510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Non-Mydriatic Auto Fundus Camera Afc-330 with Image Filing Software NA Vis-EX

K113451510(k) TraditionalCleared May 8, 2012169-day reviewClass II21 CFR 886.1120Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nidek Technologies Srl · HKI Camera, Ophthalmic, Ac-Powered

Cleared on May 8, 2012 as substantially equivalent to K014274 (Nidek, Inc., 2002) and 1 other predicate after a 169-day review. Its family tree reaches back 3 generations to 1995. 1 later device descends from it. It has been recalled 2 times.

Decided
May 8, 2012
Review
169 days
Class
Class II
Product code
HKI
Recalls
2
Predicates
3

Its longest line of descent runs back 3 generations to K941234 (1995). 4 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor2021: 1 descendant clearedK941234 · Non-Mydriatic Retinal Camera · Canon USA, Inc. · 1995K980246 · Non-Mydriatic Retinal Camera, Model CR6-45NM · Canon USA, Inc. · 1998K014274 · Non-Mydriatic Fundus Camera, Model NM-1000 · Nidek, Inc. · 20021995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant K014274 · Non-Mydriatic Fundus Camera, Model NM-1000 · Nidek, Inc. · 2002K980246 · Non-Mydriatic Retinal Camera, Model CR6-45NM · Canon USA, Inc. · 1998K941234 · Non-Mydriatic Retinal Camera · Canon USA, Inc. · 19950
  1. Root · 3 generations backK941234 Non-Mydriatic Retinal CameraCanon USA, Inc. · 1995
  2. 2 generations backK980246 Non-Mydriatic Retinal Camera, Model CR6-45NMCanon USA, Inc. · 1998
  3. 1 generation backK014274 Non-Mydriatic Fundus Camera, Model NM-1000Nidek, Inc. · 2002
  4. This deviceK113451 Non-Mydriatic Auto Fundus Camera Afc-330 with Image Filing Software…Nidek Technologies Srl · 2012
  5. Descendant · cites it directly · 0 of its ownK202097 Fundus CameraHuvitz Co., Ltd. · 2021

Most influential descendants

Showing 1 of 1 · See all
  • K202097Fundus CameraHuvitz Co., Ltd. · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K941234Non-Mydriatic Retinal CameraCanon USA, Inc.1995—
−2K980246Non-Mydriatic Retinal Camera, Model CR6-45NMCanon USA, Inc.1998—
−1K014274Non-Mydriatic Fundus Camera, Model NM-1000Nidek, Inc.2002—
0 (this device)K113451Non-Mydriatic Auto Fundus Camera Afc-330 with Image Filing Software NA Vis-EXNidek Technologies Srl2012—
+1K202097Fundus CameraHuvitz Co., Ltd.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
FIG. 1 — Lineage of K113451, 1995–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HKI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HKI device

  1. #1 K253176 TearViewscore 12.7 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
*K014274Non-Mydriatic Fundus Camera, Model NM-1000
Summary page 2
…TRADITIONAL 510(K) PREMARKET NOTIFICATION Section 05 510(k) Summary (Cont.) PREDICATE DEVICES Nidiek NM-1000 (K014274) Carl Zeiss Meditec FF450 with Fundus Camera and VISLIPAC System (K01 1977) Nidek Technologies Orion (K07023 ) INDICATIONS FOR U…
Nidek, Inc.2002—
*K011877FF450 Plus Fundus Camera and Visupac Digital Image Archiving System, or FF450 Plus…
Our reading of the summary gave K011977, which is not a number in FDA’s lists. Corrected to the number one digit away whose FDA name carries the summary’s words for the device, “Carl Zeiss Meditec FF450 with Fundus Camera and VISUPAC System”.
Carl Zeiss Jena GmbH2001—
Kept off the family treeK070023FDA’s K070023 is Microsurgeon Microwave Tissue Ablation Device, Model Mtad-100
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
Microsurgeon, Inc.2007—

Also cleared against K014274: 4 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K202097Fundus CameraHuvitz Co., Ltd.20210

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 16, 2014Class IIZ-1980-2014Device DesignImage taken by AFC-330 has a white spot which may affect diagnosis or evaluation of image.
Aug 1, 2013Class IIZ-2046-2013Software designUnder certain circumstances, the Left eye image may be saved as the Right eye image.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 3 companies; a new generation was cleared every 3.9 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K202097 (Huvitz Co., Ltd., 2021) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K014274.

How does its category compare?

All HKI devices

HKI Camera, Ophthalmic, Ac-Powered: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · HKI

4in 2025

2007: 22008: 52009: 112010: 62011: 72012: 52013: 72014: 42016: 42017: 22018: 62019: 42020: 52021: 42022: 12023: 52024: 32025: 42026 (partial year): 2
2007high 11 (2009) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HKI

167in 2025

2008: 29 days median2009: 104 days median2010: 140 days median2011: 160 days median2012: 179 days median2013: 141 days median2014: 135 days median2016: 146 days median2018: 71 days median2019: 73 days median2020: 182 days median2021: 155 days median2023: 259 days median2024: 28 days median2025: 167 days median
2007high 259 (2023) · low 28 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HKI

1in 2018

2012: 12016: 12017: 12018: 1
2007high 1 (2012) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HKI

1in 2023

2007: 32008: 12009: 12012: 22013: 222014: 32015: 32016: 662017: 202018: 142020: 52021: 32022: 12023: 1
2007high 66 (2016) · low 0 (2010)2025

MAUDE reports naming product code HKI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—03
200852901
20091110401
2010614000
2011716000
2012517912
20137141022
2014413503
20150—03
20164146166
20172—120
2018671114
201947300
2020518205
2021415503
20221—01
2023525901
202432800
2025416700
20262—0—

What is it indicated for?

The Non-Mydriatic Auto Fundus Camera AFC-330 with Image Filing Software NAVIS-EX is intended to capture, display, store and manipulate images of the retina and the anterior segment of the eye, to aid in diagnosing or monitoring diseases of the eye that may be observed and photographed.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113451: Non-Mydriatic Auto Fundus Camera Afc-330 with Image Filing Software NA Vis-EX.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113451

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026