510(k) · Cleared
ClearedDMRC Nanoflow DMRC Nanocomposite Ebond Sure Etch Gel Sure Etch Liquid
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Danville Materials, Inc. · EBF Material, Tooth Shade, Resin
Cleared on Sep 25, 2012 after an 89-day review. FDA’s file for it is a 510(k) statement, which names no predicate.
- Decided
- Sep 25, 2012
- Review
- 89 days
- Class
- Class II
- Product code
- EBF
- Recalls
- None
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in EBF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new EBF device
- #1 K202846 Tera Harzscore 18.5 · 7 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
How does its category compare?
All EBF devicesEBF Material, Tooth Shade, Resin: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.
FDA decisions per year · EBF
15in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · EBF
43in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · EBF
1in 2020
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · EBF
1in 2025
MAUDE reports naming product code EBF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 19 | 49 | 1 | 6 |
| 2008 | 14 | 79 | 42 | 10 |
| 2009 | 20 | 84 | 1 | 8 |
| 2010 | 19 | 75 | 6 | 31 |
| 2011 | 19 | 81 | 7 | 31 |
| 2012 | 11 | 86 | 0 | 62 |
| 2013 | 13 | 153 | 0 | 246 |
| 2014 | 10 | 112 | 0 | 195 |
| 2015 | 17 | 123 | 0 | 119 |
| 2016 | 20 | 205 | 0 | 143 |
| 2017 | 15 | 108 | 4 | 42 |
| 2018 | 15 | 115 | 0 | 8 |
| 2019 | 17 | 159 | 2 | 11 |
| 2020 | 15 | 190 | 1 | 2 |
| 2021 | 13 | 178 | 0 | 2 |
| 2022 | 17 | 63 | 0 | 2 |
| 2023 | 9 | 87 | 0 | 8 |
| 2024 | 18 | 120 | 0 | 5 |
| 2025 | 15 | 43 | 0 | 1 |
| 2026 | 16 | 59 | 2 | — |
https://510kreport.com/device/K121880 · printed Nov 6, 2026
Cite this page
510(k) Report. “K121880: DMRC Nanoflow DMRC Nanocomposite Ebond Sure Etch Gel Sure Etch Liquid.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K121880
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026