510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Iolmaster 500

K122418510(k) TraditionalCleared Apr 12, 2013247-day reviewClass II21 CFR 886.1850Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Carl Zeiss Meditec, AG · HJO Biomicroscope, Slit-Lamp, Ac-Powered

Cleared on Apr 12, 2013 as substantially equivalent to K101182 (Carl Zeiss Meditec, AG, 2010) after a 247-day review. Its family tree reaches back 6 generations to 1996. 9 later devices descend from it. It has been recalled 4 times.

Decided
Apr 12, 2013
Review
247 days
Class
Class II
Product code
HJO
Recalls
4
Predicates
1

Its longest line of descent runs back 6 generations to K955442 (1996). 20 ancestors in all, the oldest K781994 (1979). 9 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor2003: 1 other ancestor1996: 2 other ancestors1994: 2 other ancestors1993: 2 other ancestors1979: 1 other ancestor1990: 1 other ancestor1987: 2 other ancestors1986: 1 other ancestor1997: 1 other ancestor2016: 1 descendant cleared2015: 1 descendant cleared2019: 2 descendants cleared2017: 1 descendant cleared2021: 1 descendant cleared2020: 1 descendant cleared2018: 1 descendant cleared2025: 1 descendant clearedK955442 · Kowa SC-1200 Photo Slit Lamp System · Kowa Optimed, Inc. · 1996K991284 · Anterior Eye-Segment Analysis System · Nidek, Inc. · 1999K012873 · Documenting Laser Slit Lamp, DP 2010 · Bausch & Lomb, Inc. · 2001K030393 · Optical Low Coherence Reflectometry (Olcr) Pachymeter · Haag-Streit Holding AG · 2003K082891 · Lenstar, Model LS900 · Haag-Streit AG · 2009 · recalledK101182 · Iolmaster 500 · Carl Zeiss Meditec, AG · 2010 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 6 other ancestors4 generations back: 4 other ancestors5 generations back: 2 other ancestors6 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 5 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant K101182 · Iolmaster 500 · Carl Zeiss Meditec, AG · 2010 · recalledK082891 · Lenstar, Model LS900 · Haag-Streit AG · 2009 · recalledK030393 · Optical Low Coherence Reflectometry (Olcr) Pachymeter · Haag-Streit Holding AG · 2003K012873 · Documenting Laser Slit Lamp, DP 2010 · Bausch & Lomb, Inc. · 2001K991284 · Anterior Eye-Segment Analysis System · Nidek, Inc. · 1999K955442 · Kowa SC-1200 Photo Slit Lamp System · Kowa Optimed, Inc. · 1996−50
  1. Root · 6 generations backK955442 Kowa SC-1200 Photo Slit Lamp SystemKowa Optimed, Inc. · 1996
  2. 5 generations backK991284 Anterior Eye-Segment Analysis SystemNidek, Inc. · 1999
  3. 4 generations backK012873 Documenting Laser Slit Lamp, DP 2010Bausch & Lomb, Inc. · 2001
  4. 3 generations backK030393 Optical Low Coherence Reflectometry (Olcr) PachymeterHaag-Streit Holding AG · 2003
  5. 2 generations backK082891 Lenstar, Model LS900Haag-Streit AG · 2009 · Recalled
  6. 1 generation backK101182 Iolmaster 500Carl Zeiss Meditec, AG · 2010 · Recalled
  7. This deviceK122418 Iolmaster 500Carl Zeiss Meditec, AG · 2013
  8. Descendant · cites it directly · 4 of its ownK143275 IOLMaster700Carl Zeiss Meditec, Inc. · 2015
  9. Descendant · 2 generations on · 3 of its ownK170171 IOLMaster 700Carl Zeiss Meditec, AG · 2017 · Recalled
  10. Descendant · cites it directly · 3 of its ownK152311 Pentacam AXLOculus Optikgerate GmbH · 2016 · Recalled
  11. Descendant · 3 generations on · 1 of its ownK173771 IOLMaster 700Carl Zeiss Meditec, AG · 2018 · Recalled
  12. Descendant · 4 generations on · 0 of its ownK252633 Unity DX (Udx)Cylite Pty., Ltd. · 2025

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2015: 12016: 12017: 12018: 12019: 22020: 12021: 12025: 1
2015high 2 (2019) · low 0 (2022)2026

