510(k) · Cleared
Recalled · Class II · 2019Smith & Nephew, Inc. Intramedullary Nail System Instrumentation
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Smith & Nephew, Inc. · HSB Rod, Fixation, Intramedullary And Accessories
Cleared on Oct 19, 2012 as substantially equivalent to K811002 (Richard'S Medical Equip., Inc., 1981) and 17 other predicates after an 88-day review. Its family tree reaches back 5 generations to 1998. It has been recalled once.
- Decided
- Oct 19, 2012
- Review
- 88 days
- Class
- Class II
- Product code
- HSB
- Recalls
- 1
- Predicates
- 18
Its longest line of descent runs back 5 generations to K981529 (1998). 35 ancestors in all, the oldest K811002 (1981). No later device cites it as a predicate yet.
- Root · 5 generations backK981529 Titanium Intramedullary Nail, Titanium Locking ScrewSmith & Nephew, Inc. · 1998 · Recalled
- 4 generations backK040212 Trigen IntertanSmith & Nephew, Inc. · 2004 · Recalled
- 3 generations backK040656 Modification to Smith & Nephew Intramedullary Hip ScrewSmith & Nephew, Inc. · 2004
- 2 generations backK050226 Asian Intramedullary Hip Screw NailsSmith & Nephew, Inc. · 2005
- 1 generation backK092748 Sureshot Tan Nails and AccessoriesSmith & Nephew, Inc. · 2010 · Recalled
- This deviceK122170 Smith & Nephew, Inc. Intramedullary Nail System InstrumentationSmith & Nephew, Inc. · 2012
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −5 | K981529 | Titanium Intramedullary Nail, Titanium Locking Screw | Smith & Nephew, Inc. | 1998 | 2 |
| −4 | K040212 | Trigen Intertan | Smith & Nephew, Inc. | 2004 | 6 |
| −3 | K040656 | Modification to Smith & Nephew Intramedullary Hip Screw | Smith & Nephew, Inc. | 2004 | — |
| −2 | K050226 | Asian Intramedullary Hip Screw Nails | Smith & Nephew, Inc. | 2005 | — |
| −1 | K092748 | Sureshot Tan Nails and Accessories | Smith & Nephew, Inc. | 2010 | 1 |
| 0 (this device) | K122170 | Smith & Nephew, Inc. Intramedullary Nail System Instrumentation | Smith & Nephew, Inc. | 2012 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1981 | 1 | 0 |
| 1989 | 1 | 0 |
| 1992 | 1 | 0 |
| 1995 | 5 | 0 |
| 1996 | 1 | 0 |
| 1998 | 3 | 0 |
| 1999 | 2 | 0 |
| 2001 | 3 | 0 |
| 2002 | 1 | 0 |
| 2003 | 4 | 0 |
| 2004 | 3 | 0 |
| 2005 | 2 | 0 |
| 2006 | 2 | 0 |
| 2011 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HSB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HSB device
- #1 K200869 Gamma3 Systemscore 19.8 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (18)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K811002 | Ender PinsSummary page 1…ponent metallic bone fixation appliances and accessories Device Class: Class II Panel Code: Orthopaedics/87 Predicate Devices: Ender Nail (K811002); TriMax Nail System (K964163); Friedi Gliding Nail modified (K974409); Titanium Nail System (K9… | 1981 |
| * | K964163 | Trimax Nail SystemSummary page 1…and accessories Device Class: Class II Panel Code: Orthopaedics/87 Predicate Devices: Ender Nail (K811002); TriMax Nail System (K964163); Friedi Gliding Nail modified (K974409); Titanium Nail System (K981529); Friedi Gliding Nails a… | 1996 |
| * | K974409 | Friedl Gliding Nail SystemSummary page 1…I Panel Code: Orthopaedics/87 Predicate Devices: Ender Nail (K811002); TriMax Nail System (K964163); Friedi Gliding Nail modified (K974409); Titanium Nail System (K981529); Friedi Gliding Nails addition smaller nails (K020240); IP-XS C… | 1998 |
| * | K981529 | Titanium Intramedullary Nail, Titanium Locking ScrewSummary page 1…te Devices: Ender Nail (K811002); TriMax Nail System (K964163); Friedi Gliding Nail modified (K974409); Titanium Nail System (K981529); Friedi Gliding Nails addition smaller nails (K020240); IP-XS Compression Nail System (K032548);… | 1998 |
