510(k) Report

510(k) · Cleared

Cleared

Gyrus Amci PK Button Electrode

K120567510(k) TraditionalCleared Jun 20, 2012114-day reviewClass II21 CFR 876.4300

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Gyrus Acmi, Inc. · FAS Electrode, Electrosurgical, Active, Urological

Cleared on Jun 20, 2012 as substantially equivalent to K093181 (Gyrus Acmi, Inc., 2010) and 2 other predicates after a 114-day review. Its family tree reaches back 6 generations to 1991. 3 later devices descend from it.

Decided
Jun 20, 2012
Review
114 days
Class
Class II
Product code
FAS
Recalls
None
Predicates
3

Its longest line of descent runs back 6 generations to K911904 (1991). 22 ancestors in all, the oldest K903323 (1990). 3 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2010: 2 other ancestors2008: 1 other ancestor2003: 1 other ancestor1999: 4 other ancestors1998: 1 other ancestor1990: 1 other ancestor2000: 3 other ancestors1997: 1 other ancestor1996: 1 other ancestor1994: 1 other ancestor2017: 1 descendant cleared2022: 1 descendant cleared2024: 1 descendant clearedK911904 · Models Ues-10 & PSD-10 · Olympus Corp. · 1991K970184 · Olympus Ues-20 Electrosurgical System and Its Associated Accessories · Olympus America, Inc. · 1997×2: 2 generations cleared close together, K023767 (2002), K030194 (2003) K100275 · Working Element, Resection Button Electrode for Plasma Vaporizaton, HF-Resection Electrode Loops and Bands · Olympus Winter & Ibe GmbH · 2010 · recalledK102781 · Resection Button Electrode for Plasma Vaporization, HF-Resection Electrode Rollers and Needles, Loops and Bands · Olympus Winter & Ibe GmbH · 2011 · recalled1990200020102020
  1. Root · 6 generations backK911904 Models Ues-10 & PSD-10Olympus Corp. · 1991
  2. 5 generations backK970184 Olympus Ues-20 Electrosurgical System and Its Associated AccessoriesOlympus America, Inc. · 1997
  3. 4 generations backK023767 Ues-30 Electrosurgical Unit and It's Associated AccessoriesThe Olympus Optical Co. · 2002
  4. 3 generations backK030194 Xues-41 Endoscopic Electrosurgical UnitThe Olympus Optical Co. · 2003 · Recalled
  5. 2 generations backK100275 Working Element, Resection Button Electrode for Plasma Vaporizaton…Olympus Winter & Ibe GmbH · 2010 · Recalled
  6. 1 generation backK102781 Resection Button Electrode for Plasma Vaporization, HF-Resection…Olympus Winter & Ibe GmbH · 2011 · Recalled
  7. This deviceK120567 Gyrus Amci PK Button ElectrodeGyrus Acmi, Inc. · 2012
  8. Descendant · cites it directly · 2 of its ownK163090 Plasma Edge SystemChirurgie Innovation · 2017
  9. Descendant · 2 generations on · 1 of its ownK213135 Plasma Edge resection and vaporization electrodes, Plasma Edge…Chirurgie Innovation · 2022
  10. Descendant · 3 generations on · 0 of its ownK240504 PLASMA EDGE SystemLamidey Noury Medical · 2024

Most influential descendants

Showing 3 of 3 · See all
  • K163090Plasma Edge SystemChirurgie Innovation · 2017 · cites it directly2descendants
  • K213135Plasma Edge resection and vaporization electrodes, Plasma Edge working elements, Adaptor…Chirurgie Innovation · 2022 · 2 generations on1descendant
  • K240504PLASMA EDGE SystemLamidey Noury Medical · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K911904Models Ues-10 & PSD-10Olympus Corp.1991—
−5K970184Olympus Ues-20 Electrosurgical System and Its Associated AccessoriesOlympus America, Inc.1997—
−4K023767Ues-30 Electrosurgical Unit and It's Associated AccessoriesThe Olympus Optical Co.2002—
−3K030194Xues-41 Endoscopic Electrosurgical UnitThe Olympus Optical Co.20033
−2K100275Working Element, Resection Button Electrode for Plasma Vaporizaton…Olympus Winter & Ibe GmbH20101
−1K102781Resection Button Electrode for Plasma Vaporization, HF-Resection Electrode…Olympus Winter & Ibe GmbH20111
0 (this device)K120567Gyrus Amci PK Button ElectrodeGyrus Acmi, Inc.2012—
+1K163090Plasma Edge SystemChirurgie Innovation2017—
+2K213135Plasma Edge resection and vaporization electrodes, Plasma Edge working…Chirurgie Innovation2022—
+3K240504PLASMA EDGE SystemLamidey Noury Medical2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199410
199610
199710
199810
199940
200030
200310
200810
201020

