# K113776: Flashpak

- Pathway: 510(k) (Traditional)
- Applicant: Symmetry Medical
- Decided: Aug 8, 2012 after 230 days
- Product code: FRG (Wrap, Sterilization)
- Device class: 2
- Regulation: 21 CFR 880.6850
- Recalls of this device: 3 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K871202 (1987). 1 ancestor in all.
1 later device descends from it; 1 cites it directly.
0 of 3 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K871202: Flashpak Container System, Performance Instruments Corp. (1987) Summary page 2.

## Most influential descendants
- K141826: Flashpak Sterilization Container System (2015), 0 descendants of its own

## Recalls of this device
- Nov 29, 2016 · Class II · Z-1011-2017: Sealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.
- Nov 29, 2016 · Class II · Z-1012-2017: Sealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.
- Nov 29, 2016 · Class II · Z-1010-2017: Sealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.

## Indications for use
FlashPak® is a reusable rigid container system to be used during flash sterilization by hospitals and healthcare facilities. It is intended to enable flash sterilization of the enclosed devices and prevent recontamination during immediate transport to the point of use. The container is compatible with gravity-displacement flash sterilization using a 10 minute cycle at 132° C and with pre-vacuum flash sterilization using a 4 minute cycle at 132° C. FlashPak is recommended for surface sterilization of stainless steel instruments and for lumens with the following limits: gravity-displacement (5.5mm inner diameter or larger and up to 184mm in length), pre-vacuum (1mm inner diameter or larger and up to 203mm in length).

Source page: https://510kreport.com/device/K113776