| * | K020240 | Slim Gliding Nail SystemSummary page 1…l System (K964163); Friedi Gliding Nail modified (K974409); Titanium Nail System (K981529); Friedi Gliding Nails addition smaller nails (K020240); IP-XS Compression Nail System (K032548); TriGen Straight Humeral Nail System (K032722); Friedl… | 2002 |
| * | K032548 | IP-XS Compression Nail SystemSummary page 1…odified (K974409); Titanium Nail System (K981529); Friedi Gliding Nails addition smaller nails (K020240); IP-XS Compression Nail System (K032548); TriGen Straight Humeral Nail System (K032722); Friedl Gliding Nails Gliding nails with chamfered blade (K… | 2003 |
| * | K032722 | Trigen Straight Humeral Nail SystemSummary page 1…9); Friedi Gliding Nails addition smaller nails (K020240); IP-XS Compression Nail System (K032548); TriGen Straight Humeral Nail System (K032722); Friedl Gliding Nails Gliding nails with chamfered blade (K033763) TriGen InterTAN Nail (K040212)… | 2003 |
| * | K033763 | Modification to Slim Gliding Nail SystemSummary page 1…IP-XS Compression Nail System (K032548); TriGen Straight Humeral Nail System (K032722); Friedl Gliding Nails Gliding nails with chamfered blade (K033763) TriGen InterTAN Nail (K040212) TriGen 1300 TAN Nails (K(040462); Intramedullary Hip Screw (K(04… | 2003 |
| * | K040212 | Trigen IntertanSummary page 1…; TriGen Straight Humeral Nail System (K032722); Friedl Gliding Nails Gliding nails with chamfered blade (K033763) TriGen InterTAN Nail (K040212) TriGen 1300 TAN Nails (K(040462); Intramedullary Hip Screw (K(040656); TriGen Adolescent TAN (K… | 2004 |
| * | K040462 | Trigen Trochanteric Antegrade NailSummary page 1…tem (K032722); Friedl Gliding Nails Gliding nails with chamfered blade (K033763) TriGen InterTAN Nail (K040212) TriGen 1300 TAN Nails (K(040462); Intramedullary Hip Screw (K(040656); TriGen Adolescent TAN (K(040929) TriGen Hindfoot Fusion N… | 2004 |
| * | K040656 | Modification to Smith & Nephew Intramedullary Hip ScrewSummary page 1…liding nails with chamfered blade (K033763) TriGen InterTAN Nail (K040212) TriGen 1300 TAN Nails (K(040462); Intramedullary Hip Screw (K(040656); TriGen Adolescent TAN (K(040929) TriGen Hindfoot Fusion Nail (HFN) (K(043052); Titanium Knee F… | 2004 |
| * | K040929 | Trigen Adolescent TanSummary page 1…TriGen InterTAN Nail (K040212) TriGen 1300 TAN Nails (K(040462); Intramedullary Hip Screw (K(040656); TriGen Adolescent TAN (K(040929) TriGen Hindfoot Fusion Nail (HFN) (K(043052); Titanium Knee Fusion Nail (1(050938); TriGen Meta… | 2004 |
| * | K043052 | Trigen Hindfoot Fusion NailSummary page 1…en 1300 TAN Nails (K(040462); Intramedullary Hip Screw (K(040656); TriGen Adolescent TAN (K(040929) TriGen Hindfoot Fusion Nail (HFN) (K(043052); Titanium Knee Fusion Nail (1(050938); TriGen Meta-Nail Retrograde Femoral and Tibial Nails (K(061019);… | 2004 |
| * | K050938 | Knee Fusion NailSummary page 1…ullary Hip Screw (K(040656); TriGen Adolescent TAN (K(040929) TriGen Hindfoot Fusion Nail (HFN) (K(043052); Titanium Knee Fusion Nail (1(050938); TriGen Meta-Nail Retrograde Femoral and Tibial Nails (K(061019); TriGen Stainless Steel Knee Fusion Nails… | 2005 |
| * | K061019 | Trigen Meta-Nail Retrograde Femoral and Tibial NailsSummary page 1…TriGen Hindfoot Fusion Nail (HFN) (K(043052); Titanium Knee Fusion Nail (1(050938); TriGen Meta-Nail Retrograde Femoral and Tibial Nails (K(061019); TriGen Stainless Steel Knee Fusion Nails (K(061783); SURESHOT TAN Nails and Accessories (K(092748);… | 2006 |
| * | K061783 | Knee Fusion NailSummary page 1…Titanium Knee Fusion Nail (1(050938); TriGen Meta-Nail Retrograde Femoral and Tibial Nails (K(061019); TriGen Stainless Steel Knee Fusion Nails (K(061783); SURESHOT TAN Nails and Accessories (K(092748); 4.5mm and 5.0mm TriGen Low Profile Bone Screw (1(111025);… | 2006 |