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202210
202410
FIG. 1 — Lineage of K120567, 1991–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FAS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FAS device

  1. #1 K250522 Multi4 Systemscore 11.4 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K093181Gyrus Acmi PK Button Electrode
Summary page 2
…Cyrus ACMI, Inc. June 4, 2012 136 Turnpike Road Southborough, MA 01772 Predicate Devices Gyrus ACMI PK Button Electrode (K09318 1) - Existing design Olympus HF-Resection Electrodes (1(102781) - Existing design /desired indications Gyrus ACMI SuperPulse System…
Gyrus Acmi, Inc.2010—
*K102781Resection Button Electrode for Plasma Vaporization, HF-Resection Electrode Rollers and…
Summary page 2
…he system for optimal safe performance with the Gymus AGMI PlC® Button Electrode. This is the identical technology used in the predicate devices, including (K(093181 and K10278 1). Like the predicate K093181, and K(102781 the distal tip of the electrode is a Stainless Steel 304 button measuring 3.16mm in d…
Olympus Winter & Ibe GmbH20111
*K100816PK Superpulse Sysem Generator Model 744000
Summary page 2
…s ACMI PK Button Electrode (K09318 1) - Existing design Olympus HF-Resection Electrodes (1(102781) - Existing design /desired indications Gyrus ACMI SuperPulse System (1(100816) - Common generator Intended Use The Gyrus ACMI PlC® Button Electrode is a bipolar instrument intended for use in urological ele…
Gyrus Acmi, Inc.2010—

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K163090Plasma Edge SystemChirurgie Innovation20172

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 5 companies; a new generation was cleared every 5.6 years on median.
  • 4 of the 22 ancestors have a recall of any class.
  • Of its descendants, K163090 (Chirurgie Innovation, 2017) has gone furthest, with 2 descendants of its own.

How does its category compare?

All FAS devices

FAS Electrode, Electrosurgical, Active, Urological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · FAS

2in 2025

2010: 32011: 12012: 22015: 12016: 12017: 22021: 22022: 12024: 12025: 2
2007high 3 (2010) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FAS

150in 2010

2010: 150 days median
2007high 150 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FAS

1in 2023

2016: 22021: 42023: 1
2007high 4 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FAS

395in 2025

2007: 62008: 132009: 92010: 252011: 332012: 302013: 482014: 522015: 822016: 692017: 1212018: 702019: 612020: 1002021: 1022022: 1472023: 3292024: 3102025: 395
2007high 395 (2025) · low 6 (2007)2025

MAUDE reports naming product code FAS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—06
20080—013
20090—09
20103150025
20111—033
20122—030
20130—048
20140—052
20151—082
20161—269
20172—0121
20180—070
20190—061
20200—0100
20212—4102
20221—0147
20230—1329
20241—0310
20252—0395
20260—0—

What is it indicated for?

The Gyrus AGMI PK® Button Electrode is a bipolar instrument intended for use in urological electrosurgical procedures involving the ablation or removal of soft tissue and where associated hemnostasis is required. The specific urological indication fdr the PK@ Button device is Transurethral Vaporization of the prostate (TIJVP/TVP) fdr benign prostatic hypertrophy, and for Transurethral Vaporization of bladder tumors. The PK&~ Button is not to be used to resect tissue.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K120567: Gyrus Amci PK Button Electrode.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K120567

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026