510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Hoffmann LRF (Limb Reconstruction Frame) System

K113327510(k) TraditionalCleared Mar 29, 2012140-day reviewClass II21 CFR 888.3030Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Stryker Corp. · KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

Cleared on Mar 29, 2012 as substantially equivalent to K031181 (Smith & Nephew, Inc., 2003) and 2 other predicates after a 140-day review. Its family tree reaches back 2 generations to 2000. 17 later devices descend from it. It has been recalled 3 times.

Decided
Mar 29, 2012
Review
140 days
Class
Class II
Product code
KTT
Recalls
3
Predicates
3

Its longest line of descent runs back 2 generations to K994143 (2000). 4 ancestors in all. 17 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2000: 1 other ancestor2014: 1 descendant cleared2013: 2 descendants cleared2016: 2 descendants cleared2017: 1 descendant cleared2020: 1 descendant cleared2019: 1 descendant cleared2021: 2 descendants cleared2023: 2 descendants cleared2025: 4 descendants cleared2026: 1 descendant clearedK994143 · Smith & Nephew External Fixation System [Unilateral (Linear) & Multilateral (Circular) Fixators and Accessories] · Smith & Nephew, Inc. · 2000 · recalledK031181 · External Fixation Systems · Smith & Nephew, Inc. · 2003200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 3 descendants2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 2 descendants5 generations forward: 1 descendant6 generations forward: 1 descendant7 generations forward: 2 descendants8 generations forward: 3 descendants9 generations forward: 2 descendants10 generations forward: 1 descendant K031181 · External Fixation Systems · Smith & Nephew, Inc. · 2003K994143 · Smith & Nephew External Fixation System [Unilateral (Linear) & Multilateral (Circular) Fixators and Accessories] · Smith & Nephew, Inc. · 2000 · recalled0+5+10
  1. Root · 2 generations backK994143 Smith & Nephew External Fixation System [Unilateral (Linear) &…Smith & Nephew, Inc. · 2000 · Recalled
  2. 1 generation backK031181 External Fixation SystemsSmith & Nephew, Inc. · 2003
  3. This deviceK113327 Hoffmann LRF (Limb Reconstruction Frame) SystemStryker Corp. · 2012
  4. Descendant · cites it directly · 15 of its ownK130907 Hoffmann LRF (Limb Reconstruction Frame) SystemStryker Trauma AG · 2013
  5. Descendant · cites it directly · 15 of its ownK130334 Hoffmann LRF (Limb Reconstruction Frame System)Stryker Trauma AG · 2013
  6. Descendant · cites it directly · 14 of its ownK140961 Hoffmann LRF (Limb Reconstruction Frame) SystemStryker Trauma AG · 2014
  7. Descendant · 2 generations on · 13 of its ownK153377 Hoffmann LRF (Limb Reconstruction Frame) SystemStryker GmbH · 2016 · Recalled
  8. Descendant · 3 generations on · 12 of its ownK161753 Hoffman LRF SystemStryker GmbH · 2016

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

2013: 22014: 12016: 22017: 12019: 12020: 12021: 22023: 22025: 42026: 1
2013high 4 (2025) · low 0 (2015)2026

Most influential descendants

Showing 5 of 17 · See all
  • K130907Hoffmann LRF (Limb Reconstruction Frame) SystemStryker Trauma AG · 2013 · cites it directly15descendants
  • K130334Hoffmann LRF (Limb Reconstruction Frame System)Stryker Trauma AG · 2013 · cites it directly15descendants
  • K140961Hoffmann LRF (Limb Reconstruction Frame) SystemStryker Trauma AG · 2014 · cites it directly14descendants
  • K153377Hoffmann LRF (Limb Reconstruction Frame) SystemStryker GmbH · 2016 · 2 generations on13descendants
  • K161753Hoffman LRF SystemStryker GmbH · 2016 · 3 generations on12descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K994143Smith & Nephew External Fixation System [Unilateral (Linear) & Multilateral…Smith & Nephew, Inc.200019
−1K031181External Fixation SystemsSmith & Nephew, Inc.2003—
0 (this device)K113327Hoffmann LRF (Limb Reconstruction Frame) SystemStryker Corp.2012—
+1K130907Hoffmann LRF (Limb Reconstruction Frame) SystemStryker Trauma AG2013—
+1K130334Hoffmann LRF (Limb Reconstruction Frame System)Stryker Trauma AG2013—
+1K140961Hoffmann LRF (Limb Reconstruction Frame) SystemStryker Trauma AG2014—
+2K153377Hoffmann LRF (Limb Reconstruction Frame) SystemStryker GmbH20162
+3K161753Hoffman LRF SystemStryker GmbH2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201320
201410
201620
201710
201910
202010
202120
202320
202540
202610
FIG. 1 — Lineage of K113327, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KTT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KTT device