| * | K092748 | Sureshot Tan Nails and AccessoriesSummary page 1…eta-Nail Retrograde Femoral and Tibial Nails (K(061019); TriGen Stainless Steel Knee Fusion Nails (K(061783); SURESHOT TAN Nails and Accessories (K(092748); 4.5mm and 5.0mm TriGen Low Profile Bone Screw (1(111025); Device Description Subject of this Traditional 510(… | 2010 |
| * | K111025 | Trigen Low Profile Bone ScrewSummary page 1…riGen Stainless Steel Knee Fusion Nails (K(061783); SURESHOT TAN Nails and Accessories (K(092748); 4.5mm and 5.0mm TriGen Low Profile Bone Screw (1(111025); Device Description Subject of this Traditional 510(k) Premarket Notification are the Smith & Nephew, Inc. Intramedull… | 2011 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 7, 2019 | Class II | Z-0351-2020 | Process control | LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 14 years and passes through 1 company; a new generation was cleared every 2.5 years on median.
- 13 of the 35 ancestors have a recall of any class.
- 2 other devices also cite its predicate K811002.
How does its category compare?
All HSB devicesHSB Rod, Fixation, Intramedullary And Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.
FDA decisions per year · HSB
15in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HSB
106in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HSB
18in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HSB
2,911in 2025
MAUDE reports naming product code HSB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 8 | 68 | 0 | 162 |
| 2008 | 11 | 57 | 9 | 184 |
| 2009 | 18 | 86 | 81 | 334 |
| 2010 | 14 | 117 | 18 | 454 |
| 2011 | 16 | 78 | 2 | 662 |
| 2012 | 17 | 112 | 7 | 1,079 |
| 2013 | 18 | 105 | 14 | 1,773 |
| 2014 | 19 | 147 | 3 | 1,474 |
| 2015 | 22 | 111 | 3 | 1,671 |
| 2016 | 8 | 124 | 6 | 1,920 |
| 2017 | 23 | 119 | 66 | 1,447 |
| 2018 | 17 | 134 | 56 | 3,526 |
| 2019 | 13 | 132 | 6 | 2,517 |
| 2020 | 17 | 112 | 6 | 3,299 |
| 2021 | 10 | 94 | 2 | 4,111 |
| 2022 | 12 | 172 | 1 | 2,733 |
| 2023 | 11 | 162 | 1 | 2,219 |
| 2024 | 13 | 195 | 1 | 2,869 |
| 2025 | 15 | 106 | 18 | 2,911 |
| 2026 | 18 | 151 | 1 | — |
What is it indicated for?
Smith & Nephew Trigen Adolescent TAN Nail System and Trigen 1300 TAN Nail System Instruments are accessory devices and are intended to be used to assist in the implantation of the Smith & Nephew Trigen Adolescent TAN Nail System and the Smith & Nephew Trigen 1300 TAN Nail System and their cleared indications for use. Indications for interlocking intramedullary nails include: * Simple long bone fractures * Severely comminuted, spiral, large oblique and…
Read the full statement (1,844 words)
Smith & Nephew Trigen Adolescent TAN Nail System and Trigen 1300 TAN Nail System Instruments are accessory devices and are intended to be used to assist in the implantation of the Smith & Nephew Trigen Adolescent TAN Nail System and the Smith & Nephew Trigen 1300 TAN Nail System and their cleared indications for use.
Indications for interlocking intramedullary nails include:
* Simple long bone fractures
* Severely comminuted, spiral, large oblique and segmental fractures
* Nonunions and malunions
* Poly trauma and multiple fractures
* Prophylactic nailing of impending pathologic fractures
* Reconstruction following tumor resection and grafting
* Supracondylar fractures
* Bone lengthening and shortening
* Fractures that occur in and between the proximal and distal third of the long bones being treated
In addition to the indications for interlocking intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability are indicated for the following:
* Subtrochanteric fractures with lesser trochanteric involvement
* Ipsilateral femoral shaft/neck fractures
* Intertrochanteric fractures
Smith & Nephew Trigen Nail System Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Trigen System and its cleared indications for use.
TRIGEN 0 Humeral Nail System is indicated for:
* Proximal and/or diaphyseal fractures of the humerus
* Nonunions
* Malalignments
* Pathological humeral fractures
* Impending pathological fractures
Smith & Nephew Trigen InterTAN Nail System Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Trigen InterTAN Nail System and its cleared indications for use.
TRIGEN InterTAN Nails are indicated for:
* Simple long bone fractures
* Severely comminuted spiral, long oblique, and segmental shaft
* Nonunions and malunions
* Poly trauma and multiple fractures
* Prophylactic nailing of impending pathologic fractures
* Reconstruction, following tumor resection and grafting
* Bone lengthening and shortening
* Subtrochanteric fractures
* Ipsilateral femoral shaft/neck fractures
* Intertrochanteric fractures
* Intracapsular fractures
Smith & Nephew Intramedullary Nail Systems Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Intramedullary Nail Systems and their cleared indications for use.
The Short Knee Nail, REVISION 0 Nail and Hindfoot Fusion Nail (HFN) are indicated for:
* Degeneration, deformity, or trauma of both the tibiotalar and talocalcaneal articulations of the hindfoot
* Tibiocalcaneal arthrodesis
* Combined arthrodesis of the ankle and sub-talar joints
* Avascular necrosis of the ankle and sub-talar joints
* Failed total ankle replacement with sub-talar intrusion
* Failed ankle arthrodesis with insufficient talar body
* Rheumatoid arthritis
* Severe deformity secondary to untreated talipes eguinovarus or neuromuscular disease
* Severe pilon fractures with trauma to the subtalar joint
Smith & Nephew SURESHOT TAN Nail Systems Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew SURESHOT TAN Nail Systems and their cleared indications for use.
SURESHOT TAN Nails are indicated for:
* Fractures of the femur including:
o simple long bone fractures
o severely comminuted, spiral, large oblique and segmental fractures
* Non unions and malunions
* Poly trauma and multiple fractures
* Prophylactic nailing of impending pathologic fractures
* Reconstruction, following tumor resection and grafting
* Supracondylar fractures
* Bone lengthening and shortening
* Fixation of fractures that occur in and between the proximal third and distal fourth of the femur
In addition, SURESHOT TAN Nails contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability and are indicated for the following:
* Subtrochanteric fractures;
* Intertrochanteric fractures;
* Ipsilateral femoral shaft/neck fractures;
* Intracapsular fractures.
Smith & Nephew Intramedullary Hip Screw System Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Intramedullary Hip Screw Systems and their cleared indications for use.
Intramedullary Hip Screws are indicated for:
* Intracapsular fractures of the femoral neck;
* Trochanteric or subtrochanteric fractures;
* Osteotomies for patients with diseases or deformities of the hip;
* Hip arthrodesis;
* Supracondylar fractures and distal femoral fractures using a supracondylar plate;
* Ipsilateral femoral shaft/neck fractures;
* Intertrochanteric fractures;
* Femoral neck fractures;
* Subrapital fractures;
* Comminuted neck and shaft fractures;
* Femur reconstruction following tumor resection;
* Leg length discrepancies secondary to femoral inequality;
* Prophylactic nailing of impending pathologic fractures.
Smith & Nephew Ender Nail System Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Ender Nail Systems and their cleared indications for use.
ENDER Nails are indicated for:
* Fracture of the neck, trochanteric, and subtrochanteric region of the femur;
* Distal femoral fractures with a distal fragment 10cm or longer
* Tibial shaft fractures
* Proximal humeral fractures
Smith & Nephew TriMax Nail System Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew TriMax Nail Systems and their cleared indications for use.
Femoral/Recon antegrade nails and retrograde nails are indicated for:
* Shaft fractures including severely comminuted, spiral, large oblique and segmental fractures
* Non-unions and malunions
* Bone lengthening/shortening
* Severely comminuted shaft fractures
* Pathologic fractures, pseudloarthrosis, failed osteosynthesis
* Closed supracondylar fractures
* Prophylactic nailing of impeding pathologic fractures
Additional indications for the femoral/recon antegrade include:
* Subtrochanteric fractures with lesser trochanteric involvement
* Ipsilateral femoral shaft/neck fractures
Additional indications for retrograde nails include:
* Severely comminuted supracondylar fractures with or without difficult intra-articular extension
* Fractures that require opening the knee joint to stabilize the femoral condylar segment
* Fractures above total knee implants
The TriMax Nail System is intended to be removed upon fracture healing.
Indications for interlocking intramedullary nails include:
* Simple long bone fractures
* Severely comminuted, spiral, large oblique and segmental fractures
* Nonunions and malunions
* Poly trauma and multiple fractures
* Prophylactic nailing of impending pathologic fractures
* Reconstruction following tumor resection and grafting
* Supracondylar fractures
* Bone lengthening and shortening
* Fractures that occur in and between the proximal and distal third of the long bones being treated
In addition to the indications for interlocking intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability (e.g. Femoral Antegrade Nail, Trochanteric Antegrade Nail and Femoral/Recon Antegrade Nail) are indicated for the following:
* Subtrochanteric fractures
* Intertrochanteric fractures
* Ipsilateral femoral shaft/neck fractures
In addition to the indications for interlocking intramedullary nails, devices that use a retrograde femoral surgical approach (e.g. Knee Nail, Retrograde and Supracondylar Nails) are indicated for:
* Comminuted supracondylar fractures with or without intra-articular extension
* Fractures that require opening the knee joint to stabilize the femoral condylar segment
* Fractures above total knee implants (peri-prosthetic fractures)
Smith & Nephew Knee Fusion Nail Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew Knee Fusion Nails and their cleared indications for use.
Knee Fusion Nails are indicated for:
* Intramedullary Knee Arthrodesis
Smith & Nephew IP-XS Compression Nail System Instruments are accessory devices and are intended to be used to assist in the implantation of the Smith & Nephew IP-XS Compression Nail System and its cleared indications for use.
The IP-XS Compression Nail System is inserted into the medullary canal of long bones for the:
* Alignment, stabilization, and fixation of fractures caused by disease or trauma
* The fixation of long bones that have been surgically prepared (osteotomy) for correction of deformity
* Arthrodesis
Smith & Nephew SLIM (Friedi) Gliding Nail Systems Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew SLIM (Friedi) Gliding Nails and their cleared indications for use.
The SLIM (Friedl) Gliding Nail System is an all-purpose locking nail system for ensuring primary load stability in:
* Petrochanteric femoral fractures
* Subtrochanteric femoral fractures
* Lateral femoral neck fractures
Internal fixation with the SLIM (Friedl) Gliding Nail System is indicated in all combination injuries involving the lateral femoral neck or trochanter region and femoral shaft fractures. Thanks to its biomechanical characteristics, the SLIM (Friedl) Gliding Nail System is also suitable for medical femoral neck fractures with retention of the head and simple femoral shaft fractures down to the distal metaphysis. Gliding nail fixation can also be used to secure pathological fractures or to provide weight-bearing stability after varus and valgus revision osteotomies of the proximal femur.
Smith & Nephew TriGen Low Profile Bone Screw Instruments are accessory devices and are intended to be used to assist in the implantation of Smith & Nephew TriGen Low Profile Bone Screws and their cleared indications for use.
The TRIGEN TM Low Profile Bone Screw can be used with several types of nails in Smith & Nephew's (TRIGEN TM) Titanium Nail System. The TRIGEN TM Low Profile Bone Screw therefore has the following indications:
Indications for interlocking intramedullary nails include simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated.
In addition to the indications for interlocking intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability (e.g. Femoral Antegrade Nail, Trochanteric Antegrade Nail and Femoral/Recon Antegrade Nail) are indicated for the following: subtrochanteric fractures, intertrochanteric fractures, and ipsilateral femoral shaft/neck fractures.
In addition to the indications for interlocking intramedullary nails, devices that use a retrograde femoral surgical approach (e.g. Knee Nail, Retrograde and Supracondylar Nails) are indicated for the following: comminuted supracondylar fractures with or without intra-articular extension; fractures that require opening the knee joint to stabilize the femoral condylar segment; fractures above total knee implants (peri-prosthetic fractures).
The TRIGEN TM InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fractures; subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; intracapsular fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; bone lengthening and shortening.
SURESHOT TM TAN Nails are indicated for fractures of the femur including simple long bone fractures, severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening; and for fixation of fractures that occur in and between the proximal third and distal fourth of the femur.
In addition, SURESHOT TM TAN Nails contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability and are indicated for the following: subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; and intracapsular fractures.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K122170 · printed Nov 6, 2026
Cite this page
510(k) Report. “K122170: Smith & Nephew, Inc. Intramedullary Nail System Instrumentation.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K122170
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026