510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Mega (TM) 7.5FR. 30CC Intra-Aortic Balloon Catheter and Accessories, Mega (TM) 7.5FR. 40CC Intra-Aortic Balloon Catheter

K120868510(k) SpecialCleared Apr 20, 201229-day reviewClass II21 CFR 870.3535Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datascope Corp., Cardiac Assist Division · DSP System, Balloon, Intra-Aortic And Control

Cleared on Apr 20, 2012 after a 29-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 1 later device descends from it. It has been recalled 7 times.

Decided
Apr 20, 2012
Review
29 days
Class
Class II
Product code
DSP
Recalls
7
Predicates
0

1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2012: 1 descendant cleared2010201520202025
  1. This deviceK120868 Mega (TM) 7.5FR. 30CC Intra-Aortic Balloon Catheter and Accessories…Datascope Corp., Cardiac Assist Division · 2012
  2. Descendant · cites it directly · 0 of its ownK122628 Sensation Plus 7.5FR. 40CC Intra-Aortic Balloon (Iab) Catheter and…Datascope Corp., Cardiac Assist Division · 2012 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K122628Sensation Plus 7.5FR. 40CC Intra-Aortic Balloon (Iab) Catheter and AccessoriesDatascope Corp., Cardiac Assist Division · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K120868Mega (TM) 7.5FR. 30CC Intra-Aortic Balloon Catheter and Accessories, Mega (TM)…Datascope Corp., Cardiac Assist Division2012—
+1K122628Sensation Plus 7.5FR. 40CC Intra-Aortic Balloon (Iab) Catheter and AccessoriesDatascope Corp., Cardiac Assist Division20123

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
FIG. 1 — Lineage of K120868, 2012–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DSP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSP device

  1. #1 K250542 AC3™ Range™ Intra-Aortic Balloon Pumpscore 11.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K122628Sensation Plus 7.5FR. 40CC Intra-Aortic Balloon (Iab) Catheter and AccessoriesDatascope Corp., Cardiac Assist Division20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 23, 2023Class IIZ-2219-2023Under Investigation by firmDuring IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization…
Jul 27, 2020Class IIZ-2832-2020Process change controlPotential Endotoxin Contamination
Jul 27, 2020Class IIZ-2831-2020Process change controlPotential Endotoxin Contamination
Jul 27, 2020Class IIZ-2833-2020Process change controlPotential Endotoxin Contamination
Feb 21, 2018Class IIZ-1384-2018Error in labelingThe U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
Feb 21, 2018Class IIZ-1383-2018Error in labelingThe U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
May 13, 2016Class IIZ-1935-2016Unknown/Undetermined by firmIt was discovered during post-quarantine packaging on 2/16/2016 that the IFU for the STATLOCK Sheath Stabilization device for Percutaneous Sheath Introducer (Vendor PN 2403097-1405R) was erroneously packaged with the Mega and Sensation…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K122628 (Datascope Corp., Cardiac Assist Division, 2012) has gone furthest, with 0 descendants of its own.

How does its category compare?

All DSP devices

DSP System, Balloon, Intra-Aortic And Control: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · DSP

1in 2025

2007: 22009: 22011: 22012: 22013: 12015: 12017: 32018: 12019: 32020: 22023: 12025: 1
2007high 3 (2017) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSP

140in 2019

2017: 73 days median2019: 140 days median
2007high 140 (2019) · low 73 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSP

4in 2025

2007: 32009: 142010: 82012: 32014: 12016: 9, 9 Class I2018: 2, 2 Class I2019: 32021: 5, 3 Class I2022: 33, 33 Class I2024: 12025: 4
2007high 33 (2022) · low 0 (2008)2026

Class I (47 since 2007) Class II/III not classified

Adverse event reports per year · DSP

7,422in 2025

2007: 4612008: 6172009: 6562010: 1,0192011: 5902012: 4902013: 4952014: 1,7382015: 4042016: 4192017: 2,1642018: 3,3312019: 3,5522020: 3,3932021: 4,3452022: 5,1132023: 6,9802024: 7,3172025: 7,422
2007high 7,422 (2025) · low 404 (2015)2025

MAUDE reports naming product code DSP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—3461
20080—0617
20092—14656
20100—81,019
20112—0590
20122—3490
20131—0495
20140—11,738
20151—0404
20160—9 (9)419
201737302,164
20181—2 (2)3,331
2019314033,552
20202—03,393
20210—5 (3)4,345
20220—33 (33)5,113
20231—06,980
20240—17,317
20251—47,422
20260—0—
Cite this page

510(k) Report. “K120868: Mega (TM) 7.5FR. 30CC Intra-Aortic Balloon Catheter and Accessories, Mega (TM) 7.5FR. 40CC Intra-Aortic Balloon Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K120868

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026