510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Power-Load Cot Fastener System

K113598510(k) TraditionalCleared Dec 20, 201115-day reviewClass II21 CFR 880.6910Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Stryker Medical · FPO Stretcher, Wheeled

Cleared on Dec 20, 2011 as substantially equivalent to K072598 (International Retail Services Group, LLC, 2008) after a 15-day review. Its family tree reaches back 3 generations to 1995. It has been recalled 3 times.

Decided
Dec 20, 2011
Review
15 days
Class
Class II
Product code
FPO
Recalls
3
Predicates
1

Its longest line of descent runs back 3 generations to K942948 (1995). No later device cites it as a predicate yet.

  1. Root · 3 generations backK942948 Stryker Powered StretcherStryker Corp. · 1995
  2. 2 generations backK022309 Stryker Powered Wheeled StretcherStryker Corp. · 2002 · Recalled
  3. 1 generation backK072598 Irsg I-Mover for BedsInternational Retail Services Group, LLC · 2008
  4. This deviceK113598 Power-Load Cot Fastener SystemStryker Medical · 2011
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K942948Stryker Powered StretcherStryker Corp.1995—
−2K022309Stryker Powered Wheeled StretcherStryker Corp.20026
−1K072598Irsg I-Mover for BedsInternational Retail Services Group, LLC2008—
0 (this device)K113598Power-Load Cot Fastener SystemStryker Medical2011—
FIG. 1 — Lineage of K113598, 1995–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FPO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072598Irsg I-Mover for Beds
Summary page 1
…General Hospital Classification: Class 11 Predicate Device: IRSG I-Mover for Beds, International Retail Services Group LLC (K072598) Device Description The Power-LOAD system secures an ambulance cot within an emergency transport vehicle…
International Retail Services Group, LLC2008—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 20, 2018Class IIZ-1350-2019Under Investigation by firmThe power load floor plates which hold the ambulance cot anchors in place during patient transport, may be broken.
Mar 12, 2018Class IIZ-1389-2018Nonconforming Material/ComponentThe power load floor plates which hold the ambulance cot anchors in place during patient transport, may be broken.
Oct 11, 2012Class IIZ-0566-2013Component design/selectionModel 6390 Power-LOAD systems built between September 12, 2012 and September 27, 2012 may have been built with a load switch slide that does not meet specifications causing the lifting ability of the arms to be negatively impacted.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 5.4 years on median.
  • 1 of the 3 ancestors has a recall of any class.

How does its category compare?

All FPO devices

FPO Stretcher, Wheeled: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · FPO

1in 2011

2008: 12011: 1
2007high 1 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPO

Too few FPO 510(k)s in any year since 2007 to give a median.

Recalls per year · FPO

1in 2025

2007: 62008: 32009: 122010: 12011: 62012: 12014: 12016: 12022: 22025: 12026 (partial year): 2
2007high 12 (2009) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FPO

897in 2025

2007: 1172008: 3852009: 2,3372010: 4,0492011: 5,8912012: 6,3342013: 5,7432014: 1,6332015: 3812016: 2052017: 2302018: 5732019: 6632020: 6942021: 8742022: 6942023: 1,0282024: 9562025: 897
2007high 6,334 (2012) · low 117 (2007)2025

MAUDE reports naming product code FPO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—6117
20081—3385
20090—122,337
20100—14,049
20111—65,891
20120—16,334
20130—05,743
20140—11,633
20150—0381
20160—1205
20170—0230
20180—0573
20190—0663
20200—0694
20210—0874
20220—2694
20230—01,028
20240—0956
20250—1897
20260—2—

What is it indicated for?

The Power-LOADTm cot fastening system (model 6390) is intended to assist wit loading and unloading of a compatible wheeled stretcher (ambulance cot) to and from a transport vehicle and to secure the ambulance cot during transport. The device has a maximum safe working load of 870 Ibs, which includes the weight of the ambulance cot, patient, and equipment attached to the cot (i.e. oxygen bottles, monitors, and/or pumps). The intended users of the device will be trained professionals including: emergency medical service and medical care center personnel, as well as medical first responders, service technicians and installers. The expected service life of the product is 7 years.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113598: Power-Load Cot Fastener System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K113598

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026