## Lineage
Longest line of descent: 3 generations back to K842616 (1984). 4 ancestors in all.
No later device cites it as a predicate yet.
0 of 5 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K102422: Spacelabs Medical Patient Monitors, Spacelabs Medical, Inc. (2010) Summary page 1.

## Recalls of this device
- Sep 5, 2025 · Class II · Z-0169-2026: Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with t
- Sep 5, 2025 · Class II · Z-0167-2026: Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with t
- Sep 5, 2025 · Class II · Z-0168-2026: Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with t
- May 8, 2015 · Class II · Z-1608-2015: The input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective
- Jan 28, 2015 · Class II · Z-1145-2015: Spacelabs Healthcare qube Bedside Monitors, Model 91390, are recalled because the firm has received multiple reports of qube monitors failing to boot up or returning to factory default configuration s
- Jun 6, 2014 · Class II · Z-1852-2014: Spacelabs Healthcare is recalling Qube Compact Monitor, Model 91390, due to a potential chance of losing network connection after switching from hardwired Ethernet connection to using the Qube Docking
- Oct 17, 2013 · Class II · Z-0188-2014: Spacelabs Healthcare elance Vital Signs Monitor with Option S, model 93330, is recalled because the monitor will not permit connection of any of the Spacelabs TruLink line of SpO2 sensors.
- Oct 17, 2013 · Class II · Z-0351-2014: Spacelabs Healthcare qube Compact Patient Monitor, Model 91390, may fail to power ON due to incorrect programmed microcontroller component.
- Oct 19, 2012 · Class II · Z-0403-2013: Spacelabs Healthcare has learned through several reports that the input circuit may fail and the monitor will go on battery mode. The monitor will function normally until the batteries become depleted

## Indications for use
The Spacelabs Healthcare Qube Compact Monitor (91390), functioning as either bedside or central monitors; passively displays data generated by Spacelabs Heatthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the Model 91390 when employing the Spacelabs Command Module, consist of ECG, respiration, invasive and noninvasive blood pressure, SpO 2 , temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed. The Qube is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other physiologic monitors via Flexport interfaces. These devices determine a) when an alarm condition is violated; b) the alarm priority (i.e. high, medium or low); c) alarm limits; and d) when to initiate and terminate alarm notifications. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Qube may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitors allow networkbased applications to open windows and display information on other networked monitors. The Qube is also designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Qube is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.