510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Ivue with Normative Database

K121739510(k) TraditionalCleared Jan 18, 2013219-day reviewClass II21 CFR 886.1570Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Optovue, Inc. · HLI Ophthalmoscope, Ac-Powered

Cleared on Jan 18, 2013 as substantially equivalent to K091404 (Optovue, Inc., 2010) and 1 other predicate after a 219-day review. Its family tree reaches back 6 generations to 2002. 14 later devices descend from it. It has been recalled 3 times.

Decided
Jan 18, 2013
Review
219 days
Class
Class II
Product code
HLI
Recalls
3
Predicates
2

Its longest line of descent runs back 6 generations to K012727 (2002). 9 ancestors in all. 14 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2010: 1 other ancestor2006: 1 other ancestor2005: 1 other ancestor2023: 2 descendants cleared2013: 1 descendant cleared2016: 1 descendant cleared2017: 3 descendants cleared2024: 4 descendants cleared2014: 1 descendant cleared2018: 1 descendant cleared2026: 1 descendant clearedK012727 · Humphrey OCT3 · Carl Zeiss Ophthalmic Systems, Inc. · 2002K030433 · Stratusoct with RNFL Normative Database · Carl Zeiss Meditec, Inc. · 2003 · recalledK033123 · Stratusoct with RNFL & Macula Normative Database · Carl Zeiss Meditec, Inc. · 2004 · recalledK063343 · Bioptigen Spectral Domain Ophthalmic Imaging System · Bioptigen, Inc. · 2006K071250 · CA, Model: Cam · Optovue, Inc. · 2007K091404 · Ivue, Model 100 · Optovue, Inc. · 2010 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 1 other ancestor 1 generation forward: 6 descendants2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 2 descendants5 generations forward: 1 descendant K091404 · Ivue, Model 100 · Optovue, Inc. · 2010 · recalledK071250 · CA, Model: Cam · Optovue, Inc. · 2007K063343 · Bioptigen Spectral Domain Ophthalmic Imaging System · Bioptigen, Inc. · 2006K033123 · Stratusoct with RNFL & Macula Normative Database · Carl Zeiss Meditec, Inc. · 2004 · recalledK030433 · Stratusoct with RNFL Normative Database · Carl Zeiss Meditec, Inc. · 2003 · recalledK012727 · Humphrey OCT3 · Carl Zeiss Ophthalmic Systems, Inc. · 2002−50+5
  1. Root · 6 generations backK012727 Humphrey OCT3Carl Zeiss Ophthalmic Systems, Inc. · 2002
  2. 5 generations backK030433 Stratusoct with RNFL Normative DatabaseCarl Zeiss Meditec, Inc. · 2003 · Recalled
  3. 4 generations backK033123 Stratusoct with RNFL & Macula Normative DatabaseCarl Zeiss Meditec, Inc. · 2004 · Recalled
  4. 3 generations backK063343 Bioptigen Spectral Domain Ophthalmic Imaging SystemBioptigen, Inc. · 2006
  5. 2 generations backK071250 CA, Model: CamOptovue, Inc. · 2007
  6. 1 generation backK091404 Ivue, Model 100Optovue, Inc. · 2010 · Recalled
  7. This deviceK121739 Ivue with Normative DatabaseOptovue, Inc. · 2013
  8. Descendant · cites it directly · 8 of its ownK161509 3D OCT-1 MaestroTopcon Corporation · 2016
  9. Descendant · 2 generations on · 5 of its ownK170164 3D OCT-1 MaestroTopcon Corporation · 2017
  10. Descendant · 3 generations on · 3 of its ownK231222 3D OPTICAL COHERENCE TOMOGRAPHY 3D OCT-1(type: Maestro2)Topcon Corporation · 2023
  11. Descendant · 2 generations on · 3 of its ownK171370 IMAGEnet 6 Ophthalmic Data SystemTopcon Corporation · 2017 · Recalled
  12. Descendant · 4 generations on · 2 of its ownK233561 3D Optical Coherence Tomography (3D OCT-1(type: Maestro2))Topcon Corporation · 2024

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2013: 12014: 12016: 12017: 32018: 12023: 22024: 42026: 1
2013high 4 (2024) · low 0 (2015)2026

