiology CatheterInnovative Health, LLC · 2016
  9. Descendant · cites it directly · 0 of its ownK130602 Lasso Nav Duo Loop CatheterBiosense Webster, Inc. · 2013 · Recalled

Most influential descendants

Showing 4 of 4 · See all
  • K211327Khelix Steerable Electrophysiology Catheters, Khelix Loop Fixed Electrophysiology…CathRx, Ltd. · 2022 · cites it directly0descendants
  • K153006Reprocessed Lasso NAV eco and Lasso 2515 NAV eco Variable Electrophysiology (EP) CatheterSterilmed, Inc. · 2016 · cites it directly0descendants
  • K153153Reprocessed Electrophysiology CatheterInnovative Health, LLC · 2016 · cites it directly0descendants
  • K130602Lasso Nav Duo Loop CatheterBiosense Webster, Inc. · 2013 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K002333Biosense Webster Lasso Deflectable Circular Mapping Catheter, Model D-1220Biosense Webster, Inc.20002
−3K031161Variable Lasso 2515 Circular Mapping CatheterBiosense Webster, Inc.20032
−2K081258Variable Lasso Nav Catheter, Models: D-1290-01, D-1290-02Biosense Webster, Inc.2009—
−1K093376Lasso Nav CatheterBiosense Webster, Inc.2010—
0 (this device)K113213Lasso 2515 Nav Eco Variable Catheter Lasso Nav Eco Catheter Continued From…Biosense Webster, Inc.2011—
+1K211327Khelix Steerable Electrophysiology Catheters, Khelix Loop Fixed…CathRx, Ltd.2022—
+1K153006Reprocessed Lasso NAV eco and Lasso 2515 NAV eco Variable Electrophysiology…Sterilmed, Inc.2016—
+1K153153Reprocessed Electrophysiology CatheterInnovative Health, LLC2016—
+1K130602Lasso Nav Duo Loop CatheterBiosense Webster, Inc.20131

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201620
202210
FIG. 1 — Lineage of K113213, 2000–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DRF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DRF device

  1. #1 K252794 Vision-MR™ Diagnostic Catheterscore 13.6 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K081258Variable Lasso Nav Catheter, Models: D-1290-01, D-1290-02
Summary page 1
…(21 CER 870.1220, Product Code DRF) Predicate Devices: LASSO® 2515 NAV Variable Catheter 510(k): K081258 LASSO®NAV Catheter 510(k): K093376 Manufacturer:…
Biosense Webster, Inc.2009—
*K093376Lasso Nav Catheter
Summary page 1
…Catheter 510(k): K081258 LASSO®NAV Catheter 510(k): K093376 Manufacturer: Biosense Webster, Inc. 3333 Diamond Canyon Road…
Biosense Webster, Inc.2010—

Also cleared against K081258: 3 other devices →

What descends from it? (4)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K211327Khelix Steerable Electrophysiology Catheters, Khelix Loop Fixed Electrophysiology…CathRx, Ltd.20220
K153006Reprocessed Lasso NAV eco and Lasso 2515 NAV eco Variable Electrophysiology (EP) CatheterSterilmed, Inc.20160
K153153Reprocessed Electrophysiology CatheterInnovative Health, LLC20160
K130602Lasso Nav Duo Loop CatheterBiosense Webster, Inc.20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 19, 2013Class IIZ-1617-2013Use errorBiosense Webster is recalling the 20-Pole LASSO NAV Catheter Eco because it has been incorrectly calibrated causing them to be recognized and displayed as a 10-Pole LASSO NAV Catheter Eco by the CARTO 3 System.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 1 company; a new generation was cleared every 2.9 years on median.
  • 2 of the 5 ancestors have a recall of any class.
  • Of its descendants, K211327 (CathRx, Ltd., 2022) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K081258.

How does its category compare?

All DRF devices

DRF Catheter, Electrode Recording, Or Probe, Electrode Recording: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · DRF

3in 2025

2007: 22008: 22009: 72010: 52011: 32012: 12013: 22014: 12015: 32016: 82017: 72018: 12019: 102020: 42021: 122022: 82023: 52024: 42025: 32026 (partial year): 2
2007high 12 (2021) · low 1 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DRF

45in 2023

2009: 243 days median2010: 65 days median2011: 190 days median2015: 29 days median2016: 175 days median2019: 163 days median2021: 124 days median2022: 283 days median2023: 45 days median
2007high 283 (2022) · low 29 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DRF

1in 2015

2007: 22008: 42011: 62014: 22015: 1
2007high 6 (2011) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DRF

508in 2025

2007: 1202008: 1762009: 1222010: 1302011: 2452012: 4022013: 3102014: 3492015: 3272016: 3832017: 4582018: 7232019: 6692020: 6512021: 6252022: 6132023: 6852024: 6232025: 508
2007high 723 (2018) · low 120 (2007)2025

MAUDE reports naming product code DRF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—2120
20082—4176
200972430122
20105650130
201131906245
20121—0402
20132—0310
20141—2349
20153291327
201681750383
20177—0458
20181—0723
2019101630669
20204—0651
2021121240625
202282830613
20235450685
20244—0623
20253—0508
20262—0—

What is it indicated for?

LASSO® 2515 NAV eco Variable Catheter The LASSO® 2515 NAV eco Variable Catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The catheter is designed to obtain electrograms in the atrial regions of the heart. The LASSO® 2515 NAV eco Variable Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not…

Read the full statement (154 words)

LASSO® 2515 NAV eco Variable Catheter

The LASSO® 2515 NAV eco Variable Catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The catheter is designed to obtain electrograms in the atrial regions of the heart.

The LASSO® 2515 NAV eco Variable Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 2.3.)

LASSO® NAV eco Catheter

The LASSO® NAV eco Catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The catheter is designed to obtain electrograms in the atrial regions of the heart.

The LASSO® NAV eco Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 2.3.)

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113213: Lasso 2515 Nav Eco Variable Catheter Lasso Nav Eco Catheter Continued From Line 2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113213

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026