  1. #1 K242861 TrueLok Elevatescore 21.3 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031181External Fixation Systems
Summary page 2
…parison showed that the proposed device is substantially equivalent in intended, use, materials and performance characteristics to the following predicate devices: " K03 1181 External Fixation Systems (Smith & Nephew) * K04 1706 Hoffman 11 Foot Ring * K003018 Titan Tibial Nail (T2 Nail) Non-Clini…
Smith & Nephew, Inc.2003—
*K041706Hoffmann II Foot Ring
Summary page 2
…equivalent in intended, use, materials and performance characteristics to the following predicate devices: " K03 1181 External Fixation Systems (Smith & Nephew) * K04 1706 Hoffman 11 Foot Ring * K003018 Titan Tibial Nail (T2 Nail) Non-Clinical Testing: Non-clinical laboratory testing was performe…
Howmedica Osteonics Corp.2004—
*K003018Titan Tibial Nail
Summary page 2
…als and performance characteristics to the following predicate devices: " K03 1181 External Fixation Systems (Smith & Nephew) * K04 1706 Hoffman 11 Foot Ring * K003018 Titan Tibial Nail (T2 Nail) Non-Clinical Testing: Non-clinical laboratory testing was performed for the Hoffmann LRF (Limb Recon…
Howmedica Osteonics Corp.200022

Also cleared against K031181: 8 other devices →

What descends from it? (17)

See all 17

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K130907Hoffmann LRF (Limb Reconstruction Frame) SystemStryker Trauma AG201315
K130334Hoffmann LRF (Limb Reconstruction Frame System)Stryker Trauma AG201315
K140961Hoffmann LRF (Limb Reconstruction Frame) SystemStryker Trauma AG201414

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 12, 2018Class IIZ-1526-2018Labeling mix-upsThe wrong parts were picked for packaging. The connecting Nuts were packed in the packaging labeled with Hinge coupling
Oct 25, 2012Class IIZ-0695-2013Device DesignTwo complaints have been filed where Hoffman LRF Wires have broken during load bearing application by patients.
Jul 3, 2012Class IIZ-0266-2013Device DesignThree complaints have been filed where the Hoffman LRF Telescopic Struts have broken during load bearing application by patients having a body weight greater than 250 LBS.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 2 companies; a new generation was cleared every 8.7 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • Of its descendants, K130907 (Stryker Trauma AG, 2013) has gone furthest, with 15 descendants of its own.
  • 8 other devices also cite its predicate K031181.

How does its category compare?

All KTT devices

KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · KTT

10in 2025

2007: 92008: 112009: 112010: 122011: 102012: 92013: 72014: 142015: 122016: 102017: 112018: 62019: 82020: 92021: 102022: 32023: 72024: 52025: 102026 (partial year): 5
2007high 14 (2014) · low 3 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KTT

88in 2025

2007: 55 days median2008: 73 days median2009: 94 days median2010: 148 days median2011: 157 days median2012: 140 days median2013: 87 days median2014: 129 days median2015: 60 days median2016: 142 days median2017: 184 days median2018: 149 days median2019: 110 days median2020: 200 days median2021: 83 days median2022: 25 days median2023: 29 days median2024: 90 days median2025: 88 days median2026 (partial year): 68 days median
2007high 200 (2020) · low 25 (2022)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KTT

2in 2021

2007: 22008: 12009: 22010: 12012: 42013: 72017: 12018: 142019: 32021: 22026 (partial year): 11
2007high 14 (2018) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KTT

673in 2025

2007: 452008: 662009: 452010: 1212011: 1962012: 5272013: 8982014: 4452015: 4012016: 4002017: 4852018: 5972019: 8002020: 9422021: 9612022: 6272023: 5542024: 5232025: 673
2007high 961 (2021) · low 45 (2007)2025

MAUDE reports naming product code KTT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007955245
20081173166
20091194245
2010121481121
2011101570196
201291404527
20137877898
2014141290445
201512600401
2016101420400
2017111841485
2018614914597
201981103800
202092000942
202110832961
20223250627
20237290554
20245900523
202510880673
202656811—

What is it indicated for?

The Stryker Hoffmann LRF (Limb Reconstruction Frame) System is indicated in pediatric patients and adults for the treatment and fixation of: Open and Closed Fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bo~nes Limb lengthening by epiphyseal or…

Read the full statement (131 words)

The Stryker Hoffmann LRF (Limb Reconstruction Frame) System is indicated in pediatric patients and adults for the treatment and fixation of:

Open and Closed Fractures

Post-traumatic joint contracture which has resulted in loss of range of motion

Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction

Pseudoarthrosis or non-union of long bo~nes

Limb lengthening by epiphyseal or metaphyseal distraction

Correction of bony or soft tissue deformity

Correction of segmental bony or soft tissue defects

Joint arthrodesis

Management of comminuted intra-articular fractures of the distal radius

The Stryker Hoffmann LRF System is indicated in adults for:

Osteotomy

Revision procedure where other treatments or devices have been unsuccessful

Bone reconstruction procedures

Fusions and replantations of the foot

Charcot foot reconstruction

Lisfiranc dislocations

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113327: Hoffmann LRF (Limb Reconstruction Frame) System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113327

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026