Most influential descendants

Showing 5 of 14 · See all
  • K1615093D OCT-1 MaestroTopcon Corporation · 2016 · cites it directly8descendants
  • K1701643D OCT-1 MaestroTopcon Corporation · 2017 · 2 generations on5descendants
  • K2312223D OPTICAL COHERENCE TOMOGRAPHY 3D OCT-1(type: Maestro2)Topcon Corporation · 2023 · 3 generations on3descendants
  • K171370IMAGEnet 6 Ophthalmic Data SystemTopcon Corporation · 2017 · 2 generations on3descendants
  • K2335613D Optical Coherence Tomography (3D OCT-1(type: Maestro2))Topcon Corporation · 2024 · 4 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K012727Humphrey OCT3Carl Zeiss Ophthalmic Systems, Inc.2002—
−5K030433Stratusoct with RNFL Normative DatabaseCarl Zeiss Meditec, Inc.20031
−4K033123Stratusoct with RNFL & Macula Normative DatabaseCarl Zeiss Meditec, Inc.20041
−3K063343Bioptigen Spectral Domain Ophthalmic Imaging SystemBioptigen, Inc.2006—
−2K071250CA, Model: CamOptovue, Inc.2007—
−1K091404Ivue, Model 100Optovue, Inc.20102
0 (this device)K121739Ivue with Normative DatabaseOptovue, Inc.2013—
+1K1615093D OCT-1 MaestroTopcon Corporation2016—
+2K1701643D OCT-1 MaestroTopcon Corporation2017—
+3K2312223D OPTICAL COHERENCE TOMOGRAPHY 3D OCT-1(type: Maestro2)Topcon Corporation2023—
+2K171370IMAGEnet 6 Ophthalmic Data SystemTopcon Corporation20171
+4K2335613D Optical Coherence Tomography (3D OCT-1(type: Maestro2))Topcon Corporation2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200510
200610
201010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201410
201610
201730
201810
202320
202440
202610
FIG. 1 — Lineage of K121739, 2002–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HLI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K091404Ivue, Model 100
Summary page 1
…CER§ 886.1570) Classification: Class 11 (ace. 21 CFR 886.1570) Product Code: HLI Predicate Devices. (1) iVue® (K091404) - manufactured by Optovue, Inc. (2) RTVue® with Normative Database (KIO0 1505) - manufactured by Optovue, Inc.…
Optovue, Inc.20102
*K101505Rtvue Model Rtvue 100
Summary page 2
…RTVue with NDB. iVue with NDB, based on the same Optical Coherence Tomography (OCT) technology that is used in the predicate device iVue (K09 1404) and RTVue with NDB (K 101505), is a non- invasive diagnostic device for viewing the ocular tissue structure with micrometer range resolution. Both iVue and RT…
Optovue, Inc.20101

Also cleared against K091404: 1 other device →

What descends from it? (14)

See all 14

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K1615093D OCT-1 MaestroTopcon Corporation20168
K231673P200TE (A10700)Optos Plc.20231
K233602P200TE (A10700)Optos Plc.20240
K163475iVueOptovue, Inc.20170
K133892IVUE 500Optovue, Inc.20140
K130656IfusionOptovue, Inc.20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 27, 2017Class IIZ-0008-2018OtherFDA determined that a 510(k) is needed for the Vault Mapping software (also known as Lens Fitting).
Mar 16, 2016Class IIZ-1498-2016Under Investigation by firmA software anomaly was detected that affects Optovue iVue and iVue 500 with Normative Database that may display images of the same eye rather than right and left if user scrolls on the map.
May 14, 2013Class IIIZ-1441-2013Software designColors from the NDB comparison for Ganglion Cell Complex thickness in the singular OU report in iVue version 3.0 (US) and 3.1(international) are different from those in the individual iWellness ID and OS reports.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 4 companies; a new generation was cleared every 2.6 years on median.
  • 4 of the 9 ancestors have a recall of any class.
  • Of its descendants, K161509 (Topcon Corporation, 2016) has gone furthest, with 8 descendants of its own.
  • 1 other device also cites its predicate K091404.

How does its category compare?

All HLI devices

HLI Ophthalmoscope, Ac-Powered: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · HLI

1in 2016

2007: 12010: 22011: 32012: 22013: 22014: 42015: 22016: 1
2007high 4 (2014) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HLI

112in 2014

2011: 36 days median2014: 112 days median
2007high 112 (2014) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HLI

1in 2021

2008: 12009: 12017: 22021: 1
2007high 2 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HLI

2in 2024

2008: 12009: 12013: 12018: 12020: 12022: 12023: 22024: 2
2007high 2 (2023) · low 0 (2007)2025

MAUDE reports naming product code HLI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20080—11
20090—11
20102—00
201133600
20122—00
20132—01
2014411200
20152—00
20161—00
20170—20
20180—01
20190—00
20200—01
20210—10
20220—01
20230—02
20240—02
20250—00
20260—0—

What is it indicated for?

The iVue is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (0CC), optic disc, cornea, and anterior chamber of the eye. The iVue with Normative Database is a quantitative tool for the comparison of retina, retinal nerve fiber…

Read the full statement (110 words)

The iVue is a non-contact, high resolution tomographic imaging device. It is intended for in vivo imaging, axial cross-sectional, and three-dimensional imaging and measurement of anterior and posterior ocular structures, including retina, retinal nerve fiber layer, ganglion cell complex (0CC), optic disc, cornea, and anterior chamber of the eye. The iVue with Normative Database is a quantitative tool for the comparison of retina, retinal nerve fiber layer, ganglion cell complex, and optic disc measurements to a database of known normal subjects. The iVue with Normative Database is indicated for use as a device to aid in the diagnosis, documentation, and management of ocular health and diseases in the adult population.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K121739: Ivue with Normative Database.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K121739

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026