PMT Truscan Surface Electrode

K122376510(k) TraditionalCleared Dec 19, 2012135-day reviewClass II21 CFR 882.1320

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pmt Corp. · GXY Electrode, Cutaneous

Cleared on Dec 19, 2012 as substantially equivalent to K032278 (Ambu, Inc., 2004) after a 135-day review. Its family tree reaches back 1 generation to 2004. 1 later device descends from it.

Decided
Dec 19, 2012
Review
135 days
Class
Class II
Product code
GXY
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K032278 (2004). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2017: 1 descendant clearedK032278 · Ambu Neuroline Single Patient Eeg/EP Cup Electrode · Ambu, Inc. · 2004200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant K032278 · Ambu Neuroline Single Patient Eeg/EP Cup Electrode · Ambu, Inc. · 20040
  1. Root · 1 generation backK032278 Ambu Neuroline Single Patient Eeg/EP Cup ElectrodeAmbu, Inc. · 2004
  2. This deviceK122376 PMT Truscan Surface ElectrodePmt Corp. · 2012
  3. Descendant · cites it directly · 0 of its ownK171102 Ives MR Conditional Cup ElectrodeIves Egg Solutions · 2017

Most influential descendants

Showing 1 of 1 · See all
  • K171102Ives MR Conditional Cup ElectrodeIves Egg Solutions · 2017 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K032278Ambu Neuroline Single Patient Eeg/EP Cup ElectrodeAmbu, Inc.2004—
0 (this device)K122376PMT Truscan Surface ElectrodePmt Corp.2012—
+1K171102Ives MR Conditional Cup ElectrodeIves Egg Solutions2017—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
FIG. 1 — Lineage of K122376, 2004–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GXY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GXY device

  1. #1 K232052 Ceribell Instant EEG Headbandscore 16.3 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K032278Ambu Neuroline Single Patient Eeg/EP Cup ElectrodeAmbu, Inc.2004—
*—Grass Instruments
Named without a submission number
—

Also cleared against K032278: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171102Ives MR Conditional Cup ElectrodeIves Egg Solutions20170

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K171102 (Ives Egg Solutions, 2017) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K032278.

How does its category compare?

All GXY devices

GXY Electrode, Cutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · GXY

6in 2025

2007: 72008: 92009: 172010: 32011: 62012: 92013: 72014: 92015: 62016: 62017: 92018: 142019: 92020: 42021: 122022: 52023: 132024: 72025: 62026 (partial year): 12
2007high 17 (2009) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GXY

148in 2025

2007: 88 days median2008: 136 days median2009: 159 days median2010: 137 days median2011: 183 days median2012: 150 days median2013: 99 days median2014: 203 days median2015: 119 days median2016: 170 days median2017: 116 days median2018: 135 days median2019: 113 days median2020: 108 days median2021: 195 days median2022: 81 days median2023: 132 days median2024: 111 days median2025: 148 days median2026 (partial year): 70 days median
2007high 203 (2014) · low 81 (2022)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GXY

2in 2023

2009: 42010: 12011: 52012: 42023: 2
2007high 5 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GXY

58in 2025

2007: 142008: 172009: 152010: 142011: 162012: 162013: 152014: 112015: 192016: 152017: 112018: 102019: 192020: 112021: 212022: 342023: 252024: 512025: 58
2007high 58 (2025) · low 10 (2018)2025

MAUDE reports naming product code GXY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007788014
20089136017
200917159415
20103137114
20116183516
20129150416
2013799015
20149203011
20156119019
20166170015
20179116011
201814135010
20199113019
20204108011
202112195021
2022581034
202313132225
20247111051
20256148058
202612700—

What is it indicated for?

The PMT® TruScan® Surface Electrodes are indicated for cutaneous use in the general recording and monitoring of the Electroencephalograph (EEG) and Evoked Potential (EP). The PMT® TruScan® Surface Electrodes are CT compatible and MR Conditional under the following conditions: Static magnetic field strength of 1.5-T only Maximum spatial gradient magnetic field of 5,000 Gauss/cm (50T/m) or less The connector hub must be placed near the center MR system's…

Read the full statement (139 words)

The PMT® TruScan® Surface Electrodes are indicated for cutaneous use in the general recording and monitoring of the Electroencephalograph (EEG) and Evoked Potential (EP).

The PMT® TruScan® Surface Electrodes are CT compatible and MR Conditional under the following conditions:

Static magnetic field strength of 1.5-T only

Maximum spatial gradient magnetic field of 5,000 Gauss/cm (50T/m) or less

The connector hub must be placed near the center MR system's bore, and must be at least 20-cm from the wall of the MR system's bore at all times.

The extension cable must be disconnected from the PMT TruScan Surface Electrode before scanning and MUST remain disconnected throughout the entire MR scan.

Normal Operating Mode of operation for the MR system with a maximum whole body averaged specific absorption rate (SAR) of 2.0-W/kg for 15 minutes of scanning (i.e., per pulse sequence)

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K122376: PMT Truscan Surface Electrode.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K122376

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026