# K120771: Unicel DXH 800 Coulter Cellular Analysis System

- Pathway: 510(k) (Abbreviated)
- Applicant: Beckman Coulter, Inc.
- Decided: Mar 22, 2013 after 373 days
- Product code: GKZ (Counter, Differential Cell)
- Device class: 2
- Regulation: 21 CFR 864.5220
- Recalls of this device: 11 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
7 later devices descend from it; 1 cites it directly.
1 of 8 devices in its family (counting itself) has a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. The summary cites one number, which could not be confirmed as its predicate (listed below).

## Cited in the summary but kept off the family tree
- K081645: Agilis NXT Steerable Introducer, Model G408324, St Jude Medical (2008). Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree

## Most influential descendants
- K140911: Unicel DXH 800 Coulter Cellular Analysis System/ Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System (2014), 6 descendants of its own
- K193124: Unicel DxH 800 Coulter Cellular Analysis System (2020), 1 descendant of its own
- K162414: UniCel DxH SlideMaker Stainer Coulter Cellular Analysis System (2016), 1 descendant of its own
- K240252: UniCel DxH 900 Coulter Cellular Analysis System; UniCel DxH Slidemaker Stainer II Coulter Cellular Analysis System; UniCel DxH 690T Coulter Cellular Analysis System (2024), 0 descendants of its own
- K230878: QScout Lab; QScout RLD (2023), 0 descendants of its own

## Recalls of this device
- Aug 13, 2025 · Class II · Z-2575-2025: Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
- Apr 24, 2017 · Class II · Z-2158-2017: There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The 
- Apr 24, 2017 · Class II · Z-2157-2017: There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The 
- Dec 19, 2016 · Class II · Z-1013-2017: Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not fl
- Dec 19, 2016 · Class II · Z-1014-2017: Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not fl
- Dec 1, 2015 · Class II · Z-0433-2016: Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple order
- Dec 1, 2015 · Class II · Z-0434-2016: Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple order
- Dec 1, 2015 · Class II · Z-0432-2016: Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple order
- Apr 13, 2015 · Class II · Z-1616-2015: Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in th
- Nov 8, 2013 · Class II · Z-1828-2014: Beckman Coulter is recalling the UniCel DxH800 and UniCel DxH600 Coulter Analysis Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cel
- Sep 24, 2013 · Class II · Z-0147-2014: Beckman Coulter is initiating a recall for UniCel DxH 800 Coulter Cellular Analysis System (PN 629029) and UniCel DxH 600 Coulter Cellular Analysis System (PN B23858) because the company's internal in

## Indications for use
The UniCel® DxH 800 Analyzer is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel® DxH 800 Analyzer identifies and enumerates the parameters indicated below on the following sample types: * Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF * Pre-Diluted Whole Blood (Venous and Capillary) o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV * Body Fluids (cerebrospinal, serous and synovial) o TNC and RBC