510(k) · Cleared
ClearedCheckmate Metatarso-Phalangeal (MTP) Arthrodesis System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Arthrosurface, Inc. · HRS Plate, Fixation, Bone
Cleared on Feb 6, 2012 as substantially equivalent to K101240 (DePuy Orthopaedics, Inc., 2010) and 2 other predicates after a 47-day review. Its family tree reaches back 10 generations to 1996. 1 later device descends from it.
- Decided
- Feb 6, 2012
- Review
- 47 days
- Class
- Class II
- Product code
- HRS
- Recalls
- None
- Predicates
- 3
Its longest line of descent runs back 10 generations to K962616 (1996). 87 ancestors in all, the oldest K792022 (1979). 1 later device descends from it, 1 citing it directly.
- Root · 10 generations backK962616 Synthes Dorsal Distal Radius Plate (DDRP) SystemSynthes (Usa) · 1996
- 6 generations between
- 3 generations backK071563 Peri-Loc Periarticular Locked Plating SystemSmith & Nephew, Inc. · 2007
- 2 generations backK090675 Smith & Nephew, Inc. VLP Foot Plating, Screw System and AccessoriesSmith & Nephew, Inc. · 2009
- 1 generation backK100776 Synthes 2.4 MM/2.7 MM Variable Angle LCP Forefoot/Midfoot SystemSynthes (Usa) · 2010
- This deviceK113762 Checkmate Metatarso-Phalangeal (MTP) Arthrodesis SystemArthrosurface, Inc. · 2012
- Descendant · cites it directly · 0 of its ownK162391 AlignMATE™ Lapidus Arthrodesis SystemArthrosurface, Inc. · 2017
Show all 10 generations
- Root · 10 generations backK962616 Synthes Dorsal Distal Radius Plate (DDRP) SystemSynthes (Usa) · 1996
- 9 generations backK982732 Synthes (Usa) Distal Radius Plate System (DRPS)Synthes · 1998
- 8 generations backK023007 Distal Radius Fracture Repair SystemHand Innovations, Inc. · 2002
- 7 generations backK030198 Hand Innovations, Inc. Distal Radius Fracture Repair SystemHand Innovations, Inc. · 2003
- 6 generations backK032705 Distal Radius Fracture Repair SystemHand Innovations, Inc. · 2003
- 5 generations backK050932 Distal Volar Radius Anatomical Plate SystemHand Innovations, Inc. · 2005 · Recalled
- 4 generations backK051735 Peri-Loc Locking Bone Plates and Locking Bone Screws for the Upper…Smith & Nephew, Inc. · 2005
- 3 generations backK071563 Peri-Loc Periarticular Locked Plating SystemSmith & Nephew, Inc. · 2007
- 2 generations backK090675 Smith & Nephew, Inc. VLP Foot Plating, Screw System and AccessoriesSmith & Nephew, Inc. · 2009
- 1 generation backK100776 Synthes 2.4 MM/2.7 MM Variable Angle LCP Forefoot/Midfoot SystemSynthes (Usa) · 2010
- This deviceK113762 Checkmate Metatarso-Phalangeal (MTP) Arthrodesis SystemArthrosurface, Inc. · 2012
Most influential descendants
Showing 1 of 1 · See all- K162391AlignMATE™ Lapidus Arthrodesis SystemArthrosurface, Inc. · 2017 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −10 | K962616 | Synthes Dorsal Distal Radius Plate (DDRP) System | Synthes (Usa) | 1996 | — |
| −9 | K982732 | Synthes (Usa) Distal Radius Plate System (DRPS) | Synthes | 1998 | — |
| −8 | K023007 | Distal Radius Fracture Repair System | Hand Innovations, Inc. | 2002 | — |
| −7 | K030198 | Hand Innovations, Inc. Distal Radius Fracture Repair System | Hand Innovations, Inc. | 2003 | — |
| −6 | K032705 | Distal Radius Fracture Repair System | Hand Innovations, Inc. | 2003 | — |
| −5 | K050932 | Distal Volar Radius Anatomical Plate System | Hand Innovations, Inc. | 2005 | 1 |
| −4 | K051735 | Peri-Loc Locking Bone Plates and Locking Bone Screws for the Upper Extremity | Smith & Nephew, Inc. | 2005 | — |
| −3 | K071563 | Peri-Loc Periarticular Locked Plating System | Smith & Nephew, Inc. | 2007 | — |
| −2 | K090675 | Smith & Nephew, Inc. VLP Foot Plating, Screw System and Accessories | Smith & Nephew, Inc. | 2009 | — |
| −1 | K100776 | Synthes 2.4 MM/2.7 MM Variable Angle LCP Forefoot/Midfoot System | Synthes (Usa) | 2010 | — |
| 0 (this device) | K113762 | Checkmate Metatarso-Phalangeal (MTP) Arthrodesis System | Arthrosurface, Inc. | 2012 | — |
| +1 | K162391 | AlignMATE™ Lapidus Arthrodesis System | Arthrosurface, Inc. | 2017 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1979 | 1 | 0 |
| 1989 | 1 | 0 |
| 1990 | 1 | 0 |
| 1991 | 1 | 0 |
| 1992 | 1 | 0 |
| 1993 | 2 | 0 |
| 1994 | 5 | 0 |
| 1995 | 2 | 0 |
| 1996 | 6 | 0 |
| 1997 | 1 | 0 |
| 1998 | 3 | 0 |
| 1999 | 3 | 0 |
| 2000 | 8 | 0 |
| 2001 | 8 | 0 |
| 2002 | 4 | 0 |
| 2003 | 3 | 0 |
| 2004 | 6 | 0 |
| 2005 | 5 | 0 |
| 2006 | 7 | 0 |
| 2007 | 2 | 0 |
| 2008 | 2 | 0 |
| 2009 | 3 | 0 |
| 2010 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2017 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HRS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HRS device
- #1 K222465 Tyber Medical Anatomical Plating Systemscore 27.5 · 16 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K101240 | Closed Fusion Plate, MTP Plate, Talus Plate, Model 824077021, 22; 824677021, 22…Summary page 2…d Device are substantially equivalent to the following previously cleared and commercially marketed devices: " DePuy ALPS Small Bone Locked Plating System K101240 * Arthrex Low Profile Plate and Screw System K052614 * Synthes 2.4 mmI2.7 mm Variable Angle (VA)-LCP… | 2010 |
| * | K052614 | Low Profile Plate and Screw SystemSummary page 2…ared and commercially marketed devices: " DePuy ALPS Small Bone Locked Plating System K101240 * Arthrex Low Profile Plate and Screw System K052614 * Synthes 2.4 mmI2.7 mm Variable Angle (VA)-LCP K100776 ForefootM idfoot System Comparative static and dynamic… | 2005 |
| * | K100776 | Synthes 2.4 MM/2.7 MM Variable Angle LCP Forefoot/Midfoot SystemSummary page 2…cked Plating System K101240 * Arthrex Low Profile Plate and Screw System K052614 * Synthes 2.4 mmI2.7 mm Variable Angle (VA)-LCP K100776 ForefootM idfoot System Comparative static and dynamic four point bending test results, along with comparative dimensional a… | 2010 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K162391 | AlignMATE™ Lapidus Arthrodesis System | 2017 | 0 |
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 16 years and passes through 5 companies; a new generation was cleared every 1.6 years on median.
- 19 of the 87 ancestors have a recall of any class.
- Of its descendants, K162391 (Arthrosurface, Inc., 2017) has gone furthest, with 0 descendants of its own.
- 8 other devices also cite its predicate K101240.
How does its category compare?
All HRS devicesHRS Plate, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.
FDA decisions per year · HRS
54in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HRS
90in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HRS
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HRS
2,306in 2025
MAUDE reports naming product code HRS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 35 | 63 | 2 | 339 |
| 2008 | 35 | 87 | 11 | 311 |
| 2009 | 47 | 85 | 3 | 435 |
| 2010 | 49 | 100 | 16 | 491 |
| 2011 | 43 | 88 | 5 | 773 |
| 2012 | 59 | 106 | 4 | 1,663 |
| 2013 | 56 | 89 | 11 | 2,643 |
| 2014 | 76 | 57 | 0 | 2,413 |
| 2015 | 56 | 97 | 7 | 3,095 |
| 2016 | 44 | 109 | 5 | 2,276 |
| 2017 | 57 | 122 | 3 | 2,220 |
| 2018 | 40 | 116 | 2 | 2,663 |
| 2019 | 46 | 89 | 25 | 3,359 |
| 2020 | 48 | 88 | 1 | 2,790 |
| 2021 | 40 | 87 | 9 | 2,442 |
| 2022 | 47 | 144 | 3 | 1,700 |
| 2023 | 40 | 120 | 1 | 2,049 |
| 2024 | 54 | 96 | 0 | 1,849 |
| 2025 | 54 | 90 | 1 | 2,306 |
| 2026 | 51 | 82 | 3 | — |
What is it indicated for?
The CheckMate TM Metatarso-Phalangeal (MTP) Arthrodesis System is intended for use in stabilization and fixation of the 1st MTP joint in the foot for fusion, osteotomy, nonunion, malunion or revision surgery.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K113762 · printed Nov 6, 2026
Cite this page
510(k) Report. “K113762: Checkmate Metatarso-Phalangeal (MTP) Arthrodesis System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113762
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026