Most influential descendants

Showing 5 of 9 · See all
  • K143275IOLMaster700Carl Zeiss Meditec, Inc. · 2015 · cites it directly4descendants
  • K170171IOLMaster 700Carl Zeiss Meditec, AG · 2017 · 2 generations on3descendants
  • K152311Pentacam AXLOculus Optikgerate GmbH · 2016 · cites it directly3descendants
  • K173771IOLMaster 700Carl Zeiss Meditec, AG · 2018 · 3 generations on1descendant
  • K252633Unity DX (Udx)Cylite Pty., Ltd. · 2025 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K955442Kowa SC-1200 Photo Slit Lamp SystemKowa Optimed, Inc.1996—
−5K991284Anterior Eye-Segment Analysis SystemNidek, Inc.1999—
−4K012873Documenting Laser Slit Lamp, DP 2010Bausch & Lomb, Inc.2001—
−3K030393Optical Low Coherence Reflectometry (Olcr) PachymeterHaag-Streit Holding AG2003—
−2K082891Lenstar, Model LS900Haag-Streit AG20093
−1K101182Iolmaster 500Carl Zeiss Meditec, AG20104
0 (this device)K122418Iolmaster 500Carl Zeiss Meditec, AG2013—
+1K143275IOLMaster700Carl Zeiss Meditec, Inc.2015—
+2K170171IOLMaster 700Carl Zeiss Meditec, AG20171
+1K152311Pentacam AXLOculus Optikgerate GmbH20163
+3K173771IOLMaster 700Carl Zeiss Meditec, AG20181
+4K252633Unity DX (Udx)Cylite Pty., Ltd.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198610
198720
199010
199320
199420
199620
199710
200010
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201610
201710
201810
201920
202010
202110
202510
FIG. 1 — Lineage of K122418, 1996–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HJO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HJO device

  1. #1 K253834 Visuref 600score 13.1 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K101182Iolmaster 500
Summary page 2
…lOLMaster 500, the subject of this 510(k), is substantially equivalent to the predicate lOLMaster 500 (Carl Zeiss Meditec, AG) cleared for marketing under 510(k) number K101 182 in October 20 10. DEVICE DESCRIPTION The lOLMaster 500 is a non-contact biometry instrument for measurements of the eye require…
Carl Zeiss Meditec, AG20104

What descends from it? (9)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K143275IOLMaster700Carl Zeiss Meditec, Inc.20154
K152311Pentacam AXLOculus Optikgerate GmbH20163

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 27, 2015Class IIZ-0357-2016Software designIOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
Oct 27, 2015Class IIZ-0358-2016Software designIOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
Oct 23, 2013Class IIZ-0333-2014Software in the Use EnvironmentThe user database delivered for download contained an erroneous value. Specifically the "aO" constant of the lens AMO Sensar AR40E is listed as "0.0" in the internal lens database of the IOLMaster. The correct value is -2.420. This could…
Oct 23, 2013Class IIZ-0334-2014Software in the Use EnvironmentThe user database delivered for download contained an erroneous value. Specifically the "aO" constant of the lens AMO Sensar AR40E is listed as "0.0" in the internal lens database of the IOLMaster. The correct value is -2.420. This could…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 6 companies; a new generation was cleared every 2.4 years on median.
  • 3 of the 20 ancestors have a recall of any class.
  • Of its descendants, K143275 (Carl Zeiss Meditec, Inc., 2015) has gone furthest, with 4 descendants of its own.

How does its category compare?

All HJO devices

HJO Biomicroscope, Slit-Lamp, Ac-Powered: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · HJO

1in 2024

2007: 12008: 42009: 32010: 52011: 22013: 22014: 102015: 22016: 52017: 92018: 22020: 12022: 12024: 12026 (partial year): 2
2007high 10 (2014) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HJO

232in 2017

2008: 117 days median2009: 91 days median2010: 121 days median2014: 196 days median2016: 243 days median2017: 232 days median
2007high 243 (2016) · low 91 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HJO

1in 2018

2013: 22015: 22018: 1
2007high 2 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HJO

20in 2025

2007: 12008: 32009: 32010: 22011: 52012: 72013: 162014: 42015: 82016: 152017: 152018: 112019: 132020: 192021: 142022: 62023: 182024: 182025: 20
2007high 20 (2025) · low 1 (2007)2025

MAUDE reports naming product code HJO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
2008411703
200939103
2010512102
20112—05
20120—07
20132—216
20141019604
20152—28
20165243015
20179232015
20182—111
20190—013
20201—019
20210—014
20221—06
20230—018
20241—018
20250—020
20262—0—

What is it indicated for?

The IOLMaster is intended for the biometric determination of ocular measurements of axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil. For patients who are candidates for intraocular lens (IOL) implantation, the device also performs calculations to assist physicians in determining the appropriate IOL power and type for implantation.

This device is intended for use by physicians and eye-care professionals and may only be used under the supervision of a physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K122418: Iolmaster 500.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K122418

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026