510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Fortex Pedicle Screw System

K120832510(k) SpecialCleared Oct 23, 2012218-day reviewClass II21 CFR 888.3070Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

X-Spine Systems, Inc. · MNH Orthosis, Spondylolisthesis Spinal Fixation

Cleared on Oct 23, 2012 after a 218-day review. The FDA summary names its predicate without a number. 3 later devices descend from it. It has been recalled 3 times.

Decided
Oct 23, 2012
Review
218 days
Class
Class II
Product code
MNH
Recalls
3
Predicates
0

3 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2018: 2 descendants cleared2015: 1 descendant clearedReference device: K080792 · K2M Cocr Rod · K2m, Inc. · 20082010201520202025
  1. This deviceK120832 Fortex Pedicle Screw SystemX-Spine Systems, Inc. · 2012
  2. Descendant · cites it directly · 1 of its ownK152132 X-spine Pedicle Screw SystemX-Spine Systems, Inc. · 2015 · Recalled
  3. Descendant · cites it directly · 0 of its ownK180655 Viking Lumbar Polyaxial Screw SystemPresidio Surgical · 2018
  4. Descendant · 2 generations on · 0 of its ownK180153 X-spine Cortical Bone Screw SystemX-Spine Systems, Inc. · 2018

Most influential descendants

Showing 3 of 3 · See all
  • K152132X-spine Pedicle Screw SystemX-Spine Systems, Inc. · 2015 · cites it directly1descendant
  • K180655Viking Lumbar Polyaxial Screw SystemPresidio Surgical · 2018 · cites it directly0descendants
  • K180153X-spine Cortical Bone Screw SystemX-Spine Systems, Inc. · 2018 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K120832Fortex Pedicle Screw SystemX-Spine Systems, Inc.2012—
+1K152132X-spine Pedicle Screw SystemX-Spine Systems, Inc.20151
+1K180655Viking Lumbar Polyaxial Screw SystemPresidio Surgical2018—
+2K180153X-spine Cortical Bone Screw SystemX-Spine Systems, Inc.2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201820
FIG. 1 — Lineage of K120832, 2012–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Fortex Pedicle Screw System
Named without a submission number
—
†K080792K2M Cocr RodK2m, Inc.2008—

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K152132X-spine Pedicle Screw SystemX-Spine Systems, Inc.20151
K180655Viking Lumbar Polyaxial Screw SystemPresidio Surgical20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 12, 2017Class IIZ-1884-2017Labeling mix-upsThe material type on the label may incorrectly state Ti6Al-4V ELI. The rods are composed of Cobalt Chromium.
Apr 7, 2015Class IIZ-1823-2015OtherThe Continuous Radius Rod should measure 50mm in length but actually measures 45mm in length.
Mar 11, 2015Class IIZ-1512-2015Nonconforming Material/ComponentOne lot of instruments have etched on the Yoke of the Pedicle Screw a laser marking size of 7.5 x 55mm instead of 5.5 x 55mm as described on the product label.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K152132 (X-Spine Systems, Inc., 2015) has gone furthest, with 1 descendant of its own.

How does its category compare?

All MNH devices

MNH Orthosis, Spondylolisthesis Spinal Fixation: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · MNH

8in 2016

2007: 52008: 132009: 132010: 102011: 172012: 72013: 102014: 122015: 102016: 8
2007high 17 (2011) · low 0 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MNH

93in 2016

2007: 123 days median2008: 65 days median2009: 151 days median2010: 106 days median2011: 171 days median2012: 93 days median2013: 124 days median2014: 140 days median2015: 99 days median2016: 93 days median
2007high 171 (2011) · low 65 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MNH

1in 2016

2009: 72012: 12013: 12016: 1
2007high 7 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MNH

89in 2025

2007: 182008: 302009: 452010: 592011: 1072012: 4332013: 4642014: 2862015: 1522016: 2162017: 1742018: 1292019: 4032020: 4322021: 5312022: 2652023: 1212024: 3022025: 89
2007high 531 (2021) · low 18 (2007)2025

MAUDE reports naming product code MNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075123018
20081365030
200913151745
201010106059
2011171710107
20127931433
2013101241464
2014121400286
201510990152
20168931216
20170—0174
20180—0129
20190—0403
20200—0432
20210—0531
20220—0265
20230—0121
20240—0302
20250—089
20260—0—

What is it indicated for?

The Fortex Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L-5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L-3 to sacrum) with removal of the implants after the attainment of a solid fusion. In addition, the Fortex Pedicle Screw System is intended to provide immobilization and stabilization of spinal…

Read the full statement (119 words)

The Fortex Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L-5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L-3 to sacrum) with removal of the implants after the attainment of a solid fusion.

In addition, the Fortex Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K120832: Fortex Pedicle Screw System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K120